320 likes | 1.56k Views
OBJECTIVE . 10/1/2011. 2. 1. Describe the three attributes of circulating blood and their relationships.2. Identify types of clients in which hemodynamic monitoring would be indicated.3. Describe the types of catheters used for hemodynamic monitoring.4. Discuss the normal and abnormal values obtained through hemodynamic monitoring as they relate to specific client situations.5. List the potential complications in use of hemodynamic monitoring devices. 6. Explain nurs9444
E N D
1. 10/2/2011 1 HEMODYNAMIC MONITORING
2. 10/2/2011 2
3. HEMODYNAMIC MONITORING 10/2/2011 3 DEFINITION:
Using invasive technology to provide
quantitative information about vascular capacity, blood volume; pump
effectiveness, and tissue perfusion
4. HEMODYNAMIC MONITORING 10/2/2011 4 ATTRIBUTES OF CIRCULATING BLOOD:
1. PRESSURE
2. RESISTANCE
3. Flow
5. HEMODYNAMIC MONITORING 10/2/2011 5 TYPES OF CATHETERS USED FOR HEMODYNAMIC MONITORING:
1. Pulmonary Artery Catheter (swan ganz)
2. Arterial Pressure Catheters
3. Central Venous Pressure or CVP monitoring
6. HEMODYNAMIC MONITORING 10/2/2011 6 INDICATIONS FOR HEMODYNAMIC MONITORING:
1. Potential or actual alteration in CO
2. Potential or actual alteration in fluid volume
7. HEMODYNAMIC MONITORING 10/2/2011 7 PULMONARY ARTERY MONITORING EQUIPMENT
1. Invasive catheter (swan ganz)
2. Transducer
3. Amplifier/recorder
8. HEMODYNAMIC MONITORING 10/2/2011 8 SWAN GANZ CATHETER
a. Distal Lumen: the PA (pulmonary artery)
b. Proximal Lumen: (CVP port)
c. Inflation Balloon
D. Thermostat Lumen
e. Additional Lumens
* VIP - venous infusion port * Pacing Port-
9. 10/2/2011 9
10. HEMODYNAMIC MONITORING 10/2/2011 10 Assessment Data, Pressures, Waveforms of the swan ganz catheter
a. RAP or CVP
b. PAP -
c. PCWP (PAWP)
d. Waveforms: RAP, RVP, PAP, PCWP
e. Using Pressures to Calculate Other Hemodynamic Variables
* CO, CI, MAP, SVR, PVR
11. HEMODYNAMIC MONITORING 10/2/2011 11 NURSING RESPONSIBILITIES
1. Pre-Insertion:
a. Patient & significant others teaching.
b. Thorough baseline assessmentc. Equipment set-upd. Positioning of the cliente. Equipment check (calibration, zero
etc.)
12. HEMODYNAMIC MONITORING 10/2/2011 12 NURSING RESPONSIBILITIES
2. Insertion,
(patient is often awake for the procedure)
a. Collaboration
b. Assist with equipment preparation
c. Monitor patients response to treatment
e. Recording
13. HEMODYNAMIC MONITORING 10/2/2011 13 NURSING RESPONSIBILITIES
3. Post-Insertion
Chest X-ray for placemen
Sterile OCCLUSIVE dressing, stabilization of the catheter
Patient comfort
Assess and document pressures/data
Catheter maintenance
Monitor patients response
14. HEMODYNAMIC MONITORING 10/2/2011 14 NURSING RESPONSIBILITIES
Maintenance of the System:
a. Label and date lines, change tubing as per
unit policy
b. Maintain & change dressing
c. Alert for waveform and pressure changes
d. Maintain pressure/pressure bag
e. Keep balloon deflated between PCWP
readings f. Collaboration and communication with
physiciang. Remember to care for the patient as well as
your equipment
15. HEMODYNAMIC MONITORING 10/2/2011 15 POTENTIAL COMPLICATIONS
1. Dysrhythmias2. Infection 3. Pneumothorax4. Pulmonary infarction 5. Balloon rupture and air embolism6. Pulmonary artery rupture
7. Endocarditis
16. HEMODYNAMIC MONITORING 10/2/2011 16 ARTERIAL PRESSURE MONITORING:
1. Indications
2. Assessment data:
* MAP = (diastolic x 2)+(systolic)
3
* MAP - ICP = CPP
17. HEMODYNAMIC MONITORING 10/2/2011 17 ARTERIAL PRESSURE MONITORING:
3. Arterial Waveform
a. Aortic valve opens and blood is ejected from LV and is recorded as an increase in pressure
b. Highest point represents systole
c. Lowest point is diastolic pressure
18. HEMODYNAMIC MONITORING 10/2/2011 18 ARTERIAL PRESSURE MONITORING:
4. Nursing Responsibilities
5. Potential Complications:
a. Thrombosis at catheter tip b. Infection c. Bleeding d. Loss of flow distal to extremity
19. Nursing Responsibilities 10/2/2011 19 Prepare the setup and the circuits
Assist in the procedure
Caring for equipments as well as for the patient.
Calibration
Obtaining the readings
Interpreting the results and informing the physician
Providing the suitable management
Documenting
20. 10/2/2011 20 Swan Ganz Cathetarization
21. PROCEDURE FOR SETTING UP A HOMODYNAMIC CIRCUIT 10/2/2011 21 Equipment Required:
1. 2 single transducer kit
2. 2 Pressure bags
3. 2 500 mL bag of .9 NaCl with1,000 units of
heparin
4. 2 transducer clamp
5. 2 pressure cable connected to Protocol
monitor
22. 10/2/2011 22
23. 10/2/2011 23
24. 10/2/2011 24
25. 10/2/2011 25
26. 10/2/2011 26