Hemoglobin determination Biotest Hemoglobin SystemAccuracy and precision Biotest Medizintechnik GmbH
Products • Cuvette • Biotest Hemoglobin • Measuring device • Biotest Hemoglobin Tester
Cuvette Blood capillary, reagent and measuring cuvette in one. Biotest Hemoglobin Tester Photometer with integrated control and digital display Function
Function - Cuvette • Chemical reaction Haemolysis by sodium desoxycholat Sodium nitrite Sodium azid Methemo-globin Free Haemoglobin Azid- Methaemo-globin Erythrocytes
Function - Cuvette • Absorbance curve
20,00 y = 1,0064x - 0,0773 R² = 0,9992 15,00 10,00 5,00 5,0 g/dl 10,0 g/dl 15,0 g/dl 20,0 g/dl Measurements Linearity and precision, n = 20 measurements by each blood sample. Biotest Hemoglobin System, g/dl Mean value HiCN, i g/dl
Evaluation Function - Tester • Photometer; two wave lengths • Evaluation and calculation • Reading
Evaluation Self calibration • 100% green light, 100% red light and dark current • Compared to moment of calibration • If necessary automatic adjustment of LED current
Resolution Max deviation = +/- 0,75 % or +/- 30 steps Absorption 256 g/L 0 = 100 % light 470 – 880 nm 12 bits = 4000 ”steps” Resolution
Calibration Absorption 256 g/L Two points of calibration for linearity; 80 and 220 g/L Three control levels for linearity 120, 150, 180 g/L 0 = 100 % light 470 – 880 nm 12 bits = 4000 ”steps”
Function - Tester • Calculation of Hb Extinction Turbidity Turbidity 500 570 600 700 800 880 Wave length [nm]
Capillary blood Hb Time 0 min 1 min 2 min 3 min
Results - Introduction • Accuracy • How close to true value is the result • Precision • How close to each other are repeated measurements with the same sample
Results - Introduction • Accuracy and precision High accuracy Low precision Low accuracy High precision High accuracy High precision
Results - Introduction • Accuracy: • Deviation (mean) in g/L • Preferred; deviation in % • Precision: • Standard deviation SD • Preferred; Coefficient of variation, CV %
Results Precision with control blood (n=50) 30 Mean value 12,39 g/dl SD 0,084 g/dl CV 0,68 % 25 20 Numbers 15 10 5 0 12,1 12,2 12,3 12,4 12,5 12,6 12,7 Result [g/dl]
Features • Cuvette • Aimed for Hb-measurement with Biotest-Hemoglobin-Tester and compatible instruments • High accuracy and precision • Can be used for venous, capillary and arterial blood • Developed to be price worth due to latest production and control technique
Features • Tester • Automatic calibration: no calibration or test cuvette needed • Cuvette holder minimizing the risk of contamination • Automatic monitoring of power supply with fast loading of batteries (1 hour loading > 40 hours of measuring) • Automatic shut off with latest value saved • Measures only cuvettes with properly haemolysed blood
Summary • High accuracy and precision • Easy and reliable system • Long time use on battery power