Imaging Sequences part I

Imaging Sequences part I. Gradient Echo Spin Echo Fast Spin Echo Inversion Recovery. Goals of Imaging Sequences. generate an RF signal perpendicular to  0 generate tissue contrast minimize artifacts. z.  0. y. x. Measuring the MR Signal. RF signal from precessing protons.

wendell
Download Presentation

Imaging Sequences part I

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Imaging Sequencespart I • Gradient Echo • Spin Echo • Fast Spin Echo • Inversion Recovery

  2. Goals of Imaging Sequences • generate an RF signal perpendicular to 0 • generate tissue contrast • minimize artifacts

  3. z 0 y x Measuring the MR Signal RF signal from precessing protons RF antenna

  4. Gradient Echo • simplest sequence • alpha flip gradient-recalled echo • 3 parameters • TR • TE • flip angle • reduced SAR • artifact prone

  5. Gradient Echo dephase gradient rephase signal  RF pulse FID gradient recalled echo

  6. z z 0 RF y x y x Partial Flip 0 ML M MXY t=t0 t=t0+ MXY =M sin() ML=M cos()

  7. y y x x z z Dephasing in the xy-planeview from the top Mxy dephase Mxy phase coherency phase dispersion

  8. y y Mxy x x z Mxy z phase coherency minus t2* decay phase dispersion Rephasing in the xy-planeview from the top rephase

  9. z 0 y x MR Signal During Rephasing RF signal “echo” RF antenna

  10. T2* decay • occurs between the dephasing and the rephasing gradients • rephasing incompletely recovers the signal • signal loss is greater with longer TEs • decay generates image contrast

  11. T2* decay • T2* decay is always faster than T2 decay • gradient echo imaging cannot recover signal losses from • magnetic field inhomogeneity • magnetic susceptibility • water-fat incoherence

  12. T2 and T2* Relaxation • T2 is the spin-spin relaxation time • T2M is the contribution to relaxation induced by inhomogeneities of the main magnet (predominant factor) • T2MS is the contribution to relaxation induced by magnetic susceptibility in the object

  13. T2 and T2* Relaxation • T2* relaxation influences contrast in gradient echo imaging • T2 relaxation influences contrast in spin echo imaging

  14. Gradient Echopulse timing  RF slice phase readout echo signal TE

  15. Gradient Echoadvantages • faster imaging • can use shorter TR and shorter TEs than SE • low flip angle deposits less energy • more slices per TR than SE • decreases SAR • compatible with 3D acquisitions

  16. Gradient Echodisadvantages • difficult to generate good T2 weighting • magnetic field inhomogeneities cause signal loss • worse with increasing TE times • susceptibility effects • dephasing of water and fat protons

  17. Gradient Echochanging TE TE 9 FA 30 TE 30 FA 30 susceptibility effect T2* weighting

  18. Gradient Echomagnetic susceptibility post-surgical change “blooming” artifact

  19. Gradient Echoin-phase / opposed-phase TE 13.42 TE 15.66 in-phase opposed-phase

  20. Water/Fat Dephasing • MR signal is a composite of fat and water in the imaging voxel • water and fat resonate at slightly different frequencies • cyclic variation in relative phase of fat and water resonance results in signal variations dependent on TE times

  21. In-Phase / Opposed-Phase TE Times (msec)

  22. Gradient Echo • image contrast depends on sequence • conventional GR scan • aka GRASS, FAST • decreased FA causes less T1 weighting • increased TE causes more T2* weighting

  23. Conventional GRTE 20, FA 15

  24. Gradient Echo • Spoiled GR • aka SPGR, RF-FAST • spoiling destroys accumulated transverse coherence • maximizes T1 contrast

  25. Gradient Echo • Contrast enhanced GR • aka SSFP, CE-FAST • infrequently used because of poor S/N • generates heavily T2* weighted images

  26. Gradient Echo • other varieties • MTC • T2 - like weighting • IR prepped • 180 preparatory pulse • DE (driven equilibrium) prepped • 90-180-90 preparatory pulses • T2 contrast

  27. MTC GRTE 13, FA 50

  28. Spin Echo • widely used sequence • 90-180-echo • 2 parameters • TR • TE • generates T1, PD, and T2 weighted images • minimizes artifacts

  29. Spin Echo gradient frequency encode readout  RF pulse  RF pulse signal FID spin echo

  30. Gradient versus Spin Echo

  31. z z 0 RF y x y x 900 Flip 0 Before ML=M MXY=0 After ML=0 MXY=M t=t0 t=t0+

  32. y y x x z z Dephasing in the xy-planeview from the top Dephasing begins immediately after the 900RF pulse. Mxy Mxy phase coherency phase dispersion 900 RF t=0 t=TE/2

  33. y y Mxy x x z z Mxy phase coherency minus t2 decay phase dispersion t=TE/2 t=TE Rephasing in the xy-planeview from the top 1800 RF

  34. z z z z y x y x y x y x 1800 Flip dephased rephased 1800 RF 900 RF t=0 t=TE/2 t=TE

  35. Spin Echopulse timing   RF slice phase readout echo signal TE

  36. WNMR Race 900 RF t=0

  37. WNMR Race

  38. WNMR Race 1800 RF t=TE/2

  39. WNMR Race t=TE

  40. Effects of the 1800 Pulse • eliminates signal loss due to field inhomogeneities • eliminates signal loss due to susceptibility effects • eliminates signal loss due to water/fat dephasing • all signal decay is caused by T2 relaxation only

  41. Spin Echoadvantages • high signal to noise • least artifact prone sequence • contrast mechanisms easier to understand

  42. Spin Echodisadvantages • higher SAR than gradient echo because of 900 and 1800 RF pulses • long TR times are incompatible with 3D acquisitions

  43. Spin Echo Contrast • T1 weighted • short TR (450-850) • short TE (10-30) • T2 weighted • long TR (2000 +) • long TE (> 60) • PD weighted • long TR, short TE

  44. Spin Echo Contrast • T1 weighted - T1 relaxation predominates • Short TE minimizes differences in T2 relaxation • Short TR maximizes differences in T1 relaxation • T2 weighted - T2 relaxation predominates • Long TE maximizes differences in T2 relaxation • Long TR minimizes differences in T1 relaxation

  45. Spin Echo Contrast T1 weighted T2 weighted

  46. Spin Echo Contrast PD weighted T2 weighted

  47. Summary • Detection of the MR signal only occurs in the transverse plane • Gradient echo • Alpha degree pulse, dephase-rephase-echo • Contrast (T1/T2/T2*) depends on sequence type • Spin echo • 90 degree pulse, dephase, 180 degree pulse, rephase-echo • T1 weighted: short TR, short TE • PD weighted: long TR, short TE • T2 weighted: long TR, long TE