250 likes | 384 Views
LISST-SL plots. R0681302. Pump takes ~2 mins to lock in Data no good. R1091106. No apparent issue with pump voltage. Apart from the 2 high conc values and open-ended PSD data looks good. R1101437. Pump not locked on for the first minute; possibly lock-on after 1 min?
E N D
R0681302 Pump takes ~2 mins to lock in Data no good
R1091106 No apparent issue with pump voltage. Apart from the 2 high conc values and open-ended PSD data looks good
R1101437 Pump not locked on for the first minute; possibly lock-on after 1 min? Upcast probably OK. Note constant tau, but variation in MC. PSD looks fine.
R1101500 Pump not locked on. Data not OK.
R1120717 Velocity very low, so pump shuts off then starts up again.
R1120727 Pitot tube contamination, not reliable data.
R1121523 Pump resetting. Data not OK.
R1121618 Pump resetting after ~1 minute. First minute may be OK..
R1160408 Pump seems to track towards the end, last ~10 data points should be OK.
R1301058 Pump never locked in. Data no good (velocity, depth OK). Why max velocity in the middle?
R1301440 Pump never locked in. Data no good (velocity, depth OK).
R1301447 Excellent pump lock-in and tracking of velocity. Good PSD, good concentration.
R1311450 Excellent pump lock-in and tracking of velocity.
R1311507 Good pump lock-in for the last half.