250 likes | 276 Views
Dr. G. Kalender presents Anaconda Stentgraft key features, modular design, indications, and innovative technology for complex iliac anatomy. Learn about implantation techniques and clinical trial results on 22nd June 2019 at Deineka Conference Room, DRK Kliniken Berlin.
E N D
EVAR Overview and stent-graft implantation in complex iliac anatomy Angiomed Workshop Deineka Conference Room 22TH JUNE 2019 Dr. G. KalenderDRK Kliniken Berlin
Anaconda Stentgraft • modular design • flexible • safe • repositionable • easilyprobedwithmagnetwire Dr. G. Kalender
Indications MorphologysuitableforEndovascularAneurysmRepair (EVAR) Proximal aortic neck lengthof≥15mm with non-significantcalcificationand/orthrombus Native proximal neck diameter17,5- 31 mm Anaconda™ ONE-LOK™ Adequateiliacorfemoralvesselaccess Native iliacarterydiametersof8,5- 21 mm Distal Fixation LengthOf ≥20mm Dr. G. Kalender
Modular concept wide range of main bodys and iliac legs durable ultrathin- polyester ring- stent design (nitinol) Dr. G. Kalender
bifurcated main body proximal ring contralateral radiopaque markers secondary ring valley hooks peek hooks leg docking radiopaque distal radiopaque marker contralateral cannulation flare with radiopaque markers Dr. G. Kalender
Iliac leg proximal radiopaque marker individual ring stents with radiopaque alignment markers distal radiopaque marker Dr. G. Kalender
Iliac leg – flared und tapered proximal radiopaque marker transition marker distal radiopaque marker Dr. G. Kalender
fast and easy cannulation patentedmagnetwirecannulation Dr. G. Kalender
Fully repositionable Dr. G. Kalender
Range of sizes • bifurcated main body • diameter 21,5 – 34 mm • length 65 mm • Iliac leg • diameter 10 – 23 mm • length 60 – 140 mm Dr. G. Kalender
Anaconda Stentgraft System NEW ONE-LOK™ BODY: Additional ring stents: maximizethelumen und bloodflow large lumenforipsilateralcannulation standarddiameter in thedockingzone indicated to treat infra-renal neck angulations > 90° Dr. G. Kalender
12 Proactive Sealing System Excellent sealing function - dual, proximal ring stent design with hooks - security against stent graft migration Dr. G. Kalender
FDA Clinical Trial Phase II One Year Results of the Vascutek Anaconda™. ONE-LOK EVAR trial. Christopher Kwolek, MD. Randy Moore, MD. On behalf of the Anaconda™ Investigators 23 hospitals (18 USA, 5 Canada), 195 patients, >15 mm neck of aneurysm, <60° angulation 98 % technical success, no conversion 0% mortality, 100 % open iliac leg 98% no endoleak Typ I oder III Typ II endoleak: 32 % (30 Tagen), 26% (1Jahr) summary: Anaconda- stentgraft- safe and effektive after 12 months Dr. G. Kalender
Fenestrated Anaconda™ • Body sectionwithoutstents: • maximise the available area for fenestrations • optimal flexibility for varying patient anatomies • fullyrepositionablesystemfor optimal positioning • nodeformationoffenestrationbystent material Dr. G. Kalender
Fenestrated Anaconda™ patented magnet system: significantly reduce bifurcate body cannulation time dual proximal ring stent design with hooks provides excellent sealing function multiple independent ring stent design provides maximum flexibility for varying patient anatomies minimises the potential for kinking Dr. G. Kalender
Fenestrated Anaconda™ > 3000 implantations > 8000 targetvessels since 06/2010 VacutekTerumo News, fenestratedanaconda, Jahrgang XVIV 2015 Dr. G. Kalender
Custom Made LegExample Concepts Dr. G. Kalender
Design Possibilitys Sealingring, Hooks, Fenestrations… Dr. G. Kalender
Access Possibilitys Dr. G. Kalender
POST-OP Pictures Dr. G. Kalender
Device Customisations Pleating • Reduces fabric in region of aortic narrowing Dr. G. Kalender
Additional Fenestration Support Tapered Graft • Helps to alleviate excess fabric overlapping with fenestrations • Reduces the fabric diameter in cases of more extreme narrowing Dr. G. Kalender
Flared Graft Extra-long Body • Increases the fabric diameter in cases of more extreme widening • 90-100mm long. Required joining ring to attach two pieces of fabric together and the contralateral flare is removed. Dr. G. Kalender
Legs Lateral • Useful when close to the bifurcation of an existing device or aorta Dr. G. Kalender