David Joy . Dr. David C. Joy Distinguished Professor Materials Science and Engineering University of Tennessee Knoxville, TN D.Phil., University of Oxford (UK): A Study of Magnetic Domains in the SEM Hitachi payroll. A Clean Machine.
E N D
David Joy • Dr. David C. Joy • Distinguished Professor • Materials Science and EngineeringUniversity of Tennessee Knoxville, TN • D.Phil., University of Oxford (UK): • A Study of Magnetic Domains in the SEM • Hitachi payroll
A Clean Machine • The FEGSEM cannot operate except in a clean, ultra-high vacuum. (Numerous caveats for low vacuum and ESEM operation.) This requires items such as: • Scroll Pumps • Turbo Molecular Pumps • Ion Pumps • as well as associated gauges • and even with the cleanest of vacuums there is still The Dark Side of SEM that must be faced….
Terminology • “Low vacuum” = High pressure • “High Vacuum” = Low pressure
Common Vacuum Units • There are many varied units that are used to specify pressures • The Torr, the Bar and the Pascal are in common use... • .. but the Pascal is the SI recommended unit for pressure and so is the best choice for documentation • 1 Atmospheric pressure is 760 mm Hg = 1 Bar = 105 Pa • 1 Torr = 1 mm Hg • 1 Torr = 1/760 of an atmosphere = 132 Pa • 1 milliTorr = 0.13Pa = 1 μmHg • 1mbar = 1/1000 Atm = 0.76 Torr = 100Pa • 1 Pa = 7.6 milliTorr = 7.6 μmHg
Pressure: Units of Measure • Pressure exerted by a column of fluid: • P ≡ F/A = mg/A = ghA/A = gh h • 1 Atm (mean sea level) = 760 Torr = 1013 mBar = 1.01x105 Pa = 101.3 kPa = 14.7 psi = 34 ft. water • Average atmospheric pressure in SLC is about 635 Torr, 12.3 psi, 28.4 ft water…
“Kinds of Pressure” • Gauge Pressure: measured with respect to ambient. • Absolute pressure: measured with respect to vacuum • Car tires, basketballs, boilers, LN2 tanks, JFB/MEB compressed air supply… • Vacuum systems, cathode ray tubes, light bulbs, barometers
Mean Free Path in Gases With sufficient accuracy for approximate calculations we may take: λ = 7 x 10-3/p mbar-cm λ = 5 x 10-3/p Torr-cm λ = 5/p μmHg-cm
Qualitative Vacuum Ranges Low vacuum (SEC) 760 to 1 Torr Medium vacuum (SEC) 1 to 10-3 Torr High vacuum (Chamber) 10-3 to 10-6 Torr Very high (Column) 10-6 to 10-9 Torr Ultra-high (Gun) 10-9 and lower laminar molecular FEGSEMs contain examples of each vacuum level
Vacuum pumps • For each of the vacuum ranges identified earlier there is one or more type of pump that is best • Pumps are always used in combination - one pump is used to start the next • The sequencing of the pump down is crucial and so this is done under computer control
Scroll Pumps • Scroll pumps are the foundation of clean vacuum systems • They consist of two Archimedes’ screws machined into aluminum plates mounted so that the spirals interleave • One plate is held fixed while the other oscillates. Gas is trapped between the spirals and forced out to the exit port • Pumping speed is constant from Atmospheric pressure down to about 1000Pa and the ultimate pressure is about 10Pa • Scroll pumps are oil-free and require neither inlet nor outlet valves The world’s oldest pump technology – Archimedes’ screw
Alternatively…. • Roughing can also be carried out using a diaphragm pump • Oscillation of the diaphragm alternately pulls gas in one port and then expels it through the other. • Oil free pumps are clean but typically a factor of 3x slower, and 3x more expensive, than pumps containing oil – but worth the wait and expense
Rotary Vane Mechanical Pump • Robust • Inexpensive • Operates to ambient pressure • Single stage and two stage
Turbomolecular pump • Archimedian screw - runs at 20000+ rpm • Needs electronic protection / control for the bearings in case of loss of power • Produces a clean, oil-free, high vacuum down to 10-6 Pa (10-8 Torr) • Must be backed: scroll pump, diaphragm pump or rotary oil pump.
