Correlation between visual impression and instrumental colour determination for LEDs János Schanda Professor Emeritus of the University of Pannonia, Hungary
Overview • Colour of LEDs • Problems with the colorimetry of LEDs • Photometric and colorimetric fundamentals • New colorimetric system? • Objective colorimetry of LEDs • Instrumental problems • LED standards • Recommendations
Colour of LEDs • LEDs are narrow band emitters • Small errors in colour matching functions produce errors. • Visual match differs from colorimetric match
Photometric and colorimetric fundamentals • Colorimetry is based on a trichromatic match between test colour stimulus and three primary (matching) colour stimuli • Transformation from real R,G,B primaries to imaginary X,Y,Z primaries • The CIE 1931 2° colorimetric system incorporated the CIE 1924 photometric observer
CIE 1924 photometric observer • Based on • Flicker photometry • Distinctintness of boarder • Small step colour difference • Describes • Visual acuity type observation: reading, observing fine details • NOT BRIGHTNESS
Spectral luminous efficiency functions and a new proposedy colour matching function
2° standard and proposed, cone fundamental based, colour matching functions (CMFs)
Broad-band – RGB-LED visual and instrumental colour match • Error decreased by 50 % or more Standard CMFs CMFs based on CIE TC 1-36 recommendation
LED colour characteristics • LEDs are narrow band emitters • bandwidth approx. 10 nm – 30 nm • Blue … Green: InGaN • Yellow … Red: AlInGaP • Both the absolute intensity and the wavelength of the emission maximum is temperature dependent • temperature dependence is composition dependent • largest changes with Red LEDs • Unusual spatial light characteristics • Solution of measurement problems caused by • Spectral mismatch: spectrometric measurement • Spectral mismatch: tristimulus colorimetry • Temperature dependence • Geometric misalignment
LED colour characteristics • LEDs are narrow band emitters • bandwidth approx. 10 nm – 30 nm • Blue … Green: InGaN • Yellow … Red: AlInGaP • Both the absolute intensity and the wavelength of the emission maximum is temperature dependent • temperature dependence is composition dependent • largest changes with Red LEDs • Unusual spatial light characteristics • Solution of measurement problems caused by • Spectral mismatch: spectrometric measurement • Spectral mismatch: tristimulus colorimetry • Temperature dependence • Geometric misalignment
LED colour characteristics • LEDs are narrow band emitters • bandwidth approx. 10 nm – 30 nm • Blue … Green: InGaN • Yellow … Red: AlInGaP • Both the absolute intensity and the wavelength of the emission maximum is temperature dependent • temperature dependence is composition dependent • largest changes with Red LEDs • Unusual spatial light characteristics • Solution of measurement problems caused by • Spectral mismatch: spectrometric measurement • Spectral mismatch: tristimulus colorimetry • Temperature dependence • Geometric misalignment
Irradiation inhomogeneity in measurement plane What should be reported? Problem of reproducible alignment
LED colour characteristics • LEDs are narrow band emitters • bandwidth approx. 10 nm – 30 nm • Blue … Green: InGaN • Yellow … Red: AlInGaP • Both the absolute intensity and the wavelength of the emission maximum is temperature dependent • temperature dependence is composition dependent • largest changes with Red LEDs • Unusual spatial light characteristics • Solution of measurement problems caused by • Spectral mismatch: spectrometric measurement • Spectral mismatch: tristimulus colorimetry • Temperature dependence • Geometric misalignment
Spectrometric measurements • Critical parameters of spectrometer • Sampling interval and bandpass: 10 nm sampling produces u’,v’ errors of several units in 3rd decimal. • Highly over sampling OK (CCD spectrometers) • Wavelength scale error: 0.5 nm error produces u’,v’ errors of several units in 3rd decimal. • Stray light: LED measurement compared to incandescent lamp if 10-4 stray light produces u’,v’ errors of several units in 3rd decimal. • Experiments showed even larger errors:
LED colour characteristics • LEDs are narrow band emitters • bandwidth approx. 10 nm – 30 nm • Blue … Green: InGaN • Yellow … Red: AlInGaP • Both the absolute intensity and the wavelength of the emission maximum is temperature dependent • temperature dependence is composition dependent • largest changes with Red LEDs • Unusual spatial light characteristics • Solution of measurement problems caused by • Spectral mismatch: spectrometric measurement • Spectral mismatch: tristimulus colorimetry • Temperature dependence • Geometric misalignment
Goodness of fit characterization • Modified f1’ method, • No illuminant • Independently for • Separately for Red, Green, Blue LEDs • Example: V(l) channel • Use coloured LED standards • f1’ provides estimate of error to be expected • Detector spectral responsivity measurement, not standardized properly
Partial f1’ error index 1 LED dominant wavelength ranges andthe dominant wavelength value of the standard LEDs
Partial f1’ error index 2 Maximal photometrical errors to be expected if the partial f1’ values are below the given limiting values
Correcting tristimulus colour measurement of LEDs by matrix transformation • Modern tristimulus colorimeters have four input channels. One can • just add the signals of the xs and xl channels (no matrixing) • Use the four channels for improving accuracy • Add a fifth channel • Optimization was performed for the LEDs shown previously
LED colour characteristics • LEDs are narrow band emitters • bandwidth approx. 10 nm – 30 nm • Blue … Green: InGaN • Yellow … Red: AlInGaP • Both the absolute intensity and the wavelength of the emission maximum is temperature dependent • temperature dependence is composition dependent • largest changes with Red LEDs • Unusual spatial light characteristics • Solution of measurement problems caused by • Spectral mismatch: spectrometric measurement • Spectral mismatch: tristimulus colorimetry • Temperature dependence • Geometric misalignment
Standard LED • Temperature and current stabilized LED for luminous flux measurement
Standard LED • Temperature and current stabilized LED for ALI measurement
LED luminance standard • TechnoTeam LED based Peltier-cooled luminance standard
LED colour characteristics • LEDs are narrow band emitters • bandwidth approx. 10 nm – 30 nm • Blue … Green: InGaN • Yellow … Red: AlInGaP • Both the absolute intensity and the wavelength of the emission maximum is temperature dependent • temperature dependence is composition dependent • largest changes with Red LEDs • Unusual spatial light characteristics • Solution of measurement problems caused by • Spectral mismatch: spectrometric measurement • Spectral mismatch: tristimulus colorimetry • Temperature dependence • Geometric misalignment
ALI measurement • Input clamp of ALI tube • Clamp for 5 mm LED
Reference LED Standard LED Detector with different filters DUT LED Flux and thermal measurement • TeraLED complex colorimetric and thermal measuring system • Radiometric • Photometric • Colorimetric • Thermal • measurements Calibration S, YREF-W Measurement of YT and YREF-T ΦT
Characteristics of Standard LEDs: Stabilization during the first two minutes
Summary • Based on visual observations a new LMS cone fundamental based colorimetry is recommended • Careful tristimulus colorimetry can be more accurate than low cost spectrometric techniques • Use temperature stabilized LEDs of similar colour as the test samples • Use good alignment for the LEDs in ALI measurements
Summary • Current sate of the art in user’s laboratory: • Spectral mismatch uncertainty: 1 – 2 % • Geometric alignment uncertainty: <+/-0,002% • Temperature dependence: +/- 0,2 %