An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. POTENTIOMETRY

  2. Potentiometric Analysis • Based on potential measurement of electrochemical cells without any appreciable current • The use of electrodes to measure voltages from chemical reactions

  3. Applications of Potentiometric Analysis

  4. Components of a Potentiometric Cell • Reference electrode • Salt bridge • Analyte • Indicator electrode RE SB A IE – Eref + Ej + Eind

  5. Reference electrode • Half-cell with known potential (Eref) • Left hand electrode (by convention) • Easily assembled • Rugged • Insensitive to analyte concentration • Reversible and obeys Nernst equation • Constant potential • Returns to original potential

  6. Indicator electrode • Generates a potential (Eind) that depends on analyte concentration • Selective • Rapid and reproducible response

  7. Salt bridge • Prevents mixing up of analyte components • Generates potential (Ej) = negligible

  8. Reference Electrodes • Standard Hydrogen Electrode • Calomel Reference Electrode • Silver/Silver Chloride Reference Electrode

  9. SHE • Hydrogen Gas Electrode • Pt (H2 (1 atm), H+ (1M)

  10. Saturated Calomel Electrode Hg2Cl2(s)+2e-2Hg(l)+2Cl-(aq) • Aka SCE • Easy to prepare • Easy to maintain • 0.2444 V at 25C • Dependent on temp • Toxic

  11. SCE

  12. Ag/AgCl Ref. Electrode AgAgCl (satd),KCl (satd) AgCl(s) + e-Ag(s)+Cl-(aq) E = 0.199 V

  13. Liquid Junction Potential • Liquid junction - interface between two solutions containing different electrolytes or different concentrations of the same electrolyte • A junction potential occurs at every liquid junction. • Caused by unequal mobilities of the + and - ions.

  14. Indicator Electrodes • Metallic IE • Electrodes of the First Kind • Electrodes of the Second Kind • Inert Metallic Electrodes (for Redox Systems) • Membrane IE • Glass pH IE • Glass IE for other cations • Liquid Membrane IE • Crystalline-Membrane IE • Gas Sensing Probes

  15. METALLIC INDICATOR ELECTRODES

  16. Electrodes of the First Kind • Pure metal electrode in direct equilibrium with its cation • Metal is in contact with a solution containing its cation. M+n(aq) + ne- M(s)

  17. Disadvantages of First Kind Electrodes • Not very selective • Ag+ interferes with Cu+2 • May be pH dependent • Zn and Cd dissolve in acidic solutions • Easily oxidized (deaeration required) • Non-reproducible response

  18. Electrodes of the Second Kind • Respond to anions by forming precipitates or stable complex • Examples: • Ag electrode for Cl- determination • Hg electrode for EDTA determination

  19. Inert Metallic (Redox) Electrodes • Inert conductors that respond to redox systems • Electron source or sink • An inert metal in contact with a solution containing the soluble oxidized and reduced forms of the redox half-reaction. • May not be reversible • Examples: • Pt, Au, Pd, C

  20. MEMBRANE ELECTRODES • Aka p-ion electrodes • Consist of a thin membrane separating 2 solutions of different ion concentrations • Most common: pH Glass electrode

  21. Glass pH Electrode

  22. Properties of Glass pH electrode • Potential not affected by the presence of oxidizing or reducing agents • Operates over a wide pH range • Fast response • Functions well in physiological systems • Very selective • Long lifespan

  23. Theory of the glass membrane potential • For the electrode to become operative, it must be soaked in water. • During this process, the outer surface of the membrane becomes hydrated. • When it is so, the sodium ions are exchanged for protons in the solution: • The protons are free to move and exchange with other ions. Charge is slowly carried by migration of Na+ across glass membrane Potential is determined by external [H+]

  24. Alkaline error • Exhibited at pH > 9 • Electrodes respond to H+ and alkali cations • C,D,E and F: measured value is < true value • Electrode also responds to other cations • Higher pH at lower [Na+]

  25. Acid error • Exhibited at pH < 0.5 • pH readings are higher (curves A and B) • Saturation effect with respect to H+

  26. Selectivity Coefficient • No electrode responds exclusively to one kind of ion. • The glass pH electrode is among the most selective, but it also responds to high concentration of Na+. • When an electrode used to measure ion A, also responds to ion X, the selectivity coefficient gives the relative response of the electrode to the two different species. • The smaller the selectivity coefficient, the less interference by X.

  27. Selectivity Coefficient • Measure of the response of an ISE to other ions Eb = L’ + 0.0592 log (a1 + kHBb1) • kHB = 0 means no interference • kHB 1 means there is interference • kHB < 1 means negligible interference

  28. LIQUID MEMBRANE ELECTRODES

  29. Liquid Membrane Electrodes • Potential develops across the interface between the analyte solution and a liquid ion exchanger (that bonds with analyte) • Similar to a pH electrode except that the membrane is an organic polymer saturated with a liquid ion exchanger • Used for polyvalent ions as well as some anions • Example: • Calcium dialkyl phosphate insoluble in water, but binds Ca2+ strongly

  30. 0.1 M CaCl2 Responsive to Ca2+

  31. Characteristics of Ca+2 ISE • Relatively high sensitivity • Low LOD • Working pH range: 5.5 – 11 • Relevant in studying physiological processes

  32. A K+-selective electrode • Sensitive membrane consists of valinomycin, an antibiotic

  33. CRYSTALLINE-MEMBRANE ELECTRODES

  34. Crystalline-Membrane Electrodes • Solid state electrodes • Usually ionic compound • Crushed powder, melted and formed • Sometimes doped to increase conductivity • Operation similar to glass membrane

  35. Crystalline-Membrane Electrodes • AgX membrane: Determination of X- • Ag2S membrane: Determination of S-2 • LaF3 membrane: Determination of F-

  36. F- Selective Electrode • A LaF3 is doped with EuF2. • Eu2+ has less charge than the La3+, so an anion vacancy occurs for every Eu2+. • A neighboring F- can jump into the vacancy, thereby moving the vacancy to another site. • Repetition of this process moves F- through the lattice.

  37. Fluoride Electrode

  38. GAS SENSING PROBES

  39. Gas Sensing Probes • A galvanic cell whose potential is related to the concentration of a gas in solution • Consist of RE, ISE and electrolyte solution • A thin gas-permeable membrane (PTFE) serves as a barrier between internal and analyte solutions • Allows small gas molecules to pass and dissolve into internal solution • O2, NH3/NH4+, and CO2/HCO3-/CO32-

  40. Gas Sensing Probe

  41. DIRECT POTENTIOMETRY • A rapid and convenient method of determining the activity of cations/anions

  42. Potentiometric Measurement Ionic composition of standards must be the same as that of analyte to avoid discrepancies Swamp sample and standard with inert electrolyte to keep ionic strength constant TISAB (Total Ionic Strength Adjustment Buffer) = controls ionic strength and pH of samples and standards in ISE measurements

  43. Potentiometric Measurement Calibration Method Standard Addition Method

  44. Special Applications: Potentiometric pH Measurement using Glass electrode One drop of solution Tooth cavity Sweat on skin pH inside a living cell Flowing liquid stream Acidity of stomach

  45. Potentiometric Titration Involves measurement of the potential of a suitable indicator electrode as a function of titrant volume Provides MORE RELIABLE data than the usual titration method Useful with colored/turbid solutions May be automated More time consuming

  46. Potentiometric Titration Curves