1 / 14

Strofio Calibration & Lunar ENAs measured by IBEX

Strofio Calibration & Lunar ENAs measured by IBEX. Frédéric Allegrini Southwest Research Institute fallegrini@swri.org (210) 522-6029. Parameter space 2 angles: elevation, azimuth Energy (velocity) Neutral/ion 30 voltages Constraint Limited calibration time. Overview.

casper
Download Presentation

Strofio Calibration & Lunar ENAs measured by IBEX

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Strofio Calibration &Lunar ENAs measured by IBEX Frédéric Allegrini Southwest Research Institute fallegrini@swri.org (210) 522-6029

  2. Parameter space 2 angles: elevation, azimuth Energy (velocity) Neutral/ion 30 voltages Constraint Limited calibration time Overview

  3. Key measurements • Voltages needed • Sensitivity: > 0.1 (counts/s)/(particles/cm3) for H2O (corresponds to 0.1A/torr ion yield and 6% transmission) [ERD_1049] • Mass resolution: m/∆m ≥ 60 [ERD_1035] • Mass range: H (1 amu) through Fe (56 amu) [ERD_1028,ERD_1030] • FOV ±10°, KOZ ±20° [ERD_1061,ERD_1059] • Velocity filter: Energy range: 0-10 eV [ERD_1784,ERD_1785] • Additional measurements: • UV attenuation and light leaks [SEN_614 , SEN_615 ] • MCP Background noise • Maximum rate of 104cts/s [ERD_ 1066]

  4. Strofio Calibration Field of view (part 1)

  5. Field-Of-View & Keep-Out-Zone KOZ FOV

  6. Casymir(Calibration System for the Mass Spectrometer Instrument ROSINA) • Beam line • Pressure ~10-9 mbar with beam • Species: Ar, Ne, H2, He, N2, … • Velocities 1-3 km/s • Size 2-10 mm dia • Strofio dedicated chamber • Pressure ~10-10 mbar • LN2 cold plate • Up to 80°C heating system • X and Z movement via external table system • ε rotation via manual actuator • Measurement ready after 2 days of pumping

  7. Geometrical Characteristics Beam • Determined by mechanical scanning the instrument: • X with table • ε with internal mechanism • Z with table (in-out of the page) ε x Z ○

  8. Field of View – No Velocity Filter

  9. Strofio Calibration Field of view (part 2)

  10. Epsilon Scan under Large Background Rejection • Determines the geometry response of the sensor • Very irregular “two horns” shape Z [mm] Ar/H2O Ar/Max(Ar) Epsilon [deg] S1A-S1B [V]

  11. Simulations vs measurements Simulations Measurements Z [mm] Z [mm] Epsilon [deg] Epsilon [deg] Testing with Strofio’s “little sister” planned for summer 2013

  12. Strofio Calibration Uv contamination

  13. UV leak test • First test performed at SwRI • ~1 solar intensity near Lyman alpha • UV leak(s) created signal slightly above background levels • Need to perform additional tests in Bern

  14. Tests in Bern Start left rate [Hz] • Illumination of the sensor over almost the whole sphere. Data points at 1 minute integration • Reported here the two scans that raised previously some questions • Intensity = 1 sun in Lyman alpha • No sensitivity detected at the 1 cts/second level • UV through Strofio aperture also tested Start right rate [Hz] Start left rate [Hz] Start right rate [Hz]