An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Accurate gamma-ray spectrometry of environmental samples: a challenge O. Sima - Bucharest University D. Arnold - PTB Braunschweig C. Dovlete - ERL Bucharest

  2. Accurate gamma-ray spectrometry of environmental samples: a challenge • Introduction • Problems in efficiency calibration of the spectrometer • Coincidence summing effects • Matrix effects • Geometry effects • GESPECOR • Summary and conclusions

  3. Introduction • Assessment of radioactivity of environmental samples: • carefully designed sampling procedures • appropriate sample preparation • accurate sample measurement • rigorous analysis of the results • Modern requirements and conditions: • low detection limits • accurate evaluation of uncertainty • high number of samples, various types, matrices, available quantities • high efficiency detectors available

  4. Problems in efficiency calibration • Low level activity + low detection limit: • high efficiency measurement conditions • volume sources • Detection efficiency for high efficiency measurements: • nuclide specific coincidence summing effects • Detection efficiency for volumic samples: • dependent upon sample matrix and density => Direct experimental calibration: - limited number of matrices - specific nuclides - expensive, problems with the management of radioactive material => Additional procedures for a complete calibration required

  5. Coincidence summing effects • Are encountered in the case of measurement of nuclides which decay through the emission of coincident radiation (cascading photons, X-rays, annihilation photons etc) • Depend on the details of the decay scheme: • Nuclide and peak specific effects • Are much enhanced in high efficiency measurement conditions • Effects: • summing out (coincidence losses from the peak) => decrease of the apparent efficiency • summing in (additional counts in the sum peak) => increase of the apparent efficiency

  6. SPECTRUM OF 22Na (WELL-TYPE DETECTOR)

  7. Coincidence summing: - sample analysis - efficiency calibration: ex: 1 l Marinelli beaker Co-60 => 1173 keV 0.926, 1332 keV 0.924 Y-88 => 898 keV 0.932, 1836 keV 0.920 ex: Well-type detector: Co-60 => 1173 keV 0.445, 1332 keV 0.424 Y-88 => 898 keV 0.472, 1836 keV 0.390 => Accurate procedures for the evaluation of the effects required

  8. Matrix effects • Matrix effects are encountered when the calibration source has a different composition and density than the sample of interest • Depend on: • sample geometry • linear attenuation coefficient • photon energy • detector parameters • Linear attenuation coefficient obtained from: • sample composition and density; • transmission experiments

  9. - Transmission factor approximated by exp(-d) ?

  10. Transmission factors (log scale). Sample: R=3.5, H=2 cm 1 0.1 exp(-d) 0.01 0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 Linear attenuation coefficient (1/cm)

  11. Relative error  (%). Soil sample, R=3.5 cm, H=2 cm. 30 20 10 0 - 10 - 20 10 100 1000 Energy (keV)

  12. Geometry effects • For samples measured close to the end cap of a closed end coaxial detector efficiency very sensitive to geometry details • For some samples (e.g. powder) it is difficult to assure exactly the standard geometry • Detectors parameters may vary in time (e.g. the entrance window of the end cap)

  13. Relative error of activity (A) (%). Soil sample, R=3.5 cm Relative error of activity (A) (%). Soil sample, R=3.5 cm 0 - 2 - 4 - 6 - 8 - 10 10 100 1000 Energy (keV)

  14. GESPECOR • Monte Carlo based software dedicated to solve problems in gamma spectrometry: • - computation of coincidence summing corrections • - computation of self-attenuation effects (matrix effects) • - computation of the efficiency • Typical applications: environmental spectrometry • - detectors: HPGe (closed end or well-type), Ge(Li) • - sources: cylinder, Marinelli, point, parallelepiped, ring • - matrix: any (known composition) or known  • - nuclides: ~ 100 (for coincidence summing effects) • Extension for very large samples: variance reduction techniques (focused photon emission, weighted emission point)

  15. Direct efficiency calibration: - typical sources (cylinder, Marinelli, point sources) - special geometries: Parallelepiped (Al-26 in meteorite samples) Spherical source (Rn-222 sources) In situ measurements Drum waste containersEfficiency transfer – less sensitive to detector details

  16. Summary and conclusions • Accurate assessment of the radioactivity of environmental samples – a challenge • Coincidence summing effects • Matrix effects • Geometry effects • The GESPECOR software can solve typical problems required by an accurate assessment of the radioactivity of environmental samples