An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ISSUES OF CLINICAL LABORATORY QUALITY ASSURANCE PROGRAMME IN PAKISTAN Brig (R) Dilshad Ahmed Khan SI (M) Professor of Pathology NUMS, Rawalpindi

  2. Quality Assurance Programme • Quality Control - The aim of quality control is simply to ensure that the results generated by the test are correct & verify that the testing system is working properly. • Quality Assurance - Ensures that the final results reported by the laboratory to the patient are correct: that the right test is carried out on the right specimen, right interpretation/result is delivered to the right person at right time

  3. Factors That Affect Quality of Patient’s Results • The condition of the specimens • Reagents & Equipment • Internal quality controls program • Participation in external quality program • The interpretation & reporting of results

  4. Unreliable QA Program ? Potential consequences of unreliable QA Performance include:- • Patient misdiagnosis • Delays in treatment • Increased costs

  5. QA Performance Failure • Pre-analytical:- • Errors before the sample reaches the laboratory • Analytical:- • Errors during the analysis of the sample • Post-analytical:- • Errors occurring after the analysis

  6. Pre – Analytical Issues Sample Collection Incorrect specimen storage Sample Transport Sample haemolysis/ Incorrect specimen container Sample left overnight at room temp/ long term storage at 4c • Delay in samples delivery/Temperature

  7. Improper Collection of Blood Samples • Sample haemolysis • LDH, potassium or inorganic phosphate • Effects of exercise • creatinekinase / CRP • Collection timing • 24 hour urine

  8. Incorrect Specimen Container • Serum or plasma • PTH, ACTH • Fluoride tubes for glucose • to inhibit glycolysis • EDTA tube • Unsuitable for calcium assay

  9. Incorrect Specimen Storage • Sample left overnight at room temperature • Falsely elevated K, Pi and red cell enzymes • short –term refrigeration • medium term freezing at –20oC • long term freezing at -80oC • Delay in sample delivery • Falsely lowered levels of unstable analytes

  10. Internal Quality Control • Daily running of two control sera • Calculation of their mean and standard deviation • Plotting them on control charts • Daily checking and Interpretation of graphs

  11. Example Mean result (x) = 100 mmol/L Standard deviation (SD) = 1.0 mmol/L Number of results (n) = 100

  12. Mean +/- 1SD x -1SD +1SD Values fall randomly about a mean value. 68% Frequency 99 100 101

  13. Mean +/- 2SD x -2SD +2SD Values fall randomly about a mean value. 95% Frequency 98 100 102

  14. Accuracy ? How correct your result is.

  15. Precision ? The reproducibility of your results.

  16. Which is more Precise ? Performing better ? Potassium mean 5 ; SD = 0.1 mmol/L Sodium mean 140; SD = 2.0 mmol/L

  17. Example Sodium has the better CV and in this example is performing better than potassium. Potassium %CV = (0.1 / 5.0) x 100% = 2.0% Sodium %CV = (2.0 / 140) x 100% = 1.4%

  18. Westgard Rules • Westgard provides multiple QC rules: • Defines acceptability of analytical process • minimises false rejections • maintains high error detection

  19. Westgard Rules

  20. Levey Jennings Chart +2SD 143 +1SD 141.5 X X X X X X X Mean X 140 X X X X X X X -1SD 138.5 X X -2SD 137

  21. Levey Jennings Chart +2SD 143 X X X +1SD 141.5 X X X X X Mean X 140 X X X X -1SD 138.5 X X X X -2SD 137

  22. Levey Jennings Chart +2SD 143 141.5 +1SD X X X X X X X X X X Mean 140 X X X X X X X -1SD 138.5 -2SD 137

  23. Levey Jennings Chart 143 +2SD X X X +1SD X 141.5 X X X X X X X X X X X X X Mean 140 -1SD 138.5 -2SD 137

  24. Shift ? Inaccurate calibration/recalibration Sudden failure or change in the light source Change in reagent formulation Change of reagent lot Sudden change in incubation temperature (enzymes ) Failure in the sampling system Failure in reagent dispense system

  25. Levey Jennings Chart? 143 X +2SD X X X +1SD X 141.5 X X Mean 140 X X X 138.5 X -1SD X X X X -2SD 137 X X

  26. Trend ? • Gradual deterioration of control materials • Deterioration of the instrument light source • Gradual accumulation of debris in sample/reagent tubing • Aging of reagents • Gradual deterioration of incubation chamber • Gradual deterioration of light filter integrity

  27. Root cause analysis

  28. External Quality Assurance? Proficiency testing is thesystem designed to objectively assess the quality of results obtained by laboratories, by means of an external agency.

