PVSpecGeGa. H. Dannerbauer, U. Klaas, J. Bouwman. Motivation: characterization of the GeGa detector Preconditions: photometric standards and PACS Internal CSs telescope temperature < 100 K, final temp not required results from commissiong phase Contents:

ulric
Download Presentation

PVSpecGeGa

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PVSpecGeGa H. Dannerbauer, U. Klaas, J. Bouwman • Motivation: • characterization of the GeGa detector • Preconditions: • photometric standards and PACS Internal CSs • telescope temperature < 100 K, final temp not required • results from commissiong phase • Contents: • carrying out dark current measurements (PCD 1.2.6) • linearity of CRE read-out (PCD 1.2.11) • dynamic range per selected integration capacity (PCD 1.2.3)

  2. PVSpecGeGa • Dark Current Measurements: • starring mode; non-standard AORs • both CSs • in buffer transmission mode for a total of 420 sec (burst mode) • RI of 2sec • bias setting: bias_d_red= 0.03 V, bias_r_red=3.e-4 V, bias_d_blu=0.08 V, bias_r_blu=3.e-4 V • we use lowst capacity for both arrays • repeated twice and for both blue filters • should also address stray light aspects blue red from PICC-MA-TR-051 (FM-IST)

  3. PVSpecGeGa • Linearity of CRE read-out • Dynamic range per selected integration capacitor red, modul 10, pixel 10 from saturation behaviour and ramp linearity (ILT): test report PICC-MA-TR-045 should be soon released.

  4. PVSpecGeGa • Linearity/Dynamic Range measurements: • starring mode • non-standard AOR • 1) on calibrators • in loss-less compression mode (burst mode) • per observations one module (spatial pixel); in total five spatial pixels • selected moduls covers both supply groups • 2) on PACS Internal CSs • buffer transmission mode (burst mode) • repeated twice and for both blue filters • only on one CS1 is enough

  5. PVSpecGeGa from latest PV plan V0.83

  6. PVSpecGeGa Update/Modifications x2 5 11.5 5 +1/8s loss-less compression mode x2 we are thinking to repeat this block =22.1 hrs from latest PV plan V0.83

  7. PVSpecGeGa AOR for Linearity Measurements:

  8. PVSpecGeGa

  9. PVSpecGeGa from latest PV plan V0.83

  10. PVSpecGeGa • What still has to be done: • 1) DC measurements: • pointing needed also for dark current measurements (Dark Sky Field) • → to compare/have a reference to the FOV scans and the RSRF meas. on the same field • →eng measurements could be on brighter skybackground without pointing • do measurements also on CS2 • 2) Linearity/Dynamic Range measurements: • implement detector selection table for four more moduls, so far only done • for central spatial pixel • modify CUS script for buffer transmission mode measurements on CS1 • 3) CUS script should be tested during FS ILT as so far it has never been • carried out under test conditions

  11. PVSpecGeGa • Analysis: • already tools for analyzing the data are existing • in FM ILT and IST JS took care for (1) and • JeB for (2)