410 likes | 894 Views
Single Chamber Temporary Pacing Operations & Troubleshooting. Temporary Pacing Indications. Sick Sinus Syndrome (Patient Must Be Symptomatic and Documented with ECG). Symptomatic sinus bradycardia Symptomatic sinus arrest Suppression of ventricular ectopy resulting from bradycardia
E N D
Temporary Pacing Indications Sick Sinus Syndrome (Patient Must Be Symptomatic and Documented with ECG) • Symptomatic sinus bradycardia • Symptomatic sinus arrest • Suppression of ventricular ectopy resulting from bradycardia • Atrial fibrillation • Bradycardia/Tachycardia syndrome
Temporary Pacing Indications Heart Blocks • Type I and Type II second degree AV block • Acute bifascicular or trifascicular block • Complete AV block • Cardiac arrest with ventricular asystole
Other Temporary Pacing Indications Drug-Refractory Dysrhythmia • Overdrive ventricular pacing to suppress or prevent VEA (Ventricular Ectopic Activity) • Overdrive atrial pacing to “break” SVT, atrial flutter
Other Temporary Pacing Indications Cardiovascular Surgery • Coverage for anesthesia and surgery in patients with positive cardiac history • Treatment for CHB development during surgery • Augment cardiac output post operatively
Other Temporary Pacing Indications Diagnostic Uses • SSS pacing studies to determine SA node recovery • Stress testing for Coronary Artery Disease • Electrophysiologic (EP) studies
Insertion Sites Internal Jugular Vein External Jugular Vein Subclavian Vein Brachial Vein Femoral Vein
Lead Types • Endocardial/Transvenous lead • Transvenous lead is introduced into a vein and advanced into the heart • Epicardial/Myocardial lead • An epicardial lead attached to the outside of the heart is introduced through the chest wall
Lead Types • Bipolar lead system • The negative and positive electrodesare in contact with the heart
Lead Types • Unipolar lead system • The negative electrode is in contact with the heart and the positive (or ground) electrode is located elsewhere on the body
Cable Connectors • Connector pins on the lead(s) must be fully inserted in the patient connector block • Observe polarity • Distal = negative • Proximal = positive • Finger tighten only – no tools!
Temporary Pacing Parameters • Pacing rate (heart rate) • Output/stimulation threshold • Sensitivity
Pacing Rate Paced Interval Paced Interval
Output Output Pulse • The output dial regulates the current or movement of electrons Output/Current(ma) Pulse Width(ms)
Capture Depolarization of cardiac musclefollowing an electrical stimulus
Stimulation Threshold The minimum output pulse neededto consistently capture the heart 3 mA 2 mA 1 mA
Lead Maturation x3 x2 Threshold Amplitude Initial 1 2 3 4 5 10 15 Days Since Leads Applied
Sensitivity The degree that the pacing system “sees” or senses signals, controlled by the sensitivity setting which is graduated in millivolts (mV) Sensitivity (mV) 5 (mV) 2.5 (mV) 1.25 (mV)
Sensitivity The lower the setting, the more sensitive the pacemaker is to intracardial signals
NBG Codes 1st Letter 2nd Letter 3rd Letter Chamber(s) Paced A = atrium V = ventricle D = dual (both atrium and ventricle) Chamber(s) Sensed A = atrium V = ventricle D = dual O = none Response to Sensing I = inhibit (Demand mode) T = triggered D = dual O = none (Asynch) Chamber paced Chamber sensed Action or response to a sensed event V V I
VVI Demand/Inhibited • Pacemaker senses intrinsic depolarization • Paces the heart when the patient’s own rate becomes slower than the pacemaker
VOO Asynchronous (Fixed) Pacemaker will emit an output at a fixed rateregardless of intrinsic activity
Single Chamber Troubleshooting Process 1. Gather information 2. Identify the problem and possible cause 3. Identify the solution and carry outcorrective procedures
Information Gathered • Patient data • Pacemaker information • Lead information • Non-invasive tools
Undersensing Failure of the pacemaker to senseintrinsic R-waves or intrinsic P-waves
Oversensing Inhibition of the pacemaker by events thepacemaker should ignore, e.g. EMI, T-waves,and myopotential
Fusion/Pseudofusion Beats Intrinsic Beat Paced Beat Intrinsic Beat Paced Beat Fusion Beat Pseudofusion Beat Fusion Beat Pseudofusion Beat
Electromagnetic Interference (EMI) Radiated or conducted energy – either electrical or magnetic – which can interfere with the function of the pacemaker in the Demand mode (EMI) Should havepaced 2.5 mV 80 80