Illustration of the single PE technique to measure the gain of a M64 (H7546A) Multianode PMT

Illustration of the single PE technique to measure the gain of a M64 (H7546A) Multianode PMT. Wallace Raposo, Mario Vaz & Luis Villaseñor. Angra Meeting 4-5 Dec. 2008 Campinas. LED Pulser Prototype. LED activating Pulse (25 ns/div) ‏. Noise. 2 PE. 1 PE. Pulses within 20 ns of

beth
Download Presentation

Illustration of the single PE technique to measure the gain of a M64 (H7546A) Multianode PMT

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Illustration of the single PEtechnique to measure the gain of a M64 (H7546A) Multianode PMT Wallace Raposo, Mario Vaz & Luis Villaseñor Angra Meeting 4-5 Dec. 2008 Campinas

  2. LED Pulser Prototype LED activating Pulse (25 ns/div)‏

  3. Noise 2 PE 1 PE Pulses within 20 ns of right window border (Q miscalulated)‏

  4. We get ~ twice the charge

  5. Final Remark 1. H7546A From the datasheet we expect G=2x10^6 at 950 V We have measured G=3.2x10^6

  6. Final Remark 2. For V<1.1 mV we get 40199 events For V>1.1 mV we get 4398 Therefore the mean value of our Poisson distribution is 4398/40199 = 0.109 Based on this number we expect 4398*0.109/2 = 240 double PEs. From a previous plot the MPV for Single PE amplitude is ~ 1.75 mV, if we cut at Amplitude > 3.5 we get 234 events, i.e. in good agreement.

  7. Final Remark 3. From a previous plot the mean value for the Q distribution noise is 0.15 pC with a sigma of 0.1 pC, i.e. the mean value is 30% of the mean PE charge we got at HV=950 V. We conclude that this method will not work for lower HV for which G < 10^6.