An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. CEN/ISO/IEEE 11073 Standards for point-of-care device communication: Overview SCATA, London, 5th May, 2005 Melvin Reynolds, AMS Consulting, Ross-on-Wye, UKChair CEN TC251 WGIV Health informatics – Technologies for interoperability, Chair ISO TC215 WG2 Health informatics – Messages and communications; Past-chair ISO TC215 WG2.1 Health informatics – Device communications, Co-chair IEEE1073 Medical device communications. with thanks to: Thomas Norgall, Fraunhofer Institute, Erlangen, andTodd Cooper, Breakthrough Solutions, San Diego

  2. CEN/ISO/IEEE 11073 Standards Overview • Requirements • Interoperability • History • The (content) standards • Conclusions • Discussion

  3. CEN/ISO/IEEE 11073 Standards Point-of-care: the need About 20% of an ICU nurse’s time can be redeployed if charting is automated, eliminating transcription error & false recording; as well a providing more rapid feedback. Different benefits apply in other care situations.

  4. Data Processing System / PDMS CEN/ISO/IEEE 11073 Standards Point-of-care: all change Fixed configuration, e.g. anaesthesia Syringe Pumps and Infusors Patient Monitor Respirator Additional Pump – but changing periodically, and evolving over time Additional Monitor

  5. CEN/ISO/IEEE 11073 Standards Point-of-care: all change Variable configuration, e.g. intensive care Data Processing System / PDMS Many Syringe Pumps and Infusors Respirator Patient Monitor - changing frequently and within minutes

  6. CEN/ISO/IEEE 11073 Standards Interoperability Interoperability: ability of two or more systems or components to exchange information and to use the information that has been exchanged source: IEEE Standard Computer Dictionary: A Compilation of IEEE Standard Computer Glossaries, IEEE, 1990 Functional interoperability: Shared Architectures, Methods & Frameworks Semantic interoperability: Shared Data types, Terminologies, Codings

  7. CEN/ISO/IEEE 11073 Standards Interoperability Point-of-care medical device communication (IEEE1073) Charter: • Provide real-time plug-and-play interoperability for patient-connected medical devices • Facilitate the efficient exchange of vital signs and medical device data, acquired at the point-of-care, in all health care environments … Leveraging off-the-shelf technologies, scaling across a wide range of system complexities, and supporting commercially viable implementations.

  8. CEN/ISO/IEEE 11073 Standards Context Health informatics: Generally interpreted as being in ISO/OSI Level 7 Medical device communication: Works at all ISO/OSI Levels, Directly in ISO/OSI Levels 7 – 5 In Levels 1 – 4, chiefly by reference to other standards (such as IrDA or 802.11)

  9. CEN/ISO/IEEE 11073 Standards Architecture Requirements: • True interoperability across all 7-layers - from the ‘connector’ to the end application • Mechanisms to support the strong quality of service requirements placed on regulated medical devices • Maintainability as communications technology and applications change

  10. CEN/ISO/IEEE 11073 Standards Series title (National, e.g. BS) EN ISO/IEEE 11073-xyyzz Health informatics - Point-of-care medical device communication - x partition (general): yy subsection (group) - zz detail (specialised)

  11. Data & Information Definitions 11073.1xxxx 11073.2xxxx Application Profiles Transport & Physical Layers 11073.3xxxx Internetworking Support 11073.5xxxx Application Gateways 11073.6xxxx Related – some shared concepts 11073.9xxxx CEN/ISO/IEEE 11073 Standards Series structure

  12. CEN/ISO/IEEE 11073 Standards 11073-1xxxx content 11073.1xxxx Data & Information Definitions Semantics needed to communicate a device’sapplication data, status and control information. Three main components: • Nomenclature: 11073.10101 • Domain Information Model (DIM): 11073.10201 • Device Specialisations: 11073.103xx

  13. CEN/ISO/IEEE 11073 Standards 11073-10101 content Nomenclature: A set of numeric codes that identify every item that is communicated between systems.

