1 / 38

The Medley ® Pathway to Patient Safety

The Medley ® Pathway to Patient Safety. “If you find a path with no obstacles, it probably doesn’t lead anywhere”. July 2001: Investment in the Medley ® Infusion System. Hamilton first to implement hospital-wide Implemented a “pump for a pump” Guardrails were only a promise

Jims
Download Presentation

The Medley ® Pathway to Patient Safety

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The Medley® Pathway to Patient Safety Mark Duffett Leslie Gauthier

  2. “If you find a path with no obstacles, it probably doesn’t lead anywhere”

  3. July 2001: Investment in the Medley® Infusion System • Hamilton first to implement hospital-wide • Implemented a “pump for a pump” • Guardrails were only a promise • Safety features: • drug calculator • free flow protection • tamper protection • ergonomic design • Software log—new way of investigating incidents • Beginning of a partnership with Alaris Medical

  4. Initial Issues Identified • “Air-in-line” alarms • Secondary infusions: • Different, louder alarm when infusion complete • Alarm if secondary med not “unclamped” • Infusion issues related to high rates (>200mL/hr) • Keypad noises • awaking patients or a safety feature? • No graphic display of site pressure • Door Failure • VTBI rate set too low

  5. Canada Day Weekend 2001: Problems • “System/Communication Errors” appeared on units in the PICU • There was the perception that the devices powered down and stopped infusing • Investigation: • Software log confirmed the devices continued to infuse • Caused by a capacitor problem First upgrade!

  6. February 2002: Guardrails • Significant milestone for Hamilton • Consensus building for the dataset • Practice standardization But… • No bolus feature available • Limited number of drugs supported • Added complexity

  7. July 2002: Disaster Mode • Multiple incidents the week of July 16th • “Runaway infusions”, incorrect rates and programming • Response: • Communication and education for all staff • Mandated use of solusets • Reiterate mandatory use of guardrails • Eliminate bolusing from continuous infusions

  8. July 2002: Investigation • Joint investigation with HHS, St. Joseph’s and Alaris • Findings: • “Fluid Ingression” problems • 4 malfunctioning keypads identified • Log analysis showed that after each keypad malfunction ”START” was selected • Twofold problem: • Technical malfunction • User/technology interface

  9. June 2002: Results • Improved processes for incident investigation, including: • Biomedical Support 24 hrs/ 7 days/week • Medley incident form • Segregation of the Medley devices and tubings • Completion of incident reports • Feedback to Alaris: • Customer Call Centre and the turnaround time for analyzing the software logs • Why are Guardrails under the “OPTIONS” key?

  10. June 2002: Results • Provided staff an opportunity to verbalize their concerns regarding their perceptions of the safety of the Medley® system • Upgrade in August 2002 to replace keypads • Practice issues • Guardrails underused • Bolusing from continuous infusions • Competency checklists

  11. October 2002: Upgrade Hardware: • APM seals Software: • Guardrails for bolus doses of infusions • Increased capacity of Guardrails system • CQI software installed

  12. May 2003: Upgrade Hardware: • Correct “System/Communication Errors” Software: • revisions to the Guardrails • 1st download of CQI data: • 10% of all devices in the city over 6 months

  13. Dobutamine Incident RN description of event: • The pump “all of a sudden” increased the rate to 905 mL/h following the hanging of a new 250 mL bag of dobutamine Software log showed: • The pump alarmed for the VTBI absorbed • Door opened • 75 seconds later new VTBI selected • 245 entered in the dose field instead of the VTBI field • “START” selected • Guardrails were not being used

  14. Overinfusions • A number of incidents where the RN identified that the pump “ran away” • Software log showed no problems • Most likely caused by improper set loading resulting in free flow Solution to be implemented soon!

  15. December 2004: Upgrade Hardware: • Capacity for syringe platform (NICU, PICU, OR) • Corrective actions for fluid ingression Software: • Guardrails are not under “OPTIONS” • Updates to the Guardrails • to deal with issues identified in the CQI download in May 2003 Direct access to the Guardrail Software • Easier access to the CQI data • 2nd CQI download

  16. June 2004: Audit Software: • Download of CQI data • Are channel labels used appropriately? • Guardrails audit • Use of guardrails • Overrides of alerts • Appropriate patient profile used • Can we gain insight into the overrides? Hardware: • Are we using this expensive technology appropriately?

  17. June 2004: Device Use Audit

  18. June 2004: Guardrails Audit

  19. System Improvement Using CQI Data • Does the system prevent errors? • Can we increase Guardrail use? • Can we reduce nuisance alerts? • Can we improve the response to alerts?

  20. Guardrail Changes Opportunities to improve practice! CQI Data Adverse Events Prevented What Do With CQI Data?

  21. Improving Guardrails: Then Guardrails Changes Practice Changes Staff Feedback Clinical Practice

  22. Improving Guardrails: Now CQI Data Guardrails Changes Practice Changes Staff Feedback Clinical Practice

  23. 69 potentially clinically significant events = 576 events/year prevented at Hamilton Health Sciences Summary of Downloaded Data(analyzed by event)

  24. Can We Reduce Nuisance Alerts?

  25. Can We Increase Guardrail Use?

  26. Can We Increase Guardrail Use?

  27. Can We Improve the Response to Alerts?

  28. Work in Progress: Hardware • Issues with secondary medication infusions: • Alarms, clamps etc. • Correction for the free flow problem • Fluid Ingression • Batteries • Keypad Failure related to an electronic circuit • Disposables, e.g. blood filters for NICU • Allocation of this expensive resource

  29. Work in Progress: Software • Improve Guardrails dataset • Guardrails for secondary infusions • Improvements to Guardrails software • Improvement in CQI data downloading and analysis.

  30. Work in Progress: Practice • Better incident reporting • Improve front line staff’s knowledge, skill and understanding of the safety platform • Standardization of practice • Improve decision making skills related to alerts and overrides • Improve understanding of over-rides • Disseminate audit/download results.

  31. What Have We Learned? • This is a process, not a project. • This is a complex process. • Buying technology to improve patient safety isn’t an easy fix. It requires continuous investment in staff and infrastructure to support it at make it successful.

  32. It took Thomas Edison 2000 experiments to invent the light bulb. He said, “I never failed once. It just happened to be a 2000 step process”

More Related