Paul Scherrer. Institut. Bump-Bonded Module, 80 x 36 mm 2 , 366 x 157 Pixels. Bonded by Sylvan Streuli. SLS Detector Group. Paul Scherrer Institut. PILATUS Module. Module data Active area: 79.6 x 35.3 mm 2 (continuously sensitive) 157 x 366 = 57462 pixels

blades
Download Presentation

Paul Scherrer

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Paul Scherrer Institut Bump-Bonded Module, 80 x 36 mm2, 366 x 157 Pixels Bonded by Sylvan Streuli SLS Detector Group

  2. Paul Scherrer Institut PILATUS Module • Module data • Active area: 79.6 x 35.3 mm2 (continuously sensitive) • 157 x 366 = 57462 pixels • 16 chips (radiation hard) • Pixel size 0.217x0.217 mm2 • Readout-time: 5 ms • Energy range: Eg >4 keV • Minimum threshold 2 keV • Threshold adjust per pixel • Threshold RMS 6% of threshold • Rate: 1 MHz/pixel • 15-bit counter/pixel • single photon counting, no readout noise SLS Detector Group

  3. Paul Scherrer Institut Final Module Assembly Sensor Chips High Density Interconnect Wirebonds Module Control Board • Delicate and time-consuming manufacturing • Redesign MCB + HDI in one piece SLS Detector Group

  4. Paul Scherrer Institut Bank with 3 Modules SLS Detector Group

  5. Paul Scherrer Institut Module Fabrication Improved cleaning procedure prior to bump pressing: Ultra sonic cleaning at LMN Bank: Illumination with Fe-Source (6KeV) in the Lab: Module 2 (Sept 01) 15% Module 6 (April 02) 2.9% Module 5 (Nov 01) 14% Module 6 Total Nr of pixels: 57462 Defects digital: ~1100 (1.9%) Defects bump bonding: <330 (0.6%) Defects analog: Rest Total defects measured: 1670 (2.9%) SLS Detector Group

  6. Paul Scherrer Institut Flat-Field Correction of Pixel Detector Images Uncorrected Corrected for: • 1/r2 • Angle of incidence • Air absorption • Silicon absorption path SLS Detector Group

  7. SLS Detector Group Continuous fine f-slicing with the PILATUS-Detector Duty cycle = 96.7% Exposure = 500 ms Deadtime = 17 ms Recently improved to 6.7 ms readout time, giving >98% duty cycle.

  8. Paul Scherrer Institut Lysozyme: MOSFLM - ‘Autoindex” and ‘Predict’ SLS Detector Group

  9. Paul Scherrer Institut Lysozyme: MOSFLM - Refined Cell SLS Detector Group

  10. Paul Scherrer Institut Translation of DMILL chip into 0.25 mm-design 175 x 175 mm2 • New features: • 18 bit counter • simpler + faster Amplifier • 6 instead of 4 trimbits • Schmitt-trigger + pulsgenerator in digital part • Analog and comparator output 150 x 150 mm2 Submission: 25 April 8x8 pixels SLS Detector Group