  14. CEN/ISO/IEEE 11073 Standards 11073-10201 content Domain Information Model: An object oriented data model that specifies objects, attributes, attribute groups, event reports, and services that may be used to communicate device data and to control / configure the reporting of information . . . • Medical Devices and Functionalities • Measured Data and Settings • Alert Information • Remote Control • Patient Information • Communication

  15. CEN/ISO/IEEE 11073 Standards Domain Information Model

  16. ‘On-the-Wire’ Format Abstract Syntax (ASN.1) Transfer Syntax (MDER) CEN/ISO/IEEE 11073 Standards Development Process Domain Information Model

  17. CEN/ISO/IEEE 11073 Standards 11073-6xxxx content 11073.6xxxx Application Gateways Objectives: Interoperability between different application-layer protocols + Bi-directional transparency 11073 11073.3xxxx 11073.3 /.5xxxx System non-VITAL 11073.6xxxx Gateway Application (Agent/ Transport Transport Manager) Example: HL7 / 11073 Gateway: 11073.60101 HL7 - Observation Reporting Interface

  18. CEN/ISO/IEEE 11073 Standards 11073-60101 content HL7 - Observation Reporting Interface Objective: To ensure that … information … from devices using … 11073 standards is conveyed with appropriate … identification of source device system, medical device, channel and metric (… with required annotation …). For traceability, and to enable safe computation from data, the 11073… nomenclature is retained … payload of the HL7 message/document that carries the information to the target information system.

  19. CEN/ISO/IEEE 11073 Standards 11073-60101 HL7 ORI Example HL7 v2.6 message:MSH|||^~\&|Vendor_name||NUR||200312021203||ORU^R01|02110212073000458038 |P|2.3| ||||||PID|||999-99-9999 ||Smith^John^L |||||||||||||999-99-9999 ||||||||||||PV1|||CU1^^BED1||||^^^|||||||||||||||||||||||||||||||||||||||||||||OBR|||||||20031202120235|||||||||||||||||||||||||||||||||||||OBX||SN|147842^MDC_ECG_HEART_RATE^MDC^^HR^local||=^83|264864^MDC_DIM_BEAT_PER_MIN^MDC^^beat/min^local|||||R|||20031202120235|||OBX||SN|134276^MDC_EVT_ECG_V_P_C^MDC^^PVC rate^local||=^0|264864^MDC_DIM_BEAT_PER_MIN^MDC^^beat/min^local|||||R|||20031202120235|||OBX||SN|131841^MDC_ECG_AMPL_ST_LEAD_I^MDC^^ST elev lead I^local||=^0.01|266418^MDC_DIM_X_VOLT^MDC^^mV^local|||||R|||20031202120235|||OBX||SN|131842^MDC_ECG_AMPL_ST LEAD_II^MDC^^ST elev lead II^local||=^0.01|266418^MDC_DIM_X_VOLT^MDC^^mV^local|||||R|||20031202120235|||OBX||SN|131901^MDC_ECG_AMPL_ST_LEAD_III^MDC^^ST elev lead III^local||=^0.01|266418^MDC_DIM_X_VOLT^MDC^^mV^local|||||R|||20031202120235|||OBX||SN|150324^MDC_SAT_O2_ART^MDC^^SPO2^local||=^97|262688^MDC_DIM_PERCENT^MDC^^%sat^local|||||R|||20031202120235|||||460230^MDC_UPEXT_FINGER^MDC^Finger NOS^local^^1OBX||SN|149538^MDC_PLETH_PULS_RATE^MDC^^Pulse rate^local||=^83|264896^MDC_DIM_PULS_PER_MIN^MDC^^Pulse/min^local|||||R|||20031202120235|||||460230^MDC_UPEXT_FINGER^MDC^Finger NOS^local^^1OBX||ST|150302^MDC_PRESS_CUFF_DIA^MDC^^NIBPdia^local||=^80|266016^MDC_DIM_MMHG^MDC^^mmHg^local|||||R|||20031202120030|||||458572^MDC_UPEXT_ARM_UPPER^MDC^Upper arm^local^^1OBX||ST|150303^MDC_PRESS_CUFF_MEAN^MDC^^NIBP mean^local||=^93|266016^MDC_DIM_MMHG^MDC^^mmHg^local|||||R|||20031202120030|||||458572^MDC_UPEXT_ARM_UPPER^MDC^Upper arm^local^^1OBX||ST|150301^MDC_PRESS_CUFF_SYS^MDC^^NIBP sys^local||=^120|266016^MDC_DIM_MMHG^MDC^^mmHg^local|||||R|||20031202120030|||||458572^MDC_UPEXT_ARM_UPPER^MDC^Upper arm^local^^1OBX||ST|149546^MDC_PULS_RATE_NON_INV^MDC^^Pulse rate^local||=^83|264896^MDC_DIM_PULS_PER_MIN^MDC^^Pulse/min^local|||||R|||20031202120030|||||458572^MDC_UPEXT_ARM_UPPER^MDC^Upper arm^local^^1

