1k likes | 1.74k Views
MR Artifacts. Susceptibility Gradient Field RF K-Space Motion Chemical Shift Gibbs (Ringing, Truncation) Artifacts Aliasing (Wraparound) Partial Volume High Speed Imaging Effect of Field Strength. Spike (Herringbone). Bad data point/noise spike in k-space
E N D
MR Artifacts Susceptibility Gradient Field RF K-Space Motion Chemical Shift Gibbs (Ringing, Truncation) Artifacts Aliasing (Wraparound) Partial Volume High Speed Imaging Effect of Field Strength
Spike (Herringbone) • Bad data point/noise spike in k-space • Either very high / low intensity compared w/ rest of image • Spike is convolved with all other image info during FT • Since each image pixel is a weighted sum of all individual points in k-space • Results in dark stripes overlaid on image • Occurs with high duty cycle gradients sequences • Loose connection/breakdown of connections in RF coil
Uses • Can be used in cardiac imaging • Prep. Pulses applied before imaging sequence • Forms echoes in different parts of k-space • FT produces tags in grid-like pattern • Tags applied at start of each cardiac phase • Images acquired at multiple phases of cardiac cycle • Follow changes in tag position during cycle assess cardiac motion
Spike Artifact Image space K-space
Zipper Artifact • Caused by leakage of e-m energy into magnet room • Results in region of increased noise • Width of 1-2 pixels extends in frequency encode direction • Through entire series • Room shielded from outside e-m signals • Signals from equipment brought into room • OR RF shield compromised
Motion-related • Patient Motion, either: • Voluntary • non periodic • Eye movement • Swallowing • Smearing of image • Involuntary • Periodic • Respiratory • Cardiac • Pulsatile movement of vessels & CSF • Bowel motion • Coherent ghosts formed • Blurring/ghosting in phase-encode direction • Time difference in adjacent points in PE direction relatively long • = TR • Introduces phase difference between adjacent k-space lines
Phase Mis-mapping • PEG has different amplitude every TR, unlike FEG/SSG • As anatomy moves, misplaced in PE direction as PEG changes • Anatomy given different phase values depending on its position along gradient • Time delay between PE & readout • anatomy may have moved between PEG & FEG when placed into k space
Swap PE & FE Directions • Artifact occurs only in PE direction • Change axis/direction • Pick axis Produce least interference w/ ROI • Example: Sagittal C Spine • Usually FE performed in z-axis (head to foot) • Longest axis • PE would then be AP (Y axis) • However: Artifacts in AP direction • Swallowing • Carotids pulsatile motion • Ghosting over spinal cord • Swap PE & FE axes • Y gradient (AP) performs FE • Z gradient performs PE • Artifacts now harmless in head to foot direction
Use Pre-Sat Pulses • Placing pre-sat volumes over areas producing artifacts • Nullifies signal • Reduces artifact • Example: Sagittal C Spine • Pre sat pulse over throat • reduces swallowing artifact • Reduces artifact from flowing nuclei in blood vessels
Respiratory MotionREMEDIES • Breath hold • Patient cooperation req’d • May take multiple breath holds • Respiratory gating • Image acquisition only at certain phases in resp. cycle • Acquisition time ↑ • Respiratory compensation/phase reordering • ROPE (Resp. Ordered Phase Encoding) • PE steps ordered on basis phase in resp. cycle • Difficult if resp. not regular • Real-time navigator echo gating • Echo from diaphragm determines its position • Navigator echo interleaved with actual imaging sequence • Real-time monitoring • data only acquired during specific range of diaphragmatic motion
Patient Motion Without breath-holding With breath-holding; With Cardiac pulsation artifacts
Respiratory Motion Compensation Without compensation With compensation
Navigator Placement Graph shows diaphragmatic movement (white wave & green line). Aqua overlay shows navigator section from which displacement info obtained Yellow boxes: best time to image
Cardiac Pulsation • ECG gating • Time acquisition to occur @ same phase of each cardiac cycle • Coordinate excitation pulse w/ R wave of systole • Further artifact suppression if breath hold
Segmented K-Space • Echo-planar imaging • Single RF pulse/excitation • Continuous reversal of echoes using gradient pulses • Acquire all lines in k-space to form a single image • Alternate lines in k-space read in opposite direction • Prior to FT lines must be reversed • Introduces phase errors in alternate k-space lines • Errors from: • Nonlinear gradient reversal • Eddy currents • Poor shimming • Results in image ghosts
SS- vs. MS EPI Single Shot EPI Multi-Shot EPI # ghost images ↑ As # of discontinuities in k-space ↑ Errors # echoes / shot & # of segments in k-space Minimize artifact May be necessary to obtain additional navigator echo • Single additional ghost image • Reduced intensity • Shifted by ½ FOV • ½ k-space lines are different from other half • “N/2 Ghost” • Reduce artifact • Minimize phase errors
Fast Spin Echo • Segmented k-space artifacts can also occur • Minor timing errors in sequence • Between multiple RF pulses • Between data collection windows • Eddy currents • Call Service Support
Ghost Artifacts Single Shot EPI N/2 ghost multi Shot EPI Initial Image FSE
Flow Artifacts • Flowing blood source of artifacts • Ghosting in phase-encoding direction • SE sequences not as susceptible • Flow appears dark (no signal) • Blood exposed to 90° excitation pulse flows out of imaging section before 180° refocusing pulse • Blood that moved into imaging section never exposed to 90° pulse • GRE Sequences susceptible • In-flow effect bright blood
Reducing Flow Artifacts Pre-saturation Pulse • Attenuate signals upstream of imaging volume • Reduces intensity of fluid flowing into FOV • Apply saturation band adjacent to imaging section • 90° pulse • All spins tilted towards axial plane • Spoiled with gradient crusher pulses before image acquisition • Saturated spins exhibit no signal when moving into imaging volume
Reducing Flow Artifacts Cardiac Pulsation Flow-Related Artifact Suppression Cardiac pulsation artifact After saturation band applied
Reducing Flow Artifacts Flow Compensation/Gradient Moment Nulling • Flowing spins not in phase with static spins when echo forms • Flowing spins brought back into phase by motion-compensating gradient pulses • No effect on static spins • Penalty is increased echo time
Cardiac Pulsation Cardiac pulsation artifact After Motion Compensation
Flow Artifact Flow artifact in right to left direction (phase encoding) from the popliteal vessel, seen as a small bright artifact along the middle of the femoral bone.
