210 likes | 549 Views
Visible Spectrophotometer. Visible Spectrophotometer. Visible Spectrophotometer. 100 % Adjust. I 0. I. white light. detector/recorder. 100 % adjust blank distilled water. prism. I 0 = I % transmittance = (I / I 0 ) x 100 = 100 absorbance = 0. Visible Spectrophotometer.
E N D
Visible Spectrophotometer 100 % Adjust I0 I white light detector/recorder 100 % adjust blank distilled water prism I0 = I % transmittance = (I/ I0) x 100 = 100 absorbance = 0
Visible Spectrophotometer Zero Adjust I0 I white light detector/recorder zero adjust (barrier inserted) prism I = 0 % transmittance = 0 absorbance = maximum
Visible Spectrophotometer Sample I0 I white light detector/recorder sample solution prism I < I0 % transmittance = (I/ I0) x 100 = 50% absorbance = 1.7
Visible Spectrophotometer I0 I white light detector/recorder sample solution prism I0 = I % transmittance = (I0 / I) x 100 = 100 absorbance = 0 In order for blue solution to absorb light, wavelength must be changed to blue’s complement
Visible Spectrophotometer white light I0 I detector/recorder sample solution prism Wavelength is changed by either rotating prism or position of slit
“Spec 20” Visible Spectrophotometer digital analog
2. Zero adjust with sample compartment empty : 0 % T
3. 100 % adjust with distilled water (blank) in sample compartment : 100% transmittance
4. Record sample absorbance (Be certain to switch MODE to absorbance)
Beer’s Law • As the intensity of the color of the sample solution increases… • the absorbance increases. • A = b c absorbance constant path length concentration
Determination of the Concentration of an Unknown Solution • Prepare a set of standard solutions of known concentrations • Select the appropriate wavelength • Determine the absorbance of each • Determine the absorbance of the unknown solution
Sample Conc.(ppm) Absorbance 1 0.0 0 2 5.0 .0023 3 10.0 .0045 4 15.0 .0056 5 20.0 .0076 Unknown ? .0051
C = Recall Beer’s Law: A = b c + 0 y = mx + b .0051 = 0.00037 x C + 0.00030 = 13.0