An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. How to count PMT pulses Victor Kornilov Sternberg astronomical institute I count photons 25 years.

  2. Classical scheme of a photon counting Pulses from PMT A~3mV, I<10mA Pulses from amplifier to counter A~3V, I~60mA

  3. Problems • Both signals have the same frequency characteristics • And output pulses influent at input pulses • Difficulties are enlarged for pulses < 30ns • Ground problem, External noise Evident solution: To count pulses near amplifier, in this case a system of data transmission into host computer needed. (Hamamatsu and ETL produce a photometric modules with RS232 interface)

  4. Devices control and Data acquisition • Physical layer 1 – RS485 (Recommended Standard) interface • Physical layer 2 – Inexpensive components (RISC mcontrollers) • Logical layer 1 - Exchange by bytes united in addressed packets, terminated by special signals, CRC • Logical layer 2 – Nonsymmetrical structure, exchange via host only, passive, active, induced transaction • Logical layer 3 – compact commands, data packets, direct access to module parameters, containers • Interface boards, converters, drivers, software. • We use this unnamed system 5 years (described in MASS documentation).

  5. Transaction modes • 1 - Command (reply signal only) • 2 – Data request • 3 – Active data transmission

  6. Properties of the RS485 • Communicate multiple nodes bi-directionally over a single twisted pair. • Excellent noise rejection (differential balanced) • High data rate with long cable • General robustness.

  7. Considerations of RS485 • Configuration • Termination and stubs • Data rate vs cable length • Cabling • Fail-safe biasing • Grounding and shielding Useful papers: www.national.com, Application notes www.bb-europe.com, RS-422 and RS-485 Application Note

  8. Amplifier-discriminator • Preamplifier - transimpendance SA5205, band 600MHz, K=9 (Philips) • Pulse discrimination – Ultra fast comparator AD8611, 4ns • Digital control of discrimination level • Noise <200 mV • Deadtime with Hamamatsu R647 – 17ns, with R7400 ~10ns Permits to measure fluxes up to 10 Mp/s (if counter permits too…)

  9. Counter, control and exchange • Pre-counter 74HC(T)4520, fmax=60MHz • Kernel - m controller ATMega8-16 • 2 channels with 16 bits, may be 24 bits • Hardware generation of integration strobe • In system reprogramming of m-programm • PMT current control • Exchange rate up to 2Mbit/s