1 / 13

NORM: Measurement & Analysis (Standardisation & Quality Assurance) Gert Jonkers

NORM: Measurement & Analysis (Standardisation & Quality Assurance) Gert Jonkers Engineering & Analytical - GSEA/4 “ Problem Solving ” (Shell E&P Ionising Radiation/NORM HSE Expert CHP) location Shell Research & Technology Centre, Amsterdam P.O. 38000 NL-1030 BN Amsterdam

vance
Download Presentation

NORM: Measurement & Analysis (Standardisation & Quality Assurance) Gert Jonkers

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. NORM: Measurement & Analysis (Standardisation & Quality Assurance) Gert Jonkers Engineering & Analytical - GSEA/4 “Problem Solving” (Shell E&P Ionising Radiation/NORM HSE Expert CHP) location Shell Research & Technology Centre, Amsterdam P.O. 38000 NL-1030 BN Amsterdam the Netherlands

  2. Product & Waste Streams to be Analysed on NOR’s

  3. 1 Bq[NOR]/g corresponding to ... g[NOR]/g ppt ppb ppm

  4. Nuclear Based Analytical Techniques

  5. g-Spectrometry (1) – Sample Containers Petri-dishes fix geometry Sealing dishes prevents escape of 222Rn Data acquisition time ~ 60,000 s Calibration for NORM (IAEA certified ores)

  6. g-Spectrometry (2) – Setup n-type high purity Germanium detectors for low energy range

  7. g-Spectrometry (3) – Setup

  8. g-Spectrometry (4) – Spectrometry g-yield data base used for analysis standardised 226Ra 186 keV line directly used (no 235U present) Correct intensity low energy (Eg < 100 keV, 210Pb) lines for self-absorption NOR’s analysed:226Raeq, 210Pbeq, 228Raeq & 228Theq,

  9. a-Spectrometry (1) – Spectrometry • 210Po determination in all kind of materials • Digest the sample in a mixture of concentrated acids • add a polonium (208Po) isotopic spike • Plate out polonium on a clean Ag-plate • Place plate below surface barrier detector in a vacuum chamber • Analyse the a-spectrum, where the internal reference is used for efficiency of plate out

  10. a-Spectrometry (2) – Setup Method is laborious Questionable, a separate 210Po is required from an HSE point of view.

  11. a-scintillation counting (1) – Setup (disassembled) Lucas cell method

  12. g-Spectrometry (2) – Setup Assembled Counting 222Rn concentration

  13. E&P NORM REQUIRED ANALYSES DATA(HSE point of view)