1 / 71

Chromatography

Chromatography. Prof. Zoltán Juvancz D.Sc. The first chromatography by Tswett. Aim of chromatography. The compounds of interest has to be separated from matrix components. The signal to noise ratio must be as high as possible. The time consumption of analysis must be short.

yachi
Download Presentation

Chromatography

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Chromatography Prof. Zoltán Juvancz D.Sc.

  2. The first chromatography by Tswett

  3. Aim of chromatography • The compounds of interest has to be separated from matrix components. • The signal to noise ratio must be as high as possible. • The time consumption of analysis must be short.

  4. Gas chromatograms of 104 volatile organic compounds (VOC)

  5. If the peaks are narrow, more compounds can be separated

  6. Definitions • Chromatography is a separation method. (Chromatography is not an identification method like NMR, IR, MS) Chromatography consist of two phases: mobile and stationary phase. Mobile phase is forced along the column from injection to detectoras a flowing media. Stationary phase is anchored to the column wall or to the particles, which are packed into the column.

  7. Chromatographic process • The injected sample is dragged by mobile phase along the column. • The components of the sample distribute between the stationary phase and mobile phase. • If X compound has bigger affinity to stationary phase than affinity of Y compound to stationary phase, the X compound elutes later than Y from column. • The sharp injected peaks become broader and broader during their run.

  8. Schematic view of aGC

  9. Cross section of open tubular (capillary) General parameters Column length: 5-100 m Column diameters: 0.1-0.5 mm Film thickness: 0.15 – 5 µm Fused silica wall has no metal content. The metals can cause destruction of sample and stationary phase with their catalytic effects.

  10. Advantages of chromatography • Exact quantitative analysis is done even from trace compounds. • Disturbing effects of matrix compound can be eliminate. • Small material consumption • The quantization has a broad linearity range. • Analyses of several compound can be done during one run. • Chromatography is a fast analysis method. • On-line coupling are routinely solved to compound identification methods, • Well establishes instrumentation with high level automation is commercially available.

  11. Trace analysis

  12. Single cell analysis

  13. Determination of trace compound from cpmplex matrix

  14. Analyses of chlorinated pesticides with GC x GC system

  15. Determination 106 compounds during one run

  16. Determination of trace impurity of a compounds Determination of compounds having different magnitudes in their volume.

  17. Fast analysis

  18. On-line GC/MS coupling with selected ion monitoring

  19. Chromatography is highly automitized Auotomatic processes allow whole day work without human inspection. The automated processes have high reproducibility and accuracy.

  20. On line HPLC/MS coupling

  21. Determination of origin of extasy tablets

  22. Recognition of adulteration of bergamot oil

  23. Elution profile of a chromatogram Intensity Time

  24. Chromatographic peak An ideal chromatographic has Gaus shape Qualitative measure: retention time (tR) Quantitative measure: Peak area (A)

  25. Chromatographic peak • A certain compound shows same retention times under same conditions (column, flow, temperature etc.) independently from its injected amount (in the linear range of Langmuir isotherm). The identity of the compounds is partly based their retention time. (Qualitative parameter) • Peak area is linear function of the quantity of certain compound (Quantitative parameter).

  26. Chromatographic peak An ideal chromatographic has Gaus shape Qualitative measure: retention time (tr) Quantitative measure: Peak area (A)

  27. Chromatographic peak • The given compound shows same retention times under same conditions (column, flow, temperature etc.) independently from its injected amount (in the linear range of Langmuir isotherm). The identity of the compounds is partly based their retention time. • Peak area is linear function of the quantity of certain compound.

  28. Retention time: tR (qualitative measure) Hold up time: tm (The time, what the compounds spend in mobile phase) Peak with in half high: wh Peak area (A) (quantitative measure) Theoretical plate high: N (efficiency, sharpness ) N = 5,54 (tR/wh)2 Capacity ratio: k (Strength of stationary phase) k = tR-tm/tm Chromatographic expressions

  29. Theoretical plate height The peaks become broader and broader during the chromatography The later eluting peak are broader than early ones. The theoretical plate height is same for all peaks. The theoretical plate height characterize the chromatographic system. The capacity ratio shows the interaction of compound of interests (distribution between the stationary and mobile phases.

  30. Chromatographic expressions Selectivity: (Measure of interaction ratio between two compound and stationary phase) • = tR2’/ tR1’ Resolution: Rs (Measure the separation of two peaks) Rs = 1,177 (tR1-tR2)/(Wh1+Wh2) Rs : 1.5 baseline resolution

  31. Increasing of capacity ratio with increasing of volume of stationary phase

  32. Changing of capacity ratio with decreasing solvent strength of mobile phase Decreasing solvent strength results in longer retention times.

  33. Distribution of the compounds between two phases • Kd = Cm/Cst = p/q • E = KdV/(1 +KdV) Where • Kd distribution constant • C m concentration of a compound in mobile phase • Cst concentration of a compound in stationaryphase • E extraction ratio • V phase ratio • Two compounds can be separated if their distribution constants are not equal (Kdx≠ Kdy).

  34. Distribution of a compound between the mobile and stationary phases

  35. Material distribution between the stationary phase and mobile phase K= p/q

  36. Band broadening in mobile phase

  37. water flow base Stones in river Light stone Heavy stone Simulation of separation

  38. Material distribution between the stationary phase and mobile phase

  39. Separation processes Number of the equilibrium Intensity of signal Place of materials from injection point

  40. If the peaks are narrow, more compounds can be separated

  41. The slow mass transfer between the phases causes peak broadening Resistance of mass transfer processes cause that the material zone are retainedin stationary phase from the material zone in mobile phase.

  42. Band broadening in open tube The band of sample is getting broader caused by slow diffusion from the middle of mobile phase to the stationary phase.

  43. Band broadening caused by uneven flow The flow velocity is not uniform across the column.

  44. Band broadening caused by badly washed holes The compounds come in and out from the holes slowly. The diffusion processes are much slower than the flow of the mobile phase.

  45. The ideal flow of mobile phase is result of compromises U HETP =A +B/u + Cu A=Eddy Diffusion, B =Molecular Diffusion, C =Resistance to mass transfer, U =Linear velocity of mobile phase (cm/s) HETP: high equivalent theoretical plate, column HETP: N/L (L: length of)

  46. Resolution vs. efficiency, capacity ratio, selectivity P. Sandra JHRC 12 (1989) 82.

  47. Resolution-efficiency- selectivity

  48. Ways how to improve the resolution • Increase the efficiency • Longer columns • Smaller diameter of column or smaller diameter of particles • Thinner stationary phase • Optimated flow of mobile phase • Increase the capacity ratio (3-10) • Bigger amount of stationary phase • Weaker mobile phase (HPLC) • Lower analysis temperature (GC) • Increased selectivity • More selective phases • Lower analysis temperature • Derivatization

  49. GC Chirasil-Dex 10 m Higher efficiency with smaller diameter column

  50. Higher column length Longer column give better resolution, but longer analysis times.

More Related