Susceptibility Artifact • Tissues placed in magnetic field become temp. magnetized • Slightly alters local magnetic field • Difference in susceptibility between tissues • Field inhomogeneity at tissue boundaries field gradient • Spins dephase faster • Signal loss • Low signal intensity • Signal loss worst for bone-soft tissue & air-tissue boundaries • Air, bone much lower magnetic susceptibility than most tissues • Geometric distortions introduced
SE vs. EPI • SE sequences less affected • 180° refocusing pulse cancels susceptibility gradients • EPI more severely affected • Echoes are refocused by using gradients over long time period • Minimize by orienting PE gradient along same axis as susceptibly gradients • Reduce artifact by: • Use SE sequences • Reduce echo time • Increase acquisition matrix • Proper shimming over VOI to improve local field inhomogeneity
Reorienting PE Gradient Anterior-Posterior PE (same axis as susceptibility gradients) Left-Right PE direction
SE vs. GRE Susceptibility Spin echo Gradient echo
Metal Implants • Most severe susceptibility artifact • Metal > magnetic susceptibility than tissues • Typically areas of complete signal loss • Minimize effect: • Large receiver bandwidth • Decreased echo time • Fast SE with high bandwidth • Watch heating of adjacent tissue
Chemical Shift • Molecular protons surrounded by clouds of e- • In external magnetic field electric current induced • This current will induce a magnetic moment • Antiparallel to external field • Reduce local magnetic field felt by proton • “electronic shielding” • Protons in water vs. protons in fat • Significantly different chemical environments • Resonance frequencies different • Precess @ different frequencies • Chemical Shift • Larmor frequency shift between water protons & fat protons
Chemical Shift • Artifacts in frequency encode direction • Slight mis-registration of fat content • slight shift in frequency of fat protons • Amount of shift depends on: • # samples in FE direction • Receiver bandwidth Gx x
Reduce Chemical Shift • Less important in FSE imaging • Higher bandwidth receiver window used • EPI very susceptible • Long duration of sampling affects/shifts any off-resonance signal (fat) • Minimized by: • Applying frequency-selective RF pulse to nullify fat signal before imaging sequence • Successful fat sat only if magnetic field homogeneous throughout ROI • Proper shimming2 distinct signal for fat & water • STIR sequences can also be used • Fat T1 short • Can be used to suppress fat signal with inversion recovery sequence • TI needed to null signal from any tissue = 0.7 x T1
EPI & Chemical Shift Reduced chemical shift artifact; fat saturation Severe chemical shift artifact; Insufficient fat suppression
Applies to GE sequences • Since H2O precesses faster than fat it gets 360° ahead of fat in short time period • Thus times when fat & H2O are totally in phase, times when totally out of phase • Dark boundaries when out of phase • Does not only appear in FE direction as CS Artifact of 1st kind Chemical Shift of the 2nd Kind In-phase @ points A, C, E Out-of-phase @ points B, D
In- vs. Out-Of-Phase 3-cm left adrenal mass In Phase Mean signal intensity in mass = 115 Out of Phase Mean signal intensity in mass = 66 • Fat present in lesion • Benign adrenal adenoma
In- vs. Out-Of-Phase metastatic non-small cell and squamous cell carcinoma of the lung demonstrate a left adrenal mass In Phase Out of Phase • No differences in lesion signal between 2 images • No Fat present in lesion • Metastases generally do not contain fat
Aliasing (Wraparound) • Overlapping opposite side of image of signals outside FOV • Spatial encoding of objects outside FOV cannot be distinguished from inside FOV • Imaging FOV smaller than anatomy being imaged • PE direction • FOV = distance along gradient to complete 1 cycle • Gradient will move from -180° to +180° across FOV • If RF transmission coil sensitivity extends beyond FOV • Spins outside FOV excited • Will be part of next cycle in PE direction (i.e. -360° to +360° etc. )
Phase Equivalence • Phase angles of spins outside FOV essentially equivalent to spins within FOV • But on opposite sides of image • For FT: • Spins at x° = x+360° • Results in overlap of signal outside FOV with signal within FOV
Phase-Encoding PE: Modifying phase of spins in a direction of slice plane Step 1: Phase shifts range from -180° to +180° Step 2: Shifts increased by multiple of 180 (360, 540 Etc.) . . . Step N Meaningful phase range
Phase shift Outside FOV Phase = 200° Inside FOV: Phase shifts range from -180° to +180° Outside FOV: Phase shifts < -180° or > +180° Mismapped to equivalent phase inside image. Equivalent phase = -160°
FOV FOV ≥ imaged anatomy Wrap around artifacts FOV < imaged anatomy