G. Mettivier , M.C. Montesi, P. Russo

Design of a compact gamma camera with semiconductor hybrid pixel detectors: imaging tests with a pinhole collimator. G. Mettivier , M.C. Montesi, P. Russo. Università di Napoli “Federico II” and INFN Napoli, Italy. Scope of the project. Design and build a compact gamma camera .

willa
Download Presentation

G. Mettivier , M.C. Montesi, P. Russo

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Design of a compact gamma camera with semiconductor hybrid pixel detectors: imaging tests with a pinhole collimator G. Mettivier, M.C. Montesi, P. Russo Università di Napoli “Federico II” and INFN Napoli, Italy 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  2. Scope of the project • Design and build a compact gamma camera • for Tc-99m g -ray tracer (140 keV) • for sentinel lymph node (SLN) radio-guided surgery • using the medipix series of hybrid pixel detectors + collimator Design specifications of final system • CdTe pixel detector (1 mm thick) • Medipix2 readout chip (256x256 pixels, 55 mm pitch) • Spatial resolution 1-2 mm 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  3. 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  4. Scope of the present work To perform preliminary evaluations of the spatial resolution attainable with a pinhole tungsten collimator with the present imaging system • Medipix1 chip (64x64 pixels, 170 mm pitch) • 300 mm thick Silicon detector (low detection efficiency) 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  5. 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  6. Spatial resolution Point-like g source 0.6 mm FWHM at 46 mm Geometric efficiency 2x10-5 Collimator Field Of View 11 mm Si Pixel detector Medipix1 chip Chipboard or PGA Previous work Multi-hole collimator Setup d D 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  7. h b Si pixel detector Medipix1 chip This work Experimental Setup Point-like g source Pinhole Collimator PGA chip carrier 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  8. 0.35 mm h 7 mm b Experimental Setup Pinhole aperture a = 90 deg Pinhole minimum diameter d = 0.35 mm (de = 0.57 mm) Total (geometric+penetrative) on-axis efficiency S = de2 /16h2 de = effective diameter Total FWHM spatial resolution Rs2[ (h+b)de/b ]2 + (h/b)2 Ri2 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  9. Results h = 16.5 mm h = 26.5 mm Point-like Co 57 gamma source (122 keV) h = 36.5 mm h = 46.5 mm Ethr = 15 keV Dt = 15 h b = 7.5 mm 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  10. Results 0.2 cps/kBq 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  11. Results D = 0.35 mm 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  12. Conclusions Tested a prototype compact gamma camera Imaging performance at 10 mm source-collimator distance system sensitivity = 0.2 cps/kBq FOV = 20 mm Spatial resolution = 1 mm 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  13. Future Developments Use of the Medipix2 readout ASIC (256x256 pixel, 55 micron pitch) Bump-bonded to a 1 mm thick CdTe pixel With respect to the present Medipix1 + Silicon detector: - Higher detection efficiency at 140 keV (30%) - Higher sensitivity (in the order of 1 cps/kBq) - Equal or better spatial resolution (1-2 mm) - Detection threshold settable up to about 100 keV Comparable with other prototype using arrays of scintillators Spatial resolution = 3.5 mm, FOV = 9 mm, Sensitivity = 10 cps/kBq (L. R. Mac Donald et al.) 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002

  14. Acknowlegement This work has been done in the framework of the activities of the Medipix2 collaboration (www.cern.ch/MEDIPIX) which comprises 16 European research institutions. The Medipix2 chip has been designed by CERN (Geneva, Switzerland). The MUROS2 interface has been designed by NIKHEF (Amsterdam, The Netherlands). The Medisoft 4 software system has been written by Universita' di Napoli Federico II and INFN (Napoli, Italy). 4th International Workshop on Radiation Imaging Detectors, Amsterdam, The Netherlands 8-12 September 2002