1 / 17

Analyte Ionization

Analyte Ionization. Lecture 3. Yuri Kazakevich Seton Hall University. Reversed-Phase HPLC Retention. The retention of ionizable analytes on the same bonded phase can be varied by: Type of organic modifier Concentration of organic modifier Temperature pH of the mobile phase

avel
Download Presentation

Analyte Ionization

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Analyte Ionization Lecture 3 Yuri Kazakevich Seton Hall University

  2. Reversed-Phase HPLC Retention • The retention of ionizable analytes on the same bonded phase can be varied by: • Type of organic modifier • Concentration of organic modifier • Temperature • pH of the mobile phase • However the pH of the eluent is dependent upon the type and concentration of the organic modifier and the temperature.

  3. Introduction • pH of the mobile phase affects: • analyte ionization and solvation • interactions of the analytes with the stationary phase • stationary phase structure and properties • bonded phase stability • dissolution of silica matrix • pH and the type of pH modifier have a major impact on the selectivity alteration. • Recent developments in silica manufacturing and bonded phase chemistry significantly widened applicable pH range (1.5 - 10).

  4. pH Definition pH is the negative logarithm of the proton concentration in the solution pH = -log[H+] Equilibrium constants for acids are usually written in the following form:

  5. [C6H5COO-][H+] Ka = [C6H5COOH] Ionization Equilibria Ka=6.4 x10-5 , pKa=4.19 Increase of the proton concentration in the HPLC mobile phase shifts equilibrium to the left. pKb = 9.4 pKa= 4.6 pKa = 14 – pKb

  6. [C6H5NH3+] [C6H5COO-] pKb=9.4 % ionized pKa=4.2 % ionized pH pH Aniline Benzoic acid Dependencies of Analyte Ionization on the pH of the Mobile Phase

  7. Dependencies of Analyte Retention on the pH of the Mobile Phase * Basic compound: pKa=6 Acidic compound: pKa=3 pKa=Analyte is 50% ionized *Cs. Horvath, W.Melander, I.Molnar, Anal.Chem. 49 (1977) 142.

  8. Dependencies of Analyte Retention on the pH of the Mobile Phase Acid and base are both ionized at pH=5 Acid and base are both neutral at pH=5 • The retention of ionizable compounds at a certain pH is dependent on their ionization state.

  9. RNH3+ RCOOH RNH2 Base RCOO- Acid pH=2.3 Base pH=2.3 Acid pH=6.0 pH=6.0 e c n a b r o s b A 5 10 15 Retention Time (min.) Dependencies of Analyte Retention on the pH of the Mobile Phase Ionization in general decreases hydrophobicity causing a decrease of HPLC retention.

  10. Effect of pH on Acidic Analyte Retention HPLC Conditions: Mobile phase Column – Zorbax Eclipse XDB-C18, 150 mm x 4.6 mm 30% Acetonitrile Flow rate – 1.0 mL/min 70% 20mM Na2HPO4 buffer; variable pH Detection – UV 220nm acidic modifier = perchloric acid Injection volume – 1L

  11. Effect of pH on Retention Factor of Acidic Analytes HPLC Conditions: Mobile phase Column – Zorbax Eclipse XDB-C18, 150 mm x 4.6 mm 30% Acetonitrile Flow rate – 1.0 mL/min 70% 20mM Na2HPO4 buffer; variable pH Detection – UV 220nm acidic modifier = perchloric acid Injection volume – 1L

  12. C H C H C H C H N H 2 3 3 3 2 C H 3 C H 3 N C H N N H H C N N 3 2 3 2,3-dimethylaniline phenylethylamine 2-picoline pyridine pKa=9.83 pKa=5.96 pKa = 5.17 pKa = 4.70 4-ethylpyridine pKa = 5.87 Effect of pH on Basic Analyte Retention 2,4-lutidine pKa = 6.74 Chromatographic Conditions Column: 15 cm x 0.46 cm Zorbax Eclipse XDB-C18 Eluent: 90% Aqueous / 10% MeCN Buffer: 10 mM Na2HPO4•7H2O + xH3PO4 Flow rate: 1 ml/min Temp: 25oC Time (min.)

  13. Effect of pH on Basic Analyte Retention Chromatographic Conditions Column: 15 cm x 0.46 cm Zorbax Eclipse XDB-C18 Eluent: 90% Aqueous / 10% MeCN Buffer: 10 mM Na2HPO4•7H2O + xH3PO4 Flow rate: 1 ml/min Temp: 25oC

  14. Retention of Aniline as a Function of pH Aniline pKa=4.6 Chromatographic Conditions Column: 15 cm x 0.46 cm Zorbax Eclipse XDB-C18 Eluent: 90% Aqueous / 10% MeCN Aqueous: 10 mM Na2HPO4•7H2O + xHClO4 Flow rate: 1 ml/min Time (min.)

  15. Peak Fronting • A condition where the rear of the peak is steeper than the front relative to the baseline. • Related to a secondary chemical equilibrium process • -Analyte Ionization • pH of analysis is close to analyte pKa • -Reaction of analyte with mobile phase components • (ex. Aldehydes in presence of water under acidic or basic conditions)

  16. Peak Tailing • A condition where the front of the peak is steeper than the rear relative to the baseline. • Appears when the analyte concentration exceed the linear range of adsorption isotherm. • Tailing (Depends upon acidity of silanols, ionization state of basic analyte and mobile phase pH).

  17. Concluding Remarks • pH is an effective tool for adjustment of selectivity and retention • pH can be used to optimize the resolution • Reversed phase packings are most stable between pH’s 2 - 8.