Ion Chromatography

Ion Chromatography is designed with a bipolar conductivity detector to improve instrument detection performance and stability. Boasts a modular design coupled with advanced <br>circuit technology. It provides a flow range spanning from 0.001 to 9.999 ml/min and has a detection range spanning from 0 to 35,000 u00b5S/cm. It is exclusively crafted for the <br>separation, identification, and quantification of ions within a sample.

Zola2
Download Presentation

Ion Chromatography

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Ion Chromatography www.labmate.com | info@labmate.com Overview Ion Chromatography is designed with a bipolar conductivity detector to improve instrument detection performance and stability. Boasts a modular design coupled with advanced circuit technology. It provides a flow range spanning from 0.001 to 9.999 ml/min and has a detection range spanning from 0 to 35,000 µS/cm. It is exclusively crafted for the separation, identification, and quantification of ions within a sample.

  2. Specifications : Ion Chromatographic Type High pressure and low pulse Pump double plunger tandem advection pump Maximum 35 MPa (PEEK) pressure Pressure (0.2 to 0.5) ml/ min ≤ 3 % display accuracy (0.5 to 1.0) ml/ min ≤ 2 % > 1.0 ml/ min 2 % Flow stability 0.001 to 9.999 ml/ min Flow range 0.001 to 9.999 ml/ min Pressure ≤ 0.5% pulse

  3. Conductivity detector Type Temperature Control and Bipolar Conductivity Detector Detection Bipolar conductivity detection mode Cell volume ≤ 0.8 µL Detection 0 to 35000 µS/ cm range Detection ≤ 0.0020 nS/ cm resolution Output -6000 to +6000 mV voltage (adjustable) Baseline ≤ 0.05 % F.S. noise Baseline ≤ 3 % FS/ 30 min drift Operating Room temperature +5 to 60°C temperature (41 to 140°F) range

  4. Controlling ± 0.01°C temperature accuracy Maximum 10 MPa pressure Linear range ≥ 103 Instruments ≥ 0.999 linearity Quantitative ≤ 1.0 % repeatability Qualitative ≤ 1.0 % repeatability Minimum Cl ≤ 0.0005 ug/ ml detectable concentratio BrO3 ≤ 0.001 ug/ ml n Li+ ≤ 0.001 ug/ ml

  5. Flow system Plastic flow Made of PEEK materials path Six-way PEEK material, 5000 psi Valve pressure, independent automatic collection, and flow function Degassing Device Inner 300 µl volume Degassing 7.5 ml volume Maximum 10 ml/ min flow rate Degassing Degassing 90 % at 1ml/ min efficiency Organic 1 tolerance

  6. Built-in Eluent Generator Eluent KOH/ NaOH types Eluent 0.1 to 100 mM concentrat ion range Concentra 0.1 mM tion increment Flow rate 0.1 to 3.0 ml/ min change Maximum 20 MPa pressure Minimum 5 MPa pressure Power supply 150 W Dimension 350 × 470 × 650 mm

  7. Net weight 34 kg Gross weight 40 kg Features : - Featuring a gradient elution system, efficiently detects small molecules in complex samples - Incorporate temperature control for optimal analysis conditions - Includes automated components and software to streamline analysis - Built-in low-pressure degassing technology to remove bubble interference - Automated self-regenerating electrolytic microfilm suppressor - Enhanced with standard dilution function that saves time, labor, and costs - RS232 and USB interfaces Standard Accessories : Ion Chromatography host:1 High-pressure advection pump:1 Conductivity detector:1 Anion column:1 Anion guard column:1 Suppressor:1 Chromatography workstation:1 Column thermostat system:1

  8. Spare parts kits :1 Vacuum degassing pump:1 Sand core filtration devices:1 Pretreatment column:20 0.22 µm syringe filter:200 Eluent generator:1 Optional Accessories : Autosampler (20 samples) Autosampler (54 samples) Autosampler (108 samples) Anion column Guard column Anion suppressor Spare parts kits Applications : Ion Chromatography finds application in various scientific and industrial fields such as Environmental Monitoring, Laboratories, and more. Pharmaceutical Analysis, Research Labmate Scientific Email: info@labmate.com | Website: www.labmate.com