  29. Benefits of EQA Provide an inter-laboratory comparison Allows participants to identify problems with their testing process (Accuracy) Investigate factors in performance of tests (methods, Instrument, Reagents etc) Supplement internal quality control procedures Identifies improvement opportunities

  30. External QA Programme • International • Regional • National

  31. International External QC • REQAS (Rs 75,000/year) • BIORAD (Rs 80000/year) • CAP (Rs 100000/year)

  32. National External Quality Assurance Programme in Pakistan • E Q A pre-requisite for any country • Designed specifically for the country’s needs • Economical (NEQAPP Rs 2000/year) • Useful in establishing national quality goals • Fulfilling the requirement of accreditation by PNAC

  33. NEQAPP • Clinical chemistry PT programme Started in 1996 • 60 labs invited for the scheme • 47 labs enrolled • Each lab allotted a confidential code number • QC sera sent every three months • Results

  34. List of Analytes • Albumin • Bilirubin • Cholesterol • Creatinine • Glucose • Lithium • Magnesium • Osmolality • Inorganic phosphate • Potassium • Total protein • Sodium • Triglycerides • Urea • Uric acid • Acid Phosph. (Total) • ALT • AST • Alkaline phosphatase • Amylase • CK • LDH

  35. Cumulative statistics: Percentage bias Percentage Inorganic PO4 Glucose

  36. EQA: Causes of Poor Performance • Methodological/reagents reasons (33%) • Technical/intruments reasons (19%) • Clerical reasons (12%) • Blunders (11%) • Un-explained reasons (25%)

  37. Expand Scope of NEQAPP • Expand NEQAPP to all disciplines of Pathology • Funding for necessary infrastructure • Enhance QA of labs on national level

  38. PC-1 for NEQAPP • Submitted in early 2005 • Prepared By: • Brig Dilshad Ahmad Khan • Checked By: • Brig Farooq Ahmad Khan • Approved by: • Maj Gen Masood Anwar, HI(M)

  39. PC-1 for NEQAPP • Team of PNAC visited AFIP • PNAC requested Surg Gen for cooperation • Sectoral committee of PNAC for clinical labs • PC-1 for NEQAPP was approved in 2009

  40. Expand NEQAPP Programme offered in 2010 Clinical Chemistry Immunoassay/ Tumour Marker Haematology Microbiology Histopathology

  41. New Participants 1 Questionnaire EQAS Process 2. Preparation of Panels 3. Panel Distribution 4. Data Collection 5. Data Analysis 6. Final Report

  42. Enrollment Documentation ENROLMENT DOCUMENT: - To be returned to NEQAPP, AFIP Rawalpindi With Following Information: LAB NAME PATHOLOGIST ADRESS E-MAIL TELEPHONE NO PROGRAMME When the details have been entered in the enrolment document it should be sent to NEQAPP, AFIP or Register on website for registration (www. Neqapp.net)

  43. Pre-requisites- NEQAPP • List of analytes • Methods • Instruments/equipments • Reagents

  44. Clinical Chemistry

  45. Immunoassay

  46. Instrument Code

  47. Reagent Suppliers Source

  48. PT Samples • Lyophilized human serum • Similar to patient specimens • Appropriate concentration levels • Good long term stability • Easy to transport • Reconstitution errors

  49. General Chemistry Program : Specimen Design High Pool Specify lowest and highest specifications 4 Level 6 3 + High Pool Low Pool 2 1 Lyophilized human serum Low Pool Level 1