1 / 8

Resident Physics Lectures

Christensen, Chapter 7 X-Ray Beam Restrictors. Resident Physics Lectures. George David Associate Professor Department of Radiology Medical College of Georgia. Aperture Diaphragms Cones and Cylinders Collimators. Restrictors Types. Lead sheet with hole in center simple

bperry
Download Presentation

Resident Physics Lectures

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Christensen, Chapter 7 X-Ray Beam Restrictors Resident Physics Lectures George David Associate Professor Department of Radiology Medical College of Georgia

  2. Aperture Diaphragms • Cones and Cylinders • Collimators Restrictors Types

  3. Lead sheet with hole in center • simple • large penumbra (partial shadow) • Lead close to focal spot Focal Spot Aperture Diaphragms Penumbra Receptor

  4. Focal Spot Focal Spot Film Film Cones & Cylinders With cone angle as above, this functions as an aperture Reduced Penumbra

  5. Advantages • Adjustable rectangular field • fluoro may have circular field • Light beam indicates x-ray field • Two sets of shutters • longitudinal • transverse • Each set may have 2 pairs of shutters operating together Collimators One set; two pairs

  6. Focal Spot Filter Lamp Mirror Shutters (only 1 set shown) • Generally include filter & light Radiographic Collimators

  7. Numerical field size indicators • Field size dependent on source-image distance • SID • Focal spot to image receptor 8 10 12 14 17 40” 8 10 12 14 17 Collimator Features

  8. Senses cassette size, orientation, & SID • source - image distance • limits x-ray field to size of receptor • PBL does not insure tube & bucky aligned Positive Beam Limitation(PBL or Automatic Collimation) Cassette

More Related