Turbo pump performance • Turbo pumps can start even at atmospheric pressure (although they labor) and they can go down to 10-8 T • It is best to pre-pump the system with a clean backing system • TMP do not pump all gases with same efficiency - large molecules are pumped faster than smaller molecules • 1 milliT = 0.13Pa = 1 μmHg Turbo pump performance
Ion Pumps • Ionized molecules spiral in magnetic field and get buried in Ti wall coating • A large number of these structures are run in parallel to improve the pumping speed • Diode pumps only handle gases that are easily ionized (no noble gases)
The triode pump • If noble or unusual gases are expected to be found in the SEM (nitrogen, helium, counter gases from a WDS system etc.) then a triode pump must be used. • The additional electrode then makes it possible to ionize these gases
Ion pump performance • “The” UHV pump - goes to 10-9 Pa (10-11 Torr) and below in a properly designed vacuum system • Requires no backing…more than a little misleading… in fact it works best in a sealed system. Entrainment pump! • The IP requires a periodic bake-out into rough pumped system to clean the buried gas from the pump. This is done during the gun bake procedure • Check for electrical instability by slapping the pump with an open hand. Instability indicates need for a bake Ion Pump Performance
Cryogenic pump • Cryo-pumps use liquid helium and activated charcoal absorbers to pump to 10-12 T • Very high pumping speeds • No vibration or magnetic fields • But they need periodic bake-outs into a rough pump to clean the absorbers • They are expensive to run unless used with a closed-circuit (Stirling engine) liquid He pump
Vacuum Gauges • Vacuum systems must be monitored constantly to ensure satisfactory performance, but manufacturers seem to be reluctant to provide gauges which allow this to be done • Many different types of gauges are available because each only covers a limited range of pressures • Never trust a gauge unless you can check it independently Range of gauge utility
Pirani gauge • The Pirani is a dedicated low vacuum gauge device • The resistance of the hot wire changes with the rate of heat loss (conduction) to the gas • The Wheatstone bridge then measures the change in resistance of the hot wire • Pirani’s are rugged and generally reliable and rarely need attention Schematic Circuit for a Pirani (hot wire) gauge
Pirani calibration • The calibration of a Pirani depends on thermal conductivity and so on the actual gas in the system • Beware when using a crystal spectrometer as gases leaking from the counter tubes will degrade the accuracy of the Pirani gauge Correction Curve for Pirani Gauges
Penning (Cold cathode) Gauge • A Penning gauge measures the ion current flowing from the cathode to the anode • The magnetic field increases sensitivity by making the ions spiral as they travel to cause secondary ionization • Beware - a Penning gauge reads zero current when the pressure is both very low and very high. The gauge must ‘strike’ to be operational • Check with a Pirani gauge if in doubt Penning gauges require routine cleaning and testing
Capacitance Manometer Gauge head on chamber Controller and digital read-out
Capacitance Manometer • A = Annular electrode • D = Disk electrode • S = Substrate • G = Getter (in vacuum space) • Differential capacitance between annulus and disk depends on pressure difference between Test Chamber and “Getter”. (Earlier reference to “unbacked getter pump”)
Ion gauges • Pressures lower than 10-5 Torr can be measured with ion gauges (which are miniature ion pumps) or (more usually) directly from the actual ion pump • Mass spectrometer gauges (residual gas analyzers) are a desirable extra. These can measure partial pressures of e.g helium (for leak testing) or of water vapor.
O-ring seals • O-rings (from the 1950s) made it possible to build demountable vacuum systems • The rings are now made of high tech polymers such as VITON • Two kinds are in common use... • Black/shiny - has filler. Low vacuum only. Lubricate with finger grease to prevent cracking • Brown/dull - high vacuum, and bakeable. Do not grease • Do not crush or cut the ring - ensure that it is in the groove designed for it
UHV metal to metal seals • First used in the 1960s • Knife edges on the flanges cut into OFHC (oxygen free high conductivity) copper rings about 5mm thick to make an impermeable metal to metal seal • Good down to pressures as low as 10-10 Pa • Bakeable, clean, long lasting and (with care) reusable • Expensive - an 8 inch gasket costs ~ $100 • Don’t touch !
Vacuum Hygiene • Always keep vacuum systems running 24/7/365 • Use LN2 cooled maze traps, and fore-line traps, to reduce backstreaming in older machines • Do not overpump the specimen exchange chamber (SEC) as this can result in backstreaming (unique to Hitachi FEG systems) • Keep your fingers away from samples and from the specimen chamber area - wear gloves • If column contamination occurs try nitrogen purging (laminar flow) over a weekend Maze trap fitted to a rotary pump (mfp discussion)
Cleaning samples • Do not use organic solvents as these are always contaminated, even the fresh ‘electronic grade’ material in brown glass bottles • Never, never, use squeeze or spray bottles as the TEFLON filler goes into solution • Use detergents instead e.g. Alconox ‘Detergent 8’ which are bio-degradeable and leaves no residue • Carbon Dioxide ‘snow’ cleaning -no residue and good solvent action but expensive to set up. www.co2clean.com CO2 “snow gun” for sample cleaning
Clean is not for ever ... • As soon as a specimen is prepared for observation it begins to get dirty again (CCW rule: “one monolayer/sec at 10-6 Torr) • Even storing the sample in a vacuum dessicator will not prevent the growth of bacterial or microbial surface contaminant films because the source of the problem is carried in by the specimen itself • Remedial action is required As prepared After one week
Plasma cleaning • Plasma cleaning is a rapid and easy way of removing the build-up of surface contaminants • Fast and non-destructive Same sample after plasma cleaning
The Dark Side of SEM • The interaction of electrons with solids results in a variety of interactions which give us uniquely valuable information about the sample • But these same interactions can also result in either temporary or permanent damage to the sample • Know your enemy!
