Image Quality Radiographic Resolution.

Image Quality Radiographic Resolution. Demelza Green Jan 2004 Technologies for Imaging. Radiographic Resolution. The resolving power of a radiograph. How close structures can be and still appear separate - fine detail. Radiographic Resolution.

bishop
Download Presentation

Image Quality Radiographic Resolution.

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Image QualityRadiographic Resolution. Demelza Green Jan 2004 Technologies for Imaging.

  2. Radiographic Resolution • The resolving power of a radiograph. • How close structures can be and still appear separate - fine detail.

  3. Radiographic Resolution • Expressed as a measurement of line pairs per mm (lp/mm) • the number of line pairs resolved within a mm determines the quality of the recording medium. • A line pair is made of a line and a space. • Human eye is limited to resolve 5 lp/mm. • Radiographic film 10 lp/mm

  4. Noise • Noise - contributes no useful diagnostic information and serves only to detract from the quality from the image.

  5. Noise • Quantum mottle = image noise. • Fluctuation in the quantity of photons that contribute to the image formation. • Image appear mottled or blotchy. • Result of fast screens that require less exposure - less photons. • Decreases image detail.

  6. Unsharpness • Three Types • Geometric. • Movement. • Photographic.

  7. Unsharpness • SOD - source object distance • SID - source image (receptor) distance • OID - object image (receptor) distance

  8. Geometric Unsharpness (Ug) • Size of focal spot. • Image magnification - • Geometric unsharpness = • Poor film/screen contact. SID SOD Focal spot size x OID SOD

  9. Movement unsharpness (Um) • Exposure time. • Subject motion - voluntary and involuntary. • Immobilisation.

  10. Photographic Unsharpness (Up) • Screen and film crystal size • Duplitised screens/ emulsions (crossover) • Reflective screen layers.

  11. Unsharpness • Total unsharpness • = √ Ug2 + U m2+Up2

  12. Contrast • What is contrast? • The capacity of showing different densities • The whiteness of the bone against the blackness of the film and the range of greys in between.

  13. Radiographic Contrast • Subject ContrastFilm/Receptor Contrast • KVp Film Type • Tissue composition Direct exposure/ intensifying screens • Contrast Medium Processing Conditions

  14. Contrast • High Contrast film • Black/white • CXR for bony mets or rib fractures. • Low Contrast film • Shades of grey • CXR for lung lesions or chest infections.

  15. Radiographic Contrast • Tissue Composition. • Composition and thickness of tissue affect absorption. •  create a range of densities (contrast) • anatomical structures with a wide range of tissue composition will produce high subject contrast. • Those with a similar range produce low subject contrast

  16. Radiographic Contrast • Contrast media • for imaging areas of low subject contrast. • Positive contrast agents e.g. barium have high atomic no. and absorb more x-rays than the surrounding structures. • Negative contrast agents e.g. air have low atomic no. produce more density than surrounding structures.

  17. Scatter. • when the x-rays transverse the body tissues either are absorbed or scattered or transmitted without interactions • scattered photons reaching the detector form a ‘fog’ on the radiographic which blurs the image and obscures the anatomical details  degradation of the image

  18. Scatter • scattered photons can be absorbed by the body tissues  increase in patient’s dose • minimise the scattered radiation generated in the body tissues. • minimise the scattered radiation reaching the detector.

  19. Contrast • Kilovoltage. • Low KVp decreased penetration • more absorption • more density differences - high contrast

  20. Contrast • Kilovoltage • High KVp increased penetration. • less absorption. • less density differences - low contrast. • Increase in scatter - unwanted interactions with film.

  21. Scatter Radiation • Using the minimum field size (collimation); the larger the irradiated volume the higher the scatter generated in the body; minimise the irradiated volume by correct collimation

  22. Scatter radiation

  23. Digital Imaging • Define spatial resolution. • Describe factors influencing spatial resolution. • Fully describe the concept of windowing. • Define the terms window width and window level.

  24. Spatial resolution • Matrix size - greater the matrix then smaller the pixel size improved spatial detail. • As with conventional radiography detail lost during imaging process e.g. on the plate.

  25. Digital Resolution • Typical TV monitor 2.5 lp/mm • Radiographic Film 10 lp/mm •  need a tv monitor martix of at least 2048x2048 - fine line monitors - expensive. • Resolution poor.

  26. Windowing There are 2000 levels of grey. Human eye cannot detect between small level changes. Windowing is a contrast enhancing technique. Allows us to pick a section within that scale and assign grey levels depending on what we are looking at. + 1000 Grey Levels 0 - 1000

  27. Windowing • Window Width (WW) = number of grey levels displayed. • Window level (WL) = mid point of the range chosen.

  28. Windowing • Soft Tissue. white WW = 200 WL = 0 +100 0 -100 Black

  29. Windowing • Lung Tissue. white WW = 700 WL = -600 -300 -600 -900 Black

  30. Windowing • Bone. white +800 450 +100 WW = 700 WL = 450 Black

  31. Windowing • If we have a wide widow width contrast is poor, e.g. we will get a low contrast image. • A narrow WW gives grater contrast, e.g. a high contrast image.

  32. Noise • Quantum mottle = primary source of noise in CR/DR. • Is the visible density fluctuations on the image. • A result of fewer photons reaching the imaging plate to form the image. • Known as image noise.

  33. Noise • System Noise. • Less of a problem than image noise. • Arises from the processing. • Effected by - phosphor conversion fluctuations, • laser beam scanning, • ADC.