Urine Pregnancy Screening for Research Staff. Before testing sample. Date and initial kit when opening. Check product expiration, if expired discard kit. Who can perform the test?.
E N D
Before testing sample • Date and initial kit when opening. • Check product expiration, if expired discard kit.
Who can perform the test? • Following demonstration of clinical competency, research staff may perform the QuickVue One-Step hCG Urine Test. • Competency must include the performance of a known negative and positive sample. • Competency must be documented annually for employees.
Quality Control: • Perform commercial quality control at each location during monthly maintenance. • These results are documented on the QuickVue One-Step hCG QC log sheet.
Procedure: • NOTE: Do not use if foil package appears to have been opened prior to testing. • Label the Test Device with patient History number or Subjects Study ID Number. • Draw up urine into the disposable dropper and dispense 3 drops into the large round Sample well. • Read result at 3 minutes. Note: Some positive results may appear sooner.
Interpretation of Results: Negative Result: • Control: • A blue line appears in the small square labeled “Control Window”. • Document the control reaction by recording “yes”. • Patient: • There is no change in the in the large square labeled “Read Result Window”. The blue line is the only line present. • Document the reaction by using either “NEG” or as a (–), minus sign. • Record in the patient result column as “NEG”.
Interpretation of Results: Positive Result: • Control: • A blue line appears in the small square labeled “Control Window”. • Document the control reaction by recording “yes”. • Patient: • A pink line appears below the blue line in the large square labeled “Read Result Window”. • Document the reaction by using “POS” or as a (+), plus sign. • Record in patient result column as “POS”. NOTE: ANY shade of a pink line in the Read Result Window should be interpreted as a positive result.
Interpretation of Results: Invalid Result: • If a vertical blue line DOES NOT appear in the small square labeled “Control Window”. • Document the control reaction by using a “Invalid”. The test should be repeated using a new QuickVueOne-Step hCG Test.
Reporting Results • Most errors occur when recording the results incorrectly! • Fill in date, time, patient’s name and history number. • Record the reactions of both the control and patient test window on the QC Log Sheet. • Write “POS” or “NEG” as the interpretation of the patient’s test result.
Completion of Training: • Trainer must observe your performance of commercial controls with a known positive and negative hCG test before you can perform patient screening.
Standard Safe Work Practices • No food, drink allowed in the screening area • Area must have place for hand-washing • Area must have Biohazard disposal • Work area must be maintained in a clean and sanitary condition • Procedures involving body fluids are to be performed in a manner to minimize splashing, spraying, spattering, and droplet generation
Approved Trainers • Ana Garcia-Turner, RPM Ophthalmology • Barbara Kurth, RPM Radiology • Bonnie Thiele, RPM, OB/GYN • Catee Mullen, RPM Heart Center • Deborah Hannah, RPM Dermatology • Eang King, RPM CFM • Holly Tiemann, DOCR • Sharon Minda, RPM Psychiatry • Terry Ainsworth, DOCR