Radiation Damage Ionization Displacement Heating Contamination Etching Results from vacuum problems Both are usually important Unwanted Beam Interactions Intrinsic to electron beam irradiation
Unwanted beam interactions • Electron beams have bad effects on organic, polymeric, and ionic materials • This is ‘radiolysis’ Effect of 0.01mC/cm on protein protoxin Shrinkage of ArF resist 1mC/cm2 500nm
Radiolysis is…. • Radiolysis is the breaking of bonds as the result of ionization by the electron. • Electrons are the most intense source of ionizing radiation available -the typical dose in an SEM is equivalent to standing 6 foot from a 10 megaton H-bomb Compare SEM to Sun and SPEAR
Radiolysis damage in Polymers • In polymers radiolysis produces swelling or shrinking in the material and the actual loss of the sample • Despite appearances this damage is not due to heating in the sample • The effect may be reduced by coating with metal or a thin carbon layer Courtesy Dale Newbury NIST
Dose does matter • A typical SEM dose for a photo-record is about 0.1 C/cm2 or 100 el/Å2 • Typically at 1 -10el/Å2 we see a loss of crystallinity at 10-100 el/Å2 mass loss and above 100 el/Å2 limiting mass loss Dose from a single photo scan
Is a high beam energy bad? • It is often said that low beam energies minimize or eliminates beam induced damage • From casual observation this statement may appear to be true, but physics and measurements show that the truth is just the opposite • And note - even a very low energy electron (1eV) has an equivalent temperature of 10,000oK, which is hotter than the surface of the sun
‘Mythbuster’ fact Range~size damage is maximum • All electrons damage • At low energies the damage is high but limited by the range • Damage is a maximum when range & feature sizes are similar • At higher energies damage falls - energy deposition occurs outside the feature Range<<size damage limited Range>>size damage limited Adapted from Egerton (2004)
Damage in semiconductors • e- beam damage of devices shifts the threshold voltage • damage is localized in gate oxides and is usually reversible • Damage depends on the beam energy…. and generally appears to get worse as the energy is increased
Thermal damage? • Not usually a serious problem as the energy deposited is quite small • For a typical material of medium density and thermal diffusivity the temperature rise with energy and beam dose is minimal Magnification5keV 15keV 30keV 400x0.1C/nA 0.24C/nA 0.56C/nA 4000x 0.15C/nA 0.34C/nA 0.79C/nA
Other beam induced damage • In addition to Radiolysis the beam can produce ‘knock-on’ damage • In this the incident electron strikes an atom head on and knocks it out of position generating vacancies, e.g. Frenkel defects • This requires a minimum beam energy before it can occur, the value varying with the atomic number of the sample • For Carbon (Z=6) the knock-on threshold energy is about 80keV, for Silicon (Z=14) the knock-on threshold is 220keV • Not currently a problem with SEMs
Contamination and Etching • Contamination is beam induced polymerization of the hydrocarbons present on the sample surface • Etching is the removal of surface layer by impact of ions (C + H2O 2- --> CO + H2 ) • Both phenomena are affected by surface charging and often occur together • Both are temperature dependent • Your microscope is not to blame!
Modern SEMs are very clean RGA of S4300 chamber just before the specimen is inserted H2O N hydrocarbon
but samples are not.. S4300 chamber vacuum just after sample insertion hydrocarbons from the SEC pump
Contamination and Etching Electrons break down the hydrocarbon film by radiolysis. The residue charges +ve and the field pulls in fresh material for radiolysis. If water vapor is present then negative drift to + ve charged regions and can etch that area away
Low magnification • At low magnifications the hydrocarbon film is polymerized into a thin sheet. • This will charge positive (and so look black in the SE image) but is not a serious problem • Minimize by pre-exposing the sample at the lowest possible magnification prior to examination Schematic of contamination build-up at low magnification scans
Black squares... • The black squares are visible evidence of the charging that occurs • Post facto in situ removal of contamination is possible using plasma sources in the chamber although the process is slow • Use plasma cleaning before observation for best results • Example shown is by courtesy of Dr. Bryan Tracy, Spansion Inc.
High magnification • At high magnification the contamination grows a cone which scatters the beam • Avoid spot mode - always keep the beam scanning the sample • Pre-pump samples before use • Keep your hands off the sample • Avoid the use of dirty solvents • Plasma clean before use if possible ~ 0.03m Ant-hill contamination