An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Resistance to Glare Disability Provided byBlue Light–Filtering IOLs:Contralateral-Controlled Study W. Andrew Maxwell, MD, PhD California Eye Institute; Fresno, CA and Billy R. Hammond, Jr, PhD Vision Sciences Laboratory, University of Georgia; Athens, GA Disclosures: This study was funded by Alcon Research, Ltd, which also assisted with the preparation of these slides. Dr. Maxwell is a consultant to Alcon.

  2. Introduction • Background: Glare light, such as the light from oncoming headlights, can reduce visual performance. • Purpose: This study was designed to compare glare disability… …in eyes with intraocular lenses (IOLs) that filtered short-wave blue light… …versus… …contralateral eyes with IOLs that did not filter visible blue light. Schematic colors and shapes are not anatomically or scientifically accurate; they are for purposes of explanation only.

  3. Methods: Subjects and Apparatus • Eligible subjects were at least 12 months post-implantation of • a control IOL (model SA60AT; Alcon) and • a contralateral IOL that filtered blue light (spherical SN60AT or aspheric SN60WF; Alcon) • Test apparatus was a 2-channel Maxwellian-view optical system to produce a glare light and a target light on the retina of the subject: neutral density filter holder wedge (circular attenuating filter, to change glare intensity) beam splitter beam splitter eye Channel 1: Glare light Produced by a xenon lamp, to mimic xenon headlights or to approximate daytime sunlight spectrum aperture Channel 2: Visual Target mirror pupil viewer (Produced by a halogen lamp) aperture neutral density filter holder

  4. Methods: A Subject’s View of the Test • 1. Subject saw • A. a 3.1°-diameter halogen-light disk as the target • B. an annulus from the xenon channel with • a shape that was 20° inner and 22° outer diameter • an intensity that was too low to obscure the target A B Technician increased glare intensity • 2. Subject saw • A. the same central 3.1°-diameter disk, but • B. the brightness (ie, the glare) of the annulus was beginning to move toward obscuring the central target A B Technician increased glare intensity • 3. Subject saw • A. no central 3.1°-diameter disk, because • B. the glare from the annulus had obscured the central target. The intensity of the annulus of light at that point was defined as the glare disability value. A B Schematics are not exact representations; they are for purposes of explanation only.

  5. Methods: Replicates and Analyses • Measurements • Masking: The technician who took the measurements was masked to the IOL type in each eye. • Replicates: 5 measures were taken per eye, unless a subject’s values had >5% variability, in which case up to4 additional trials were conducted. If extra trials were conducted, all trials were averaged into the mean result. • Analyses • Comparisons between IOL groups were performed using paired t-tests. • The mean values are presented as log energy values (µW/cm2).

  6. Results: Demographics • At 5 clinical sites, a total of 52 subjects were enrolled. • 28 were women (54% of cohort) and 24 were men (46% of cohort). • Mean age = 76 ± 9 years • All subjects had a control SA60AT lens in 1 eye, and • 50% of subjects (26 of 52) had a contralateral model SN60AT lens(spherical IOL with the blue-filtering chromophore), and • 50% of subjects (26 of 52) had a contralateral model SN60WF lens(aspheric IOL with the blue-filtering chromophore). • The average duration of pseudophakia before entering the study was statistically similar between IOL groups, at • 5.2 ± 1.3 years for the eyes with blue-filtering IOLs, and • 5.6 ± 1.3 years for the eyes with control SA60AT IOLs, with • average duration between contralateral IOL implantations of 0.4 ± 1.1 years.

  7. Results: Monocular Distance Visual Acuity (similar between groups)

  8. Results: Glare Disability • For the overall population of 52 subjects at 5 clinical sites, significantly more glare light (P = 0.04) could be withstood by eyes with blue-filtering IOLs (1.97 ± 0.44 log µW/cm2) than by eyes with non-filtering control IOLs (1.88 ± 0.43 log µW/cm2)before losing sight of the target. • For the subpopulation of13 subjects at the Fresno site,†significantly more glare light(P < 0.05) could be withstood byeyes with blue-filtering IOLs thaneyes with non-filtering control IOLs before losing sight of the target  †Site of Fogg, Maxwell, Lanier & Remington Eye Care

  9. Discussion & Conclusions • Key finding: Compared with eyes with control IOLs, the eyes with blue-filtering IOLs had a higher mean threshold (higher mean resistance) to glare disability. • Proposed mechanism: The light entering the eye from the glare source was scattered toward the retina by the intraocular media, causing a veiling luminance over the visual target. Blue-filtering IOLs prevented some of the glare scatter. • Real-world applications: Analogous situations could occur under bright midday sunlight or when a driver is looking into oncoming headlights. By filtering such glare sources with a blue-filtering IOL, the visibility of a target within an individual’s sight line would be improved, and better edge detection could be provided, as was demonstrated by the model in this study. Read more at: • Hammond BR, Renzi LM, Sachak S, Brint SF. Contralateral comparison of blue-filtering and non-blue-filtering intraocular lenses: glare disability, heterochromatic contrast, and photostress recovery. ClinOphthalmol. 2010;4:1465-1473 • Gray R, Perkins SA, Suryakumar R, Neuman B, Maxwell WA. Reduced effect of glare disability on driving performance in patients with blue light–filtering intraocular lenses  J Cataract Refract Surg 2011;37:38-44