A). When the carrier gas is quickly switched to 100% nitrous oxide

A). When the carrier gas is quickly switched to 100% nitrous oxide (B). the halothane concentration decreases to 3% within 8 to 10 seconds A new steady-state concentration of approximately 3.5% is then attained within 1 minute. Contributor Factors:. Viscosity Density

camila
Download Presentation

A). When the carrier gas is quickly switched to 100% nitrous oxide

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A). When the carrier gas is quickly switched to 100% nitrous oxide (B). the halothane concentration decreases to 3% within 8 to 10 seconds A new steady-state concentration of approximately 3.5% is then attained within 1 minute

  2. Contributor Factors: • Viscosity • Density • relative solubility of the carrier gas • the flow-splitting characteristics of the specific vaporizer • the concentration control dial setting

  3. Safety Features: • Agent-specific, keyed filling devices • Overfilling is minimized • Firmly secured to a vaporizer manifold (Prevent vaporizer tipping ) • Contemporary interlock systems (prevent the administration of more than one inhaled anesthetic) • Hazards: • Misfilling • Contamination • Tipping (transport “T” dial setting) • Overfilling • Underfilling (sevoflurane) • Simultaneous Administration of Inhaled Anesthetics • Leaks • Environmental Considerations (MRI)

  4. Desflurane Boils at 22.8°C MAC: 6% to 7% (amount of desflurane vaporized over a given period is considerably greater than other anesthetic drugs that cause excessive cooling of the vaporizer and significantly reduce its output) Vapor pressures at 20°C enflurane 172, isoflurane 240, halothane 244, and desflurane 669 mm Hg At 1 atm and 20°C, 100 mL/min passing through the vaporizing chamber would entrain 735 mL/min of desflurane versus 29, 46, and 47 mL/min of enflurane, isoflurane, and halothane ((to produce a 1% desflurane output)) bypass flow would be approximately 73 L/min versus 5 L/min or less for the other three anesthetics

  5. Operating Principles of the Tec 6 and Tec 6 Plus Simplified schematic of the Tec 6 desflurane vaporizer

  6. The Datex-OhmedaAladin Cassette Vaporizer The heart of the vaporizer is the electronically controlled flow control valve located in the outlet of the vaporizing chamber

  7. Anesthetic Breathing Circuits Mapleson Systems Circle Breathing Systems

  8. Mapleson Systems prevention of rebreathing: • spontaneous ventilation: A > DFE > CB • controlled ventilation: DFE > BC > A

  9. Bain Circuit • The integrity of the inner tube can be assessed as described by Pethick

  10. The Traditional Circle Breathing System • The most popular breathing system in the United States • “universal F” or “single-limb circuit,” (One version of the traditional circle system), has increased in popularity in recent years • A circle system depending on the amount of fresh gas inflow can be: Semi open, semi closed, or closed • A semi open (no rebreathing), requires a very high flow of fresh gas. • A semi closed system (some rebreathing of exhaled gases), is the most common in the United States.

  11. seven primary components of circle system • (1) a fresh gas inflow source • (2) inspiratory and expiratory unidirectional valves • (3) inspiratory and expiratory corrugated tubes • (4) a Y-piece connector • (5) an overflow or pop-off valve referred to as the APL valve • (6) a reservoir bag • (7) a canister containing a carbon dioxide absorbent.

  12. Three rules must be followed • (1) a unidirectional valve must be located between the patient and the reservoir bag on both the inspiratory and expiratory limbs • (2) the fresh gas inflow cannot enter the circuit between the expiratory valve and the patient • (3) the overflow (pop-off) valve cannot be located between the patient and the inspiratory valve.

  13. The main advantages of the circle system • (1) maintenance of relatively stable inspired gas concentrations • (2) conservation of respiratory moisture and heat • (3) prevention of operating room pollution.

  14. Disadvantages • Complex design • In a recent closed-claim analysis: more than a third (25/72) of malpractice claims resulted from breathing circuit misconnections or disconnections. • valves stick: Rebreathing, barotrauma or volutrauma • Obstructed filters located in the expiratory limb: have caused increased airway pressure, hemodynamic collapse, and bilateral tension pneumothorax