1 / 28

CAS – total LWC comparison

CAS – total LWC comparison. warm clouds. Assumptions. Instruments time synchronization info isn't available for this period. Some of the instruments were not synchronized during this early stage of the campaign.

wilmet
Download Presentation

CAS – total LWC comparison

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. CAS – total LWC comparison warm clouds

  2. Assumptions • Instruments time synchronization info isn't available for this period. • Some of the instruments were not synchronized during this early stage of the campaign. • I assume the positive mixing ratio coincide with onset of LWC measurements in CAS.

  3. Limitations • Horizontal sampling line – low penetration efficiency for big droplets and drops • High concentrations and screening of big droplets by small droplets or particles cause lower LWC content. • Presented CAS LWC data for droplets >3m.

  4. Equations (+10C) Liquid water mixing ratio measured by CAS: r≡Density of dry air Liquid Water Mixing Ratio +10C: Assumption: changing chamber pressure at saturation, flux from the walls:

  5. Runs details

  6. Conditions 1243.04 P[mb],RH[%],T[K] seconds

  7. 1243.04 [Kg/Kg]

  8. 1243.06

  9. 1243.06

  10. 1243.08

  11. 1243.08

  12. 1245.01

  13. 1245.01

  14. 1245.03

  15. 1245.03

  16. 1247.01

  17. 1247.01

  18. 1247.05

  19. 1247.05

  20. 1247.07

  21. 1247.07

  22. 1248.13

  23. 1248.13

  24. 1249.03

  25. 1249.03

  26. 1249.09

  27. 1249.09

  28. conclusions • Possible reasons for discrepancies in LWC measurements: -CAS limitation that were mentioned -wrong measurement of Pressure change -wrong measurement of MBW_dew -wall losses • Should be compared to WELAS LWC