E N D
1. Energy-Dispersive X-ray Spectrometry in the AEM
3. Example of an X-ray Spectrum 2 Types of X-rays
Characteristic x-rays
elemental identification
quantitative analysis
Continuum x-rays
background radiation
must be subtracted for quantitative analysis
4. Continuum X-rays Interactions of beam electrons with nuclei of specimen atoms
Accelerating electric charge emits electromagnetic radiation
Here the acceleration is a change in direction
The good
The shape of the continuum is a valuable check on correct operation
The not-so-good
I bkg increases as ib increases
I bkg is proportional to Zmean of specimen
I max bkg rises as beam energy rises
Peak-to-background ratio
Ratio of Ichar / Ibkg sets limit on elemental detectability
5. Generation of Characteristic X-rays Mechanism
Fast beam electron has enough energy to excite all atoms in periodic table
Ionization of electron from the K-, L-, or M-shell
X-ray is a product of de-excitation
Example
Vacancy in K-shell
Vacancy filled from L-shell
Emission of a Ka x-ray (or a KLL Auger electron)
Important uses
Qualitative use x-ray energy to identify elements
Quantitative use integrated peak intensity to determine amounts of elements
6. Compute Energy of Sodium Ka Line
7. Families of Lines
8. Fluorescence Yield ? ? = fraction of ionization events producing characteristic x-rays
the rest produce Auger e–
???increases with Z
?K typical values are:
0.03 for carbon (12) K-series @ 0.3 keV
0.54 for germanium (32) K-series @ 9.9 keV
0.96 for gold (79) K-series @ 67 keV
X-ray production is inefficient for low Z lines (e.g., O, N, C) since mostly Augers produced
?? for each shell: ?K???L???M
X-ray production is inefficient for L-shell and M-shell ionizations since?
?Land??M always < 0.5:
?L = 0.36 for Au (79)
?M = 0.002 for Au (79)
9. X-ray Absorption and Fluorescence X-rays can be absorbed in the specimen and in parts of the detector
Certain x-rays fluoresce x-rays of other elements
X-rays of element A can excite x-rays from element B
Energy of A photon must be close to but above absorption edge energy of element B
Example: Fe Ka (6.40 keV) can fluoresce the Cr K-series (absorption edge at 5.99 keV)
10. EDS Dewar, FET, Crystal LN dewar is most recognizable part
To cool FET and crystal
Actual detector is at end of the tube
Separated from microscope by x-ray window
Crystal and FET fitted as close to specimen as possible
Limited by geometry inside specimen chamber
11. Electron-Hole Pair Creation Absorption of x-ray energy excites electrons
From filled valence band or states within energy gap
Energy to create an electron-hole pair
? = 3.86 eV @ 77K
(value is temperature dependent)
Within the intrinsic region
Li compensates for impurity holes
Ideally # electrons = # holes
# electron-hole pairs is proportional to energy of detected x-ray
12. Details of Si(Li) Crystal
13. X-ray Pulses to Spectrum
14. Slow EDS Pulse Processing EDS can process only one photon at a time
A second photon entering, while the first photon pulse is being processed, will be combined with the first photon
Photons will be recorded as the sum of their energies
X-rays entering too close in time are thrown away to prevent recording photons at incorrect energies
Time used to measure photons that are thrown away is “dead time”
Lower dead time -> fewer artifacts
Higher dead time -> more counts/sec into spectrum
Processor extends the “live time” to compensate
15. Things for Operator to Check Detector Performance
Energy resolution (stamped on detector)
Incomplete charge collection (low energy tails)
Detector window (thin window allows low-energy x-ray detection)
Detector contamination (ice and hydrocarbon)
Count rate linearity (counts vs. beam current)
Energy calibration (usually auto routine)
Maximum throughput (set pulse processor time constant to collect the most x-rays in a given clock time with some decrease in energy resolution)
16. Energy Resolution Natural line width ~2.3 eV (Mn Ka)
measured full width at half maximum (FWHM)
Peak width increases with statistical distribution of e-h pairs created and electronic noise:
Measured with 1000 cps at 5.9 keV
Mn Ka line
17. X-ray Windows Transmission curve for a “windowless” detector
Note absorption in Si
Transmission curves for several commercially available windows
Specific windows are better for certain elements
18. Ice Build Up on Detector Surface All detectors acquire an ice layer over time
Windowless detector in UHV acquires ~ 3µm / year
Test specimens
NiO thin film (Ni La / Ni Ka)
Cr thin film (Cr La / Cr Ka)
Check L-to-K intensity ratio for Ni or Cr
L/K will decrease with time as ice builds up
Warm detector to restore (see manufacturer)
19. Spectrometer Calibration
Calibrate spectrum using two known peaks, one high E and one low E
NiO test specimen (commercial)
Ni Ka (high energy line) at 7.478 keV
Ni La (low energy line) at 0.852 keV
Cu specimen
Cu Ka (high energy line) at 8.046 keV
Cu La (low energy line) at 0.930 keV
Calibration is OK if peaks are within 10 eV of the correct value
Calibration is important for all EDS software functions
20. Artifacts in EDS Spectra Si "escape peaks”
Si Ka escapes the detector
Carrying 1.74 keV
Small peak ~ 1% of parent
Independent of count rate
Sum peaks
Two photons of same energy enter detector simultaneously
Count of twice the energy
Only for high count rates
Si internal fluorescence peak
Photon generated in dead layer
Detected in active region
21. Expand Vertically to See EDS Artifacts
22. EDS-TEM Interface We want x-rays to come from just under the electron probe, BUT…
TEM stage area is a harsh environment
Spurious x-rays, generated from high energy x-rays originating from the microscope illumination system bathe entire specimen
High-energy electrons scattered by specimen generate x-rays
Characteristic and continuum x-rays generated by the beam electrons can reach all parts of stage area causing fluorescence
Detector can't tell if an x-ray came from analysis region or from elsewhere
23. The Physical Setup Want large collection angle, W
Need to collect as many counts as possible
Want large take-off angle, a
But W reduced as a is increased
Compromise by maxmizing W with a ~ 20° at 0° tilt angle
can always increase a by tilting specimen toward detector -- but this increases specimen interaction with continuum from specimen
24. Orientation of Detector to Specimen Detector should have clear view of incident beam hitting specimen
specimen tilting eucentric
specimen at 0° tilt
Identify direction to detector within the image
Analyze side of hole "opposite the detector”
Keep detector shutter closed until ready to do analysis
25. Spurious X-rays in the Microscope Pre-Specimen Effects
spurious x-rays => hole count due to column x-rays and stray electrons
spurious x-rays => poor beam shape from large C2 aperture
Post-Specimen Scatter
system x-rays => elements in specimen stage, cold finger, apertures, etc.
