An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Enhanced Resolution in Large Volume SrI2:Eu DetectorsJamie BrownNuclear Security Detection Workshop16th April 2019University of Surrey

  2. Why SrI2:Eu?

  3. Why SrI2:Eu? • ~120,000 photons/MeV, (c.f. 63,000 photons/MeV for LaBr3:Ce) • <2.7% FWHM reported for small crystals (<1cm3) • high Zeff • low internal radioactivity • suitability to grow large crystals • excellent proportionality • 435 nm peak emission

  4. SrI2:Eu resolutions

  5. Self-absorption- re-emission What’s the problem?

  6. GEANT4 Simulations of SrI2:Eu • Geant4 optical simulation • WLS (wavelength shifting) method • Absorption and emission spectra taken from literature • Varied intrinsic decay time and absorption length (scaled absorption spectrum) to match peak emission and effective decay time, as function of thickness • This yields a fit to 5% Eu doped SrI2, as used in Alekhin2011 M. S. Alekhin, et al. (2011). IEEE Transactions on Nuclear Science, 58(5), 2519–2527. • Can match properties reported by other authors simply by changing the absorption scaling (as a proxy for different Euconcentrations)

  7. GEANT4 Simulations of SrI2:Eu Light collection changes as function of interaction depth Leads to degradation of energy resolution in larger crystals

  8. GEANT4 Simulations of SrI2:Eu Light collection changes as function of interaction depth Leads to degradation of energy resolution in larger crystals Methods of position determination: Ratio of light arriving at two sides of crystal Decay time changes as function of interaction depth

  9. Regulated N2 environment Low H2O content (<0.5 ppm) Low O2 content (<0.5 ppm) Light collection vs. interaction depth • Experimental setup 6x6x26mm3 SrI2:Eu crystal (RMD) Single 6x6mm2SensL C-series SiPM PTFE wrapping Glove box! 137Cs source 1mm slit collimator (copper) Scan crystal across slit Caen V1730B digitizer

  10. Light collection vs. interaction depth Peak shifts of ~5.5% are observed Other authors have reported shifts of ~1% Need to stabilise/compensate for temperature shifts

  11. Methods to determine interaction depth Two-sided readout • Same 6x6x26mm3 crystal • One SiPM at each end of crystal • PTFE wrapped • Flood illumination data and collimated scan

  12. Further Work Larger Crystals • 2 larger crystals from CapeSym 13x13x26mm3 26x26x26mm3 • Expect some variation in light collection with x, y, and z Can we extend to 3d?

  13. Thank you for listening Thank you to Victor Monier for taking data. Thank you to NuSec for funding.