1 / 24

An Automatic Instrument to Measure the Absolute Components of the Earth's Magnetic Field

H.-U. Auster, M. Mandea, A. Hemshorn , E. Pulz, M. Korte. An Automatic Instrument to Measure the Absolute Components of the Earth's Magnetic Field. Outline. Fundamentals of the Method Magnetic field along a rotation axis Elimination of systematic measurement errors

brooks
Download Presentation

An Automatic Instrument to Measure the Absolute Components of the Earth's Magnetic Field

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. H.-U. Auster, M. Mandea, A. Hemshorn, E. Pulz, M. Korte An Automatic Instrument to Measure the Absolute Components of the Earth's Magnetic Field

  2. Outline • Fundamentals of the Method • Magnetic field along a rotation axis • Elimination of systematic measurement errors • Manually performed test of the method in Niemegk • Automation • Magnetometers • Mechanics & Optics • Controlling • Set up in Belsk • Outlook

  3. Motivation Manually performed absolute measurement in Hermanus by rotation of a fluxgate magnetometer about two well defined axes Ongoing activities to automate absolute measurement • Automating of DI-Flux • Manipulation of vector proton magnetometer • Automating of rotation of a vector fluxgate magnetometer about two well defined axes Auster H.U., V.Auster, A new method for performing an absolute measurement of the geomagnetic field, Meas. Sci. Technol. 14, 1013-1017, 2003

  4. Magnetic field along a rotation axis (1) Co-ordinate systems • Red: arbitrary oriented fluxgate magnetometer • Black: Geophysical reference system • Relation between both: Euler angles Rotation about Precession angle ivariable •  and  constant • Bz = Bz(υ,φ)

  5. Magnetic field along a rotation axis (2) Computation of field in rotation axis • Three independent measurements with arbitrary I necessary • Magnitude of B in direction of rotation axis becomes independent from sensor orientation angles  and  Matrix MB of measurement results Unit vector of sensor orientation

  6. Measurement Procedure • Rotation A to adjust mechanical axis to azimuth marks • Rotation B to turn the sensor about the mechanical axes • Always 360° forward and backward • 6 measurement stops each rotation direction (B and azimuth) • Rotation C for magnetometer calibration y y x x  Measurement time: 30 minutes

  7. Elimination of systematic measurement errors • Magnetometer errors by scalar calibration • Rotation about two axes sufficient for full determination of linear transfer function (offsets, scale factors, non orthogonality) • Full Earth field magnetometer necessary, high requirements on linearity (10-5) • Orientation of rotation axes • Horizontal balance by level tube, misalignment of level tube eliminated by interchanging of its ends • Azimuth by telescope, misalignment of optical axis and rotation axis eliminated by rotation of telescope

  8. Measurement Results of one year operation in Niemegk - Z Standard deviation: 1.0nT 2005 2006

  9. Measurement Results of one year operation in Niemegk - D Standard deviation after trend and readjustment correction: 0.6nT 2005 2006

  10. Measurement Results of one year operation in Niemegk - Magnetometer

  11. Steps to automation • Magnetometers • Optical control • Mechanics: 3 Rotations • Rotation about measurement axis - arbitrary angles • Turn Table - arbitrary, but well known angle • Sensor rotation - arbitrary angle • Controlling • Hardware • Software

  12. Magnetometers basket magnetometer • two digital 3-axes fluxgate magnetometers • range: 64000nT • resolution: 0.01nT • Non linearity < 10-5 • Serial & Flashcard interface • Proton-Magnetometer • Range: 20000-64000nT • resolution 0.01nT • Serial & Flashcard interface variometer scalar magnetometer

  13. Performance of Magnetometers Variometer Basket magnetometer Observatory Data

  14. Performance of Magnetometers Variometer - Obs Basket - Obs Variometer - Basket

  15. Components of the optical system • Neodym Laser • Coupled in by fibre optics • PSD 2cm x 2cm • Resolution < 0.1 mm • Protected from stray light by • Band-filter (635nm) • Black tube • Azimuth Mark • Made by ceramics • Grounded in concrete Bild PSD

  16. ~ 15 m 0.2mm  = atan(0.2mm/15 m) ~ 3'' Performance of optical system • Stability of azimuth mark: • Quarzgut 0.5 ppm/°C • Displacement < 0.1 mm (h=2m, T = 50°C)

  17. Rotation A by Pneumatics • Pointing requirements: • 1cm/20m • 0.1mm/20cm • 2arcmin • Well defined end positions necessary • 7kg has to be rotated • Low friction by bearing • Importance of surface treatment • Pressure supply necessary (2 bar)

  18. Rotation B & C by Piezo Motors Attempts to develop non magnetic motors for sensor rotation (1) Application example (2) Linear motor (3) Rotation to linear conversation (4) Final solution using gravitation

  19. Control Unit • GPS controlled Timing • Motor control • rotation about measurement axis by piezo motor • Flip mechanism by piezo motor • Pneumatic control • Turn table rotation by two valves • Laser switching and PSD read out • Magnetometer control • Pre processing of data

  20. Magnetic Requirements for Accommodation

  21. Accommodation in Belsk

  22. Outlook • System has to run permanently in Niemegk • Reliability have to be tested and improved • Redesign of laser optics and some mechanical parts • Option: Replacement of pneumatics by Piezo motor driven system • New design for lower latitudes

  23. Acknowledgement • GFZ for personal and financial investigation • All the the people designed and manufactured the mechanics (in Niemegk Potsdam Braunschweig Lindau and Garching) • Magson for magnetometers and software support • Belsk observatory for support to install the facility

  24. Measurement Results of one year operation in Niemegk - H Standard deviation: 1.4nT