S.A. Erzurum, MD FACS Aylin Sarac

Reliability of Aberrometry Versus Cycloplegic Retinoscopy in Prescribing Glasses to Undilated Child. S.A. Erzurum, MD FACS Aylin Sarac. Financial Interest. The authors have no financial interests in any of the topics discussed in this study. Purpose of Study.

turi
Download Presentation

S.A. Erzurum, MD FACS Aylin Sarac

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Reliability of Aberrometry Versus Cycloplegic Retinoscopy in Prescribing Glasses to Undilated Child S.A. Erzurum, MD FACS Aylin Sarac

  2. Financial Interest • The authors have no financial interests in any of the topics discussed in this study.

  3. Purpose of Study • Retinoscopy is a difficult skill for many eye care professionals. • Children are a difficult subpopulation for eye care professional to refract. • Inaccurate glass prescribing occurs for many children. • Dilation necessary for accurate refraction but disliked by children and family.

  4. Purpose of Study • Find a reliable method for non-specialized eye care providers to prescribe glasses for children. • Reliable method should be easy and quick. • Find method that eliminates/reduces need for dilating drops to check refraction.

  5. Method • iZON Aberrometer was machine used for comparison to cycloplegic refraction. • Images are taken before dilation. • Child sits upon parent’s lap and multiple rapid images are obtained. • Immediately upon taking image, analysis is provided of sphere, cylinder, axis, level of aberrations, and quality of image.

  6. Example of reliable image capture. • Prescription measured and reported as sphere, cylinder, and axis. • Aberrations measured and reported in green, yellow or red. • Pictures demonstrate how patient sees at night.

  7. Example of poor image and unreliable readings. No measurements are provided for aberrations. Whatever prescription was captured is reported, but may not be reliable.

  8. Example of right eye with poor image capture and left eye with excellent capture. • Right eye results for refraction may not be reliable. • Left eye results reliable.

  9. Method • Prospective consecutive series of 256 eyes in 128 pediatric patients involved in full exam. • Age range of 3 to 18 years. • iZON aberrometer images prior to dilation. • Full eye exam with cycloplegic refraction. • No patients were excluded, and all were examined by the same pediatric ophthalmologist.

  10. Results • CYCLOPLEGIC ABERROMETER (mean +/-SD) (mean +/- SD) Sphere 0.42 + 2.95 -0.05 + 2.16 Astigmatism 0.74 + 0.91 0.92 + 1.52 Spheric equiv 0.75 + 2.91 0.39 + 2.17 Age (mean + SD) 8.1 + 3.2 years; Mode 7 years Range 3-18 years Female 62 48.4% Male 66 51.6%

  11. Prevalence of Eye Disorders in Study Patients • Prevalence of Eye Disorders • Normal 120 46.9 • Esotropia 32 11.3 • Amblyopia 29 12.5 • Exotropia 9 3.5 • Cerebral palsy/ 7 2.7 developmental delay • Congenital nystagmus 4 1.6 • Cataracts 3 1.2 • IDDM 4 1.6 • Duane’s 3 1.2 • Best’s 1 0.4 • Coloboma 1 0.4

  12. Reliability of Aberrometer • Only 45 eyes of 256 (17%)had an unreliable reading and were excluded from comparison. • 26 eyes of these 45 (64%)had ocular pathology which would have directed them to sub-specialized care. • Immediately upon taking image, examiner is aware of reliability of image.

  13. Results of Comparison between two Methods • Comparison between aberrometer and cycloplegic results were made in each area of sphere, cylinder, and axis. • In all categories, correlation was high between the two methods. • Offset and slope of the best fit regression were calculated for each. • All comparisons showed the slope was very close to 1.

  14. Y cyclo sphere = 1.04 X izon sphere + 0.469 Model ANOVA F = 1552.93, p < 0.001 R = 0.934

  15. Y cyclo SPHEQ = 1.019 X izon SPHEQ + 0.400 Model ANOVA F = 718.782 p < 0.001 R = 0.865

  16. Y cyclo astig = 0.909 X izon astig – 0.026 Model ANOVA F = 766.735 p < 0.001 R = 0.887

  17. Sphere Astigmatism Spherical Equivalent

  18. Conclusions • iZON aberrometer is a reliable method of refraction to assist practitioners uncomfortable with cycloplegic refraction in children. • iZON aberrometer gives a reliable reading and a high correlation with cycloplegic refraction. • Group of patients who showed higher unreliability by aberrometry measurements had a higher chance of ocular pathology.