An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SweEt validation of Cereal samples – LC-MS/MS analysis Anne Kruse Lykkeberg; Gitte Andersen & Mette Erecius Poulsen; Nordic Pesticide Workshop, 2013, Porvoo/Borgå

  2. Agenda • Short summary of the Method • Non-linear standard curves • Validation

  3. SweEtSampleClean up

  4. LC-method • Analytical column: Genesis C18, 100x3 mm, 4 µm • Solvent A: 10/90 metanol/10 mM ammonium formate, pH 4 • Solvent B: Methanol

  5. Pyrimethanil – linear regression

  6. Pyrimethanil – quadratic regression

  7. Pyrimethanil – cubic regression

  8. Metribuzin – linear regression

  9. Injection volumen - chromatograms Fenhexamid 5 µl 2 µl 0.001 mg/kg 1.00 mg/kg

  10. Injection volumen - standard curvesEpoconazol 5 µl 2 µl

  11. Initial validation – LC-compounds • 28 compounds out of 34 had acceptable recoveries

  12. Spiroxamin • LogP = 2.89 • Low sensitivity LOD = 0.04 ng/ml • Only i SWEET – not QuEChERS • Recovery is good at the highest level

  13. Compounds with lowrecoveries

  14. Compound with highrecoveries

  15. Conclusions • Samples in ethylacetate gives non linear standard curves for manypesticides • Using Quadraticfit and i few cases Cubicfit gives • Acceptable residuals with random distributions • Acceptable recoveries at threelevels • Compounds with Log P below 0 gives a poorrecoveryusing the SweEtmethod