  20. CEN/ISO/IEEE 11073 Standards 11073-60101 HL7 ORI Example HL7 v2.6 message 11073 payload: OBX||SN|147842^MDC_ECG_HEART_RATE^MDC^^HR^local||=^83|264864^MDC_DIM_BEAT_PER_MIN^MDC^^beat/min^local|||||R|||20031202120235|||||OBX||SN|134276^MDC_EVT_ECG_V_P_C^MDC^^PVC rate^local||=^0| 264864^MDC_DIM_BEAT_PER_MIN^MDC^^beat/min^local|||||R|||20031202120235|||||OBX||SN|131841^MDC_ECG_AMPL_ST_LEAD_I^MDC^^ST elev lead I^local||=^0.01|266418^MDC_DIM_X_VOLT^MDC^^mV^local|||||R|||20031202120235|||||… OBX||SN|150324^MDC_SAT_O2_ART^MDC^^SPO2^local||=^97| 262688^MDC_DIM_PERCENT^MDC ^^%sat^local|||||R|||20031202120235|||||460230^MDC_UPEXT_FINGER^MDC^Finger NOS^local^^1OBX||SN|149538^MDC_PLETH_PULS_RATE^MDC^^Pulse rate^local||=^83|264896^MDC_DIM_PULS_PER_MIN^MDC^^Pulse/min^local|||||R|||20031202120235|||| |460230^MDC_UPEXT_FINGER^MDC^Finger NOS^local^^1… OBX||ST|150301^MDC_PRESS_CUFF_SYS^MDC^^NIBP sys^local||=^120|266016^MDC_DIM_MMHG^MDC^^mmHg^local|||||R|||20031202120030|||||458572^MDC_UPEXT_ARM_UPPER^MDC^Upper arm^local^^1OBX||ST|149546^MDC_PULS_RATE_NON_INV^MDC^^Pulse rate^local||=^83|264896^MDC_DIM_PULS_PER_MIN^MDC^^Pulse/min^local|||||R|||20031202120030|||| |458572^MDC_UPEXT_ARM_UPPER^MDC^Upper arm^local^^1

  21. CEN/ISO/IEEE 11073 Standards Conclusions • 11073 is a comprehensive system of point-of-care medical device communication standards • 11073 device types range from real-time-operating medical equipment to point-of-care test • 11073 supports wired, wireless IR (and hopefully future wireless RF) network technologies • 11073 provides plug-and-play, internetworking and application gateway capabilities

  22. CEN/ISO/IEEE 11073 Standards Conclusions • 11073 is used to a greater or lesser extent by the biggest companies in the business – often not declared! • 11073 provides an open way of accessing: device networks data for real time decision support • 11073 is being requested by health care providers & management organisations, e.g NHS NPfIT, US DHHS, as they seek point-of-care to record transparency of information

  23. CEN/ISO/IEEE 11073 Standards Contact details Thank you for your attention. Melvin ReynoldsAMS Consulting Ashcote, Walford Road, Ross-on-Wye, HR9 5PQ, UKmelvinr@ams-consulting.co.ukwww.ieee1073.org