MAGNETOMETER CALIBRATION

Ingo Richter Institut für Geophysik und Meteorologie TU Braunschweig. MAGNETOMETER CALIBRATION. Magnetsrode - MCF. Magnetsrode – Characteristics. Compensation: Dynamic v fixed (user defined) Fields - Range: -100000 nT ... +100000 nT Direction: any, 3 components

burian
Download Presentation

MAGNETOMETER CALIBRATION

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Ingo Richter Institut für Geophysik und Meteorologie TU Braunschweig MAGNETOMETER CALIBRATION

  2. Magnetsrode - MCF

  3. Magnetsrode – Characteristics • Compensation: Dynamic v fixed (user defined) • Fields - Range: -100000 nT ... +100000 nT • Direction: any, 3 components • Type : DC, AC, Arbitrary • Field-Sequence: arbitrary, user defined • Accuracy: < 0.8 nT • I/F to DUT: Analog (16 Bit) v Digital (EGSE) • DUT TC: yes, via RS232 (user defined) • Successful calibrations have been performed for: • CLUSTER, DS 1, MARS 96, ØRSTED,ROSETTA, • CASSINI, CHAMP, ROMAP, DOUBLE STAR,....

  4. Parameters A magnetometer can be characterized by the following entities: _ _ • Sensitivity S = {Sii } ,Sii= Sii (T ) • MisalignmentM = {Mij } ,Mij = Mij (T ) • Offset BO = { Box(T ), Boy(T ), Boz(T ) } • Frequency Response _ _

  5. Sensor Model B c = F-1 Bm _ _ B c ={R-1M-1 S-1}( B r -Boff - Bres) _ _ _ _ _ _

  6. Calibration Method • Offset: • Perform 2 measurements. One at 0° orientation and one at 180° • orientation. Boff = (B1 +B2) /2 . Bres = (B1-B2) /2 • Sensitivity & Misalignment: • Two methods: • a) Pure linear steps in subsequently x,yz direction • b) Spike spheres: apply randomly distributed B-Field on sphere • Generate least-square fit model to calculate S,M. • Possible models : up to 5th order polynome in x,y,z (mixed) • Frequency response: • Apply sine-signals at different discrete frequencies at constant amplitude. Calculate response function

  7. Parameters - Offset

  8. Parameters – Offset(T): Plot These parameters have been measured using thermocycles. OB & IB were placed at off-center positions in the thermal test box.

  9. Sensitivity (overview) (FS / FM-IB) Reference Measurements: T1 = 17.39°C, CoC : (Spike sphere) Sensitivities obtained from the temperature calibration at the off-center position: (linearity measurements at varying Temperature) Wanted: Temperature dependent Sensitivity at CoC: with The temperature dependent sensitivity at CoC can be derived from the off-center measurements using a geometry coefficient k

  10. Sensitivity (Results) Geometrical correction factor

  11. Parameters – S(T): Plot These parameters have been measured using thermocycles. OB & IB were placed at off-center positions in the thermal test box.

  12. 1 Misalignment (overview) Reference Measurements: T1 = 17.39°C, CoC : (Spike sphere) Misalignment angles obtained from the temperature calibration at the off-center position: (linearity measurements at varying Temperature) The off-center Misalignment Matrix is given by: Wanted: Temperature dependent Misalignment Matrix at CoC. It can be derived from the off-center measurements using a geometry correction Matrix K

  13. Misalignment (Results) The correction matrix is given by: Finally we get the desired Misalignment Matrix : and

  14. Alignment Angles (T) : Plot These angels have been measured using thermocycles. OB & IB were placed at off-center positions in the thermal test box. The plots show the deviation from The nominal value of 90° 00´.

  15. Frequency Response

  16. Documentation • Report: RO-IGM-TR-0002 • DC-Analysis: RO-IGM-TR-0003 • AC-Analysis: RO-IWF-TR-0001