Practical applications: CCD spectroscopy

Practical applications: CCD spectroscopy. Tracing path of 2-d spectrum across detector Measuring position of spectrum on detector Fitting a polynomial to measured spectrum positions Optimal extraction of spectra from CCD images with simultaneous sky background subtraction:

tuari
Download Presentation

Practical applications: CCD spectroscopy

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Practical applications: CCD spectroscopy • Tracing path of 2-d spectrum across detector • Measuring position of spectrum on detector • Fitting a polynomial to measured spectrum positions • Optimal extraction of spectra from CCD images with simultaneous sky background subtraction: • Scaling a profile + constant background • Wavelength calibration of 1-d spectra • Measurement of positions of arc-line images • Fitting a polynomial to measured positions of images of arc lines with known wavelengths

  2. x  To zenith Target spectrum x (Reference spectrum) Night sky lines  Observing hints • Rotate detector so that arc lines are parallel to columns: • To minimise slit losses due to differential refraction, rotate slit to “parallactic angle” • i.e. keep it vertical: • Spectra are then tilted or curved due to • camera distortions • Differential refraction

  3. After bias subtraction and flat fielding • Recall Lecture 3 for subtraction of B(x,), construction of flat field F(x,) and measurement of gain factor G. • Corrected image values are

  4. x  x  Tracing the spectrum • Spectra may be tilted, curved or S-distorted. • Trace spectrum via a sequence of operations: • divide into -blocks • measure centroid of spectrum in each block (fit gaussian) • fit polynomial in  to calibrate x0(). • Once this is done, use x0() to select object/sky regions on subsequent steps. x0 x0

  5.  on edge of night-sky emission line Target Ref star x  away from night-sky emission line Slices across spectrum at=const: Sky subtraction • Alignment (rotation) of CCD detector relative to grating aims to make ~const along columns. • Imperfect alignment gives slow change in  along columns. • This causes gradient, curvature of sky background when  is close to a night-sky line. • Solution: fit low-order polynomial in x to sky background data. • Alternative: fit linear function to interpolate sky from “sky regions” symmetric on either side of object spectrum: Target Ref

  6. Slice across spectrum at =const: S(x) x x1 x2 “Normal” extraction • Subtract sky fit, and sum the counts between object limits: • Dilemma: How do we pick x1, x2? • too wide: too much noise • too narrow: lose counts

  7. Slice across spectrum at =const: Starlight profile P(x) S(x) x x1 x2 Optimal extraction • 1) Scale profile to fit the data: • 2) Compute (x) from the model: • 3) -clip to “zap” cosmic-ray hits. • Iterate 1 to 3, since (x) depends on A:

  8. Estimating the profile P(x) • The fraction of the starlight that falls in row x varies along the spectrum and is given by: • This is an unbiased but noisy estimator of P(x). • It varies as a slow function of wavelength. • Plot against  and fit polynomials in  at each x. Column 20 Column 60

  9. Optimal vs. normal extraction • Pros: • Optimal extraction gives lower statistical noise. • Equivalent to longer exposure time • Incorporates cosmic-ray rejection • Cons: • Requires P(x,) slowly varying in  (point sources). • Essential papers: • Horne, K., 1986. PASP 98, 609 • Marsh, T. and Horne, K.

  10. Wavelength calibration • Select lines using peak threshold. • Measure pixel centroid xi by computing x or fitting a gaussian • Identify wavelengths i • Fit polynomial (x) to i, i=1,...,N. • Reject outliers (usually close blends) • Adjust order of polynomial to follow structure without too much “wiggling”. threshold level

  11. Dealing with flexure • Flexure of spectrograph causes position xi of a given wavelength  to drift with time. • Measure new arcs at every new telescope position. • Interpolate arcs taken every 1/2 to 1 hour when observing at same position. • Master arcfit: Use a long-exposure arc (or sum of many short arcs) to measure faint lines and fit high-order polynomial. • Then during night take short arcs to “tweak” the low-order polynomial coefficients.

  12. Statistical issues raised • Outlier rejection: what causes outliers, and how do we deal with them? • Robust statistics. • Polynomial fitting: how many polynomial terms should we use? • Too few will under-fit the data. • Too many can introduce “flailing” at ends of range. • We’ll deal with these issues in the next lecture.