Cardiac implantable electronic devices (CIED) : Pacemakers, Implantable CARDIoverter defibrillators (ICD). Moderator : Dr Manju Mani. www.anaesthesia.co.in anaesthesia.co.in@gmail.com. Brief History.
E N D
Cardiac implantable electronic devices (CIED) :Pacemakers,Implantable CARDIoverter defibrillators(ICD) Moderator : Dr Manju Mani www.anaesthesia.co.in anaesthesia.co.in@gmail.com
Brief History • First totally implantable pacemaker : into the chest of 43 yr old Arne Larsson by Dr Ake Senning in Stockholm on Oct 8 ,1958. • Introduction of external defibrillators in 1962 • First internally implanted defibrillator in 1980
pacemaker Device that provides electrical stimulation to cause cardiac contraction when intrinsic cardiac electrical activity is slow or absent
Pacemaker Functions • Stimulate cardiac depolarization • Sense intrinsic cardiac function • Respond to increased metabolic demand by providing rate responsive pacing
Indications for Pacemaker • Aquired AV block - 30 AV block • Symptomatic bradycardia • Asystole >3 sec or escape rhythm <40bpm • Post op AV block not expected to resolve • Neuromuscular disease with AV block - 20 AV block • Permanent or intermittent symptomatic bradycardia
2. After MI • Symptomatic 20 AVB or 30 AVB • Infranodal AV block with LBBB 3. Bifascicular or Trifascicular block • intermittent complete heart block with symptoms • 2 AV block • Bundle branch block
4. Sinus Node Dysfunction - with symptoms as a result of long term drug therapy - symptomatic chronotropic incompetence 5 . Hypertensive carotid sinus & neurocardiac symptoms - recurrent syncope associated with carotid sinus stimulation - Asystole of > 3 sec duration in absense of any medication
Pacemaker Components • Pulse Generator • Electronic Circuitry • Lead System
Pulse Generator • Subcutaneous or submuscular • Lithium battery • 4-10 years lifespan • long life and gradual decrease in power sudden pulse generator failure is an unlikely cause of pacemaker malfunction
Electronic Circuitry • Sensing circuit • Timing circuit • Output circuit
Bipolar Lead has both negative, (Cathode) distal and positive, (Anode) proximal electrodes Separated by 1 cm Larger diameter: more prone to fracture Compatible with ICD Unipolar Negative (Cathode) electrode in contact with heart Positive (Anode) electrode: metal casing of pulse generator Prone to oversensing Not compatible with ICD Lead System
Bipolar • current travels only a short distance between electrodes • small pacing spike: <5mm + Anode - Cathode
Unipolar • current travels a longer distance between electrodes • larger pacing spike: >20mm + Anode - Cathode
Paced Rhythm Recognition VVI / 60
I II III IV V Programmable Antitachy Chamber Chamber Response Paced Sensed to Sensing Functions/Rate Function(s) Modulation Pacemaker Code DEVELOPED AS JOINT PROJECT BY NORTH AMERICAN SOCIETY OF PACING & ELECTROPHYSIOLOGY (NASPE) AND BRITISH PACING AND ELECTROPHYSIOLOGY GROUP (BPEG) - REVISED 2002 P: Simple programmable V: Ventricle V: Ventricle T: Triggered P: Pace M: Multi- programmable A: Atrium A: Atrium I: Inhibited S: Shock D: Dual (A+V) D: Dual (T+I) D: Dual (P+S) D: Dual (A+V) C: Communicating O: None O: None O: None O: None R: Rate modulating S: Single (A or V) O: None
Common Pacemakers • VVI • Ventricular Pacing : Ventricular sensing; intrinsic QRS Inhibits pacer discharge • VVIR • As above + has biosensor to provide Rate-responsiveness • DDD • Paces + Senses both atrium + ventricle, intrinsic cardiac activity inhibits pacer d/c, no activity: trigger d/c • DDDR • As above but adds rate responsiveness to allow for exercise
Rate Responsive Pacing • When the need for oxygenated blood increases, the pacemaker ensures that the heart rate increases to provide additional cardiac output Adjusting Heart Rate to Activity Normal Heart Rate Rate Responsive Pacing Fixed-Rate Pacing Daily Activities
Determining type of pacemaker • Wallet card: 5 letter code • CXR: code visible • Single lead in ventricle: VVI • Separate leads DDD or DVI
Single Chamber • VVI - lead lies in right ventricle • Independent of atrial activity • Use in AV conduction disease
Dual Chamber • Typically in pts with nonfibrillating atria and intact AV conduction
Pacemaker Interventions • Magnet application • No universal function of magnet • Model-specific magnet, some activate reed switch asynchronous pacing at pre-set rate • Interrogation / Programming • Model-specific pacemaker programmer can non-invasively obtain data on function and reset parameters
Complications of Pacemaker Implantation • Venous access • Infection • Thrombophelbitis • Pacemaker Syndrome
