An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. EPI – Echo Planar Imaging Joakim Rydell 2004-03-26

  2. Echo Planar Imaging • Like ordinary MR, but faster • Ordinary MR: approximately one minute • EPI: less than one second • The Nobel Prize in Medicine 2003 • Lauterbur (MR) • Mansfield (EPI)

  3. Repetition, k-space • Fourier transform of image • Traversed using magnetic field gradients

  4. Repetition, pulse sequence • Slice selection • Phase encoding • Frequency encoding

  5. Introducing EPI Ordinary MR • Acquire all rows in k-space with just one excitation EPI

  6. EPI – The Movie

  7. Going even faster – Partial Fourier • Real image  hermitian symmetric k-space • Acquire half of k-space, compute the rest • Not always as good as it seems...

  8. Partial Fourier, drawbacks • Phase errors • Slowly varying • Approximately 65 % of k-space required • SNR decrease • Full k-space acquisition gives sqrt(2) SNR improvement

  9. Why use ordinary MR? • Signal decay • Low SNR • N/2 ghosting • Geometric distortion • dB/dt

  10. Signal decay • Long readout period • Weak signal near end of acquisition • Blurring in phase-encoding direction

  11. Low SNR • Fast acquisition • Less averaging • Noisy images

  12. N/2 ghosting • Alternating acqusition directions • Shifted sampling positions • N/2 ghosting • Reference scans

  13. Geometric distortion • Low bandwidth/pixel • Pixels shifted in phase encoding direction • Field maps

  14. dB/dt • Fast gradient switching induces eddy currents, causing field inhomogenity • Rapid changes in the magnetic field can cause peripheral nerve stimulation

  15. k-space trajectories • Cartesian • Blurring • N/2 ghosting • High dB/dt • Spiral • Non-uniform sampling • Cartesian spiral

  16. Ramp sampling • Sampling on gradient ramps • Non-uniform sampling • In time • In k-space

  17. Multishot • Both EPI and ordinary imaging • Collect several rows per excitation • Alignment difficulties

  18. Applications • Cardiac imaging • fMRI • ...

  19. EVI – Echo Volumar Imaging • 3D EPI • Acquire a full volume with just one excitation • Not quite possible yet