spurious x-rays => excited by electrons and x-rays generated in specimen
Coherent Bremsstrahlung
extra peaks from specimen effects on beam-generated continuous radiation
26. Test for Spurious X-rays Generated in TEM Detector for x-rays from illumination system
thick, high-Z metal acts as “hard x-ray sensor"
Uniform NiO thin film used to normalize the spurious "in hole" counts, thus
NiO film on Mo grid*
27. Spectrum from NiO/Mo Spurious x-rays
Inverse hole count (Ni Ka/ Mo Ka)
Want high inverse hole count
Fiori P/B ratio
Ni Ka/B(10 eV)
Increases with kV
Want high to improve element detectability
28. Figures of Merit for an AEM Fiori PBR = full width of Ni Ka divided by 10 eV of background
(Ni Ka) / ( Mo Ka) is inverse hole count
29. Beam Shape and X-ray Analysis Calculated probes (from Mory, 1985)
Effect on x-ray maps (from Michael, 1990)
30. Qualitative Analysis Collect as many x-ray counts as possible
Use large beam current regardless of poor spatial resolution with large beam
Analyze thicker foil region, except if light elements x-rays might be absorbed
Scan over large area of single phase => avoid spot mode
Use more than one peak to confirm each element
31. Qualitative Analysis Setup 1 Use thin foils, flakes, or films rather than self-supporting disks to reduce spurious x-rays (not always possible)
Orient specimen so that EDS detector is on the side of the specimen hole opposite where you take your analysis
Collect x-rays from a large area of a single phase
Choose thicker area of specimen to collect more counts
Tilt away from strong diffracting conditions
(no strong bend contours)
Operate as close to 0° tilt as possible (say, 5° tilt toward det.)
32. Qualitative Analysis Setup 2 Microscope Column
Use highest kV of microscope
Use clean, top-hat Pt aperture in C2 to minimize “hole count” effect
Minimize beam tails
(C2 aperture or VOA should properly limit beam angle)
Use ~ 1 nA probe current to maximize count rate
This may enlarge the electron beam (analyze smaller regions later)
Remove the objective aperture
33. Qualitative Analysis Setup 3
X-ray Spectrometer
Ensure that detector is cranked into position
Keep detector shutter closed until you are ready to analyze
Use widest energy range available (0-20 keV is normal)
0–40 keV for Si(Li) detector
0–80 keV for intrinsic Ge detector
Choose short detector time constant (for maximum countrate)
Count for a long time – 100-500 live sec
34. Peak Identification Start with a large, well-separated, high-energy peak
Try the K-family
Try the L-family
Try the M-family
Remember -- these families are related
Check for EDS artifacts
Repeat for the next largest peak
Important:
Use more than one peak for identification
If peak too small to "see", collect more counts or forget about identifying that peak; peak should be greater than 3B1/2
35. Chart of X-ray Energies (0-20 keV)
36. Chart of X-ray Energies (0-5 keV)
37. Know X-ray Family Fingerprints
38. Some Peaks will Look Similar
39. Unknown #1
40. Data Analysis for Unknown #1
41. Unknown #1
42. Unknown #2
43. Analysis of Unknown #2
44. Qualitative Analysis
45. Automatic Qualitative Analysis?
46. Automatic Qualitative Analysis Blunders
47. Summary EDS in the TEM has more pitfalls than in SEM
Use the highest kV available
Understand the effects of:
detector-specimen geometry
spurious x-rays from the illumination system
post-specimen scatter
beam shape and spatial resolution => the “witch’s hat”
Identify every peak in the spectrum
Even artifact peaks
Forget peaks of intensity < 3 x (background)1/2
Collect as many counts as possible
Use large enough beam size to obtain about 1 nA current
Qualitative analysis use:
use long counting times or
thicker electron-transparent regions with a short pulse processor time constant, if appropriate
Assume data might be used for later quantitative analysis (determine t if possible)