Venous Access • Bleeding • Pneumo / haemothorax • Air embolism
Infection • 2% for wound and ‘pocket’ infection • 1% for bacteremia with sepsis • S. aureus and S. epidermidis Rx : If bacteremic: start antibiotics, remove system, new system to be placed
Thrombophlebitis • Incidence 30-50% • b/c of collateralization only 0.5-3.5% devp symptoms • Swelling, pain, venous engorgement • Rx Heparin, lifetime warfarin
Pacemaker Syndrome • Presents with worsening of original symp post-implant of single chamber pacer - hypotension, syncope,vertigo, exercise intolerance etc • AV asynchrony retrograde VA conduction atrial contraction against closed MV + TV jugular venous distention + atrial dilation sx of CHF • Rx : dual chamber pacer
4 broad categories • Failure to Output • Failure to Capture • Inappropriate sensing: under or over • Inappropriate pacemaker rate
Failure to Output absence of pacemaker spikes despite indication to pace • dead battery • fracture of pacemaker lead • disconnection of lead from pulse generator unit • Oversensing • Cross-talk: atrial output sensed by vent lead
Failure to capture spikes not followed by a stimulus-induced complex • change in endocardium: ischemia, infarction, hyperkalemia, class III antiarrhythmics (amiodarone, bertylium)
Inappropriate sensing: Undersensing Pacemaker incorrectly misses an intrinsic deoplarization paces despite intrinsic activity • Appearance of pacemaker spikes occurring earlier than the programmed rate: “overpacing” • may or may not be followed by paced complex: depends on timing with respect to refractory period • AMI, progressive fibrosis, lead displacement, fracture, poor contact with endocardium
Undersensing • Pacemaker does not “see” the intrinsic beat, and therefore does not respond appropriately Scheduled pace delivered Intrinsic beat not sensed VVI / 60
Inappropriate sensing: Oversensing Detection of electrical activity not of cardiac origin inhibition of pacing activity • “underpacing” • pectoralis major: myopotentials oversensed • Electrocautery • MRI
VVI / 60 Oversensing • An electrical signal other than the intended P or R wave is detected ...though no activity is present Marker channel shows intrinsic activity...
Inappropriate Pacemaker Rate • Rare reentrant tachycardia seen w/ dual chamber pacers • Premature atrial or vent contraction sensed by atrial lead triggers vent contraction retrograde VA conduction sensed by atrial lead triggers vent contraction etc etc etc • Tx: Magnet application: fixed rate, terminates tachyarrthymia, • reprogram to decrease atrial sensing
Causes of Pacemaker Malfunction • Circuitry or power source of pulse generator • Pacemaker leads • Interface between pacing electrode and myocardium • Environmental factors interfering with normal function
Pulse Generator • Loose connections • Similar to lead fracture • Intermittent failure to sense or pace • Migration • Dissects along pectoral fascial plane • Failure to pace • Twiddlers syndrome • Manipulation lead dislodgement
Leads • Dislodgement or fracture (anytime) • Incidence 2-3% • Failure to sense or pace • Diagnosed with CXR, lead impedance • Insulation breaks • Current leaks failure to capture • Diagnosed with measuring lead impedance (low)
Cardiac Perforation • Early or late • Usually well tolerated • Asymptomatic inc’d pacing threshold, hiccups • Diagnosis : hiccups, pericardial friction rub CXR, Echo
Environmental Factors Interfering with Sensing • MRI • Electrocautery • Arc welding • Lithotripsy • Microwaves • Mypotentials from muscle
Pseudomalfunction: Hysteresis • Allows a lower rate between sensed events to occur; paced rate is higher Hysteresis Rate 50 ppm Lower Rate 70 ppm
Anaesthesia for insertion MAC To provide comfort To control dysrhythmias To check for proper function/capture Have external pacer & Atropine ready Continuous ECG and peripheral pulse monitoring
Temporary Pacing Methods • Invasive (Direct) cardiac Pacing • Epicardial • Stainless steel Teflon coated wires. • Endocardial • Flow directed balloon electrodes • Catheter with guidewire • With PA catheter- Side port for ventricular pacing • Non Invasive (Indirect) • Transcutaneous Pacing • Combined pacing, cardioversion and defibrillation with ECG monitoring in a single unit • Instituted quickly • Safely by minimally trained person
Cont.. • Disadvantages of transcutaneous pacing • Inability to obtain reliable capture in • Emphysema • Pneumothorax • Morbid obesity • Difficulty with lead placement • Surgical Field • Patient position • Failure of TCP to preserve AV synchrony • For patients with poor ventricular diastolic function