140 likes | 341 Views
Coronary Phantom Design for K-edge Angiography. John Jorgensen Sarah Pachtman Punam Patel Marcus Spallek Advisor: Dr. Frank Carroll. MXI Systems. MX 200 Applications Radiotherapy Crystallography Mammography Pediatrics K-edge Imaging Angiography???.
E N D
Coronary Phantom Design for K-edge Angiography John Jorgensen Sarah Pachtman Punam Patel Marcus Spallek Advisor: Dr. Frank Carroll
MXI Systems • MX 200 • Applications • Radiotherapy • Crystallography • Mammography • Pediatrics • K-edge Imaging • Angiography??? http://www.onlinetelemedicine.com/html/product/sam_images/Angiography.jpg
Project Background • Monochromatic tunable X-ray: • Reduced radiation dose to patient • Reduced contrast agent concentrations • K-edge imaging Diagram of MX200’s laser paths and major elements. http://www.mxisystems.com/mx200.html FOR MORE INFO... Free Electron Laser Laboratory http://www.vanderbilt.edu/fel/
Project Goals • Design acoronary phantom • Test contrast agent concentrations • hypersensitivity reaction to iodine (3-12% patients) • current clinical dilution is 1:16 • using monochromatic plain films it is possible to see 1:256 dilutions • Determine optimal angle for stereo imaging Tan θ = 3/30 = 0.10 θ ~ 6° 3” θ 30”
Solution Description • Radio Translucent at 35 keV • Flexible material • Fluid perfusion • Phantom Size • 7.5cm diameter • Vasculature • Size Range 5 - 1 mm • Coronary arteries on surface • Curved channels • Hollow vessels Finger imaged at 19 keV Finger imaged at 26 keV
Phantom Designs Shelley Medical Imaging Technologieswww.simutech.com SolidWorks Model Preliminary Sketch
Ideal Material ReoFlex™ 20 polyurethane resin Actual Material Z-Corp’s ZP 102 plaster-based powder Compatible with printing machine used for prototyping process Ideal for printing fine-detail or thin-walled models Radiotranslucent and flexible formulation Option to infiltrate with Synair Corporation Por-A-Mold polyurethane resin Material Considerations
November December January February March April Schedule Research Material Testing & Design Manufacturing Testing FOR MORE INFO... http://vubme.vuse.vanderbilt.edu/srdesign/2003/group12/NCIIA%20Grant%20Proposal.htm
Current Status • Work Completed • Design Finalized • Manufacturer Contacted • Prototype Being Made • Expected by Early Next Week
Future Plans • April • Create Pump for Dosimetry Testing • Start Testing • Test Dosimetry for Imaging • Test for Optimal Stereo Angle • Write Final Report • Create Poster
Testing • Create Pump for Dosimetry Testing • Run Tests • Test Dosimetry for Imaging • Acquire series of images with different dilutions of contrast agent • Test for Optimal Stereo Angle • Acquire series of images at different rotation angles
Team and Resources • Advisor: Dr. Frank Carroll, M.D. • Monochromatic x-ray machine • W.M. Keck Free-electron Laser (FEL) Center at Vanderbilt University • Materials and tools provided by BME Department and FEL Center • Polyurethane resins from outside contractors Photograph of device with laser cover removed http://www.vanderbilt.edu/fel/Xray/manuscripts/final%20report%20for%20ONR%20grant%2020021218.doc
Questions? Home