An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. S4 Environmental Disturbances Robert Schofield, U of O John Worden, Richard McCarthy, Doug Cook, Hugh Radkins, LHO Josh Dalrymple, SU • Pre-S4 “fixes” • Some S4 veto issues • Early coupling results from S4 PEM injections

  2. Up-conversion of low frequency seismic noise seismic noise from distant excavation night spectrum excites stack modes in AS_Q up-conversion reduces interferometer sensitivity night spectrum

  3. Optic motion, by itself, does not produce up-conversion. shaking ground at 1.2 Hz moves optic and produces up-conversion larger optic motion from actuator injection does not result in up-conversion

  4. Up-conversion depends on stack motion not floor motion. Small shaker on cross-beam and large shaker on floor adjusted to produce similar optic motion... ...produce similar up-conversion even though floor motion differs by 10.

  5. Ground shaking to excite specific motions Red: mainly beam-axis; Blue: mainly side-to side; Black: no shaking Beam-axis (AS_Q) Yaw (from optical lever) Pitch (optical lever) Side (from shadow sensor)

  6. Side-to-side shaking does not produce up-conversion. Ground shaking that producesbeam-axis motionproduces up-conversion, but shaking that produces side-to-side motion does not.

  7. Up-conversion for different shaking amplitudes Small fast-traveling seismic transient

  8. Simple model doesn’t work at all frequencies Predicted cutoff ~52 Hz Predicted cutoff ~12 Hz 2.1 Hz prediction ~ agrees, but 1.2 doesn’t

  9. Possible sources of up-conversion brushing cables on test mass support structure output telescope: consistent with lack of up-conversion for side-to side

  10. Up-conversion with injection to mimic pitch during wind Optical lever during periods of wind, quiet and actuator injections to “match” wind. 70-100 Hz band in AS_Q showing up-conversion during wind and actuator injection.

  11. Wind-simulating injection produced RF saturation Small fast-traveling seismic transient

  12. Wind likely produced saturation and “mystery” up-conversion. Black: wind; Red: ETMY 1.2 Hz pitch injection; Blue: quiet Up-conversion from wind differs from up-conversion from injection; wind still produced up-conversion at lower power.

  13. Up-conversion at different locations, similar velocities LHO ITMY (due to MICH) LHO ETMX Brian’s LLO ETMX measurement

  14. S4 H1-H2 coherence peak identification for Stochastic group 133 Hz: Neslab PSL chillers 330 Hz HVAC turbine SF01

  15. Pulsars in VME crates magnetometer near VME crate detects pulsar candidate frequencies. Magnetometers elsewhere don’t. Red: normal operation of VME crate Blue: VME crate on, but 7751 CPU board disconnected Black: 7751 CPU on but code not running

  16. Coherence between AS_Q and VME magnetometer 108 Hz peak magnetometer Broad-band coherence

  17. 108 Hz peak moves with air-flow to fans Blue: normal air; Red: extra air

  18. 108 Hz peak goes away when power supply fan disconnected.

  19. Summary • Pre-S4 partial fixes • a. Wandering chiller seismic peak (D.Q. flag for 16 Hz?) • b. 10 Hz from RFID • c. Transformers near test masses (ion pump supplies?) • d. Crab protection (LHO out-stations, LLO?) • Some S4 Issues • a. Up-conversion of low frequency “continuous” seismic noise • b. Veto up-converting 0.7-2 Hz seismic transients • c. Continuous environmental sources can produce AS_Q bursts • Preliminary coupling results from S4 PEM injections • a. LHO LVEA ambient sound level generally less than 1/5 SRD • above 60 Hz • b. LHO out-station ambient sound level generally less than 1/10 SRD • c. 60 Hz peak in AS_Q may be dominated by direct coupling of ambient • magnetic fields