UES: Utrecht Echelle Spectrograph

UES: Utrecht Echelle Spectrograph. High-resolution spectrograph on the 4.2m WHT (1992- 2002 ) 50,000 < R < 80,000 Spectral range 300-1100nm Cross-dispersion, (3 prisms) + 2 diffraction gratings. GTC. HORUS.

zofia
Download Presentation

UES: Utrecht Echelle Spectrograph

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. UES: Utrecht Echelle Spectrograph • High-resolution spectrograph on the 4.2m WHT (1992- 2002 ) 50,000 < R < 80,000 • Spectral range • 300-1100nm • Cross-dispersion, • (3 prisms) • + 2 diffraction • gratings

  2. GTC

  3. HORUS • Abundances everywhere (stars, clusters, comets, nebulae, ISM, planets…) • Follow-up faint metal-poor stars • Transmission spectroscopy of planets • Precision radial velocities (if ultra-stable): planets, asteroseismology • …

  4. Optical elements + detector 4k x 4k Fairchildt Collimator Prisms (3) Gratings (2) Camera 5

  5. UES optical design 3m 3m 6

  6. HORUS optical design 7

  7. pre-optics Microlense arrays (Amus) 50mm fibers (UV050125140, Ceramoptep) 8

  8. Acquisition arm 9

  9. Coupling to OSIRIS inside OSIRIS 10

  10. Design Image sliced on the Nasmyth focus into 12 square 357 µm (0.43 arcsec) microlenses feeding 50 µm diameter fibers 2 calibration fibers 4 sky fibers

  11. Light enclosed in fibers 66% 19% 84% at 1” 6% 57% at 1.5” 38% at 2”

  12. Performance • R ~ 50,000 (1.5-3.5 Å/mm and FWHM ~ 3 pixel) 16% throughput on blaze (0.4 original × 0.65 fiber × 0.8 pre-optics × 0.90 folds × 0.84@1”) • New detector is larger (40962 15 µm) • 40 mm (Δx limited by corr. plate) 61 mm (Δy CCD) • ~ 1-2 m/s long-term precision 370-570 nm contiguous 570-900 nm partially (estimated)

  13. 10ºC<T<20ºC T stability GTC foundations ≈ 0.3 K T variations damped by x5 Data (T, RH) being taken since February 2012    Coudé room

  14. Isolating structure Al 3.15 x 2.5 x 2 m (38.4 m2) stability +/- 0.1K Ref. temperature TBD Δ T = 10ºC implies losses 110w 250w chiller, 20l/min Environment

  15. Vacuum chamber • Stainless steel AISI 304 • 2.9m x 1.6m (1.5m Ø) • mass 1320 Kg • 4 isomatic dampers • Thermal transmission (238w/K): • conduction: 96w/K • convection: 140w/K • radiation (net): 2w/K

  16. All optical elements at a constant temperature 3 hangers to join bench and chamber (G10). Thermal transmission (6w/K): Conduction: 5.3w/K Convection: NA (pressure under 200 mTorr) Radiation (net): 0.7w/K Optical bench

  17. HORUS data reduction and analysis software • Basic processing: bias, flatfielding • Extraction: standard, optimal, perfectionism • Calibration: flux (rough), wavelength (critical) • RVs: cross-correlation, template matching, research! • Spectral classification: stars, galaxies, clusters • Stellar parameters: Teff, logg, [Fe/H], micro, [α /Fe] …

  18. HORUS-P

  19. HORUS-P

  20. The 2 UES gratings 21

  21. Zooming-in 22

  22. Science CCD Lab data

  23. Science Lab data Na-Ar Na-Ne Th-Ar

  24. Lab studies • Stability with temperature • Stability with image • Spectrophotometric precision • Solar center-to-limb variation Shift (pixels) Frame #

  25. HORUS • Estimates indicate 2 years of work and 350 000 euros needed • Have used internal (IAC) funding for design studies and some components (200 000 euros for detector from Spectral Instruments + about 15000 euros recently for design, fibers and other hardware) • IAC internal (Severo Ochoa) funding secured (200 000 euros) • Vacuum chamber is delayed: internal review + funding issues • Attractive alternative is to take it under a simplified form to GTC while vacuum-chamber development continues

  26. HORUS • Most high-resolution science projects do not require extreme stability • Pre-optics can be optimized for poor seeing conditions that GTC cannot use effectively now • HORS can use Nasmyth platform: it fits and with a shorter length (reduced losses) • Essentially ready to go!

  27. HORUS: tentative schedule • Basic instrument completion: June-September 2013 (cover, lab fiber link, Fold 2) • Lab tests: Nov-Dec 2013 • GTC Fiber link, acquisition arm: January –February 2014 • Calibration Unit: March-May 2014 • Comissioning: June-July 2014

  28. Thanks! • Juan Calvo, Ramon Garcia Lopez, Felix Gracia, Rafael Rebolo, Vicente Sanchez, Samuel Santana, Fabio Tenegui (IAC) • Chuck Henderson, Basil Blank (PulseRay) • Fred Hearty (Univ. Virginia)

  29. HORUS issues • Wavelength calibration: can use ICM? otherwise a supply from IACUB) • CCD had faulty electronics: now solved • CCD cryocooler: can be stored in isolated cabinet • Fiber-fed. Fibers in hand. Several fiber links produced for lab tests • Acquisition electronics: working, if too simple • Missing folder: on lease from P. L. Palle (solar lab at Teide Observatory) calibration unit is needed (ThAr hollow-cathode + power