2025 Partisol Sequential Sampler Deborah Bowe Thermo Electron Corporation
2025 Verification & Routine Maintenance • Verification • Ambient air temperature • Filter temperature • Ambient pressure • External leak check • Internal leak check • Flow rate
2025 Verification & Routine Maintenance • Routine Maintenance • Filter cassettes: inspect damage or contamination • Filter cassette magazines: Generally not necessary to clean, but if damaged certain parts can be replaced • WINS impactor: In accordance with USEPA guidelines for PM2.5 • VSCC: In accordance with U.S. EPA guidelines • External leak check: Every 4 weeks of use • PM10 Inlet: Every 14 days of use
2025 Verification & Routine Maintenance • Routine Maintenance • Particle trap filter: inspect 6 months, replace as needed • V seals: Check and clean as needed • Rainhoods: Clean air screens 6 months • Batteries: Check every 6 months replace as needed • Pump: Check every 6 months and rebuild; needed average lifetime is 12 to 18 months
2025 Backup Battery Check • Measure across 188_VBAT & PGND • Drops below 4.2VDC replace batteries
2025 Pump Check Pressure Check: Quality Pump should give 40 psi on the exhaust
Partisol 2000 Single Channel FRM Sampler Deborah Bowe Thermo Electron Corporation
2000 FRM Audit Screen Ambient temperature + 2 °C Filter temperature + 2 °C Ambient pressure + 10 mmHg
2000 FRM External Leak CheckManual method • Record reading on gauge 30 seconds after pump valve is closed • Should not drop more than 8.5 inHg in 30 seconds. Corresponds to 80ml/min rate
2000 FRM External Leak CheckSemi Automatic • Tolerance is 5” in 60 seconds
2000 FRM Internal Leak Check • Manual: 8.5”Hg in 30 seconds • Semi-automatic: 9.3”Hg in 30 seconds
2000 FRM Flow Audit Streamline Pro, FTS or Bios flow meter + 5% tolerance