An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Online In-service Training and Competency Roche Accu-Chek® Inform Glucose Meter

  2. Learning Objectives • Accu-Chek® Inform Blood Glucose Meter • Accu-Chek® Comfort Curve Test Strip • Accu-Chek® Inform Quality Control • Hospital Policies & Procedures

  3. Accu-Chek® Inform Meter Components of the system: • Accu-Chek® Inform Meter • Base unit (dock) • Dataport (not shown) • Storage Case

  4. Overview of Front of Meter • Test strip port • Alphanumeric touch screen • Menu button • Forward arrow (serves as an enter key) • Power button (purple)

  5. Overview of Back of Meter • Barcode scanner • Code key slot • Rechargeable battery

  6. Accu-Chek® Inform Base Unit • Return the Inform meter when finished with patient testing to the base unit. The battery recharges and patient results are transmitted to the LIS. • A drained battery will take 1 hour to reach 90% of its full charge, but can be used after 15 minutes of charging to perform a few tests.

  7. Accu-Chek® Inform Base Unit • A fully charged battery provides power for approximately 30 tests/day for 5 to 7 days before it will need to be recharged. • The Inform meter does not have to be docked after each test. • The Inform meter can be docked in any base unit.

  8. Accessory Box • Conveniently stores meter and supplies. • Can be stored “on-end” to save space. • Can be used to transport the Inform meter along with testing supplies when multiple patients must be tested.

  9. Accu-Chek® Comfort Curve Test Strips • Only Accu-Chek® Comfort Curve Test Strips and Glucose Control Solutions can be used with the hospital Inform glucose meter. • A code key is packaged with every box of test strips. The code key must match the number printed on the vial of strips.

  10. Accu-Chek® Comfort Curve Test Strips • The code key is inserted into the code key slot located on the reverse side of the Inform. • Test strips and control solutions are available from pharmacy.

  11. Accu-Chek® Comfort Curve Test Strips • Test strips can be used until the expiration date printed on the vial. • Control solutions must be labeled with the open and discard dates. Opened controls expire 3 months from opening or the “Use by” date on the bottle, whichever comes first.

  12. Accu-Chek® Comfort Curve Test Strips • The Comfort Curve test strip is plastic coated to prevent surface contamination. • Test strips are stored in the tightly capped original container. • Test strips not stored properly will be affected by temperature and humidity.

  13. Accu-Chek® Comfort Curve Test Strips • Improperly stored Test Strips will produce inaccurate results and must be discarded!

  14. Quality Control (QC) • Both low and high level control must be tested every 24 hours. • QC can be tested at any time. Each time QC is tested, the 24-hour clock resets. • IF QC is not tested in 24 hours, the Inform meter prevents patient testing until both low and high level QC are tested.

  15. Quality Control (QC) • QC should also be tested following these conditions: • Patient test results are questionable. • New test strip vial is opened. • New control solutions are opened. • The Inform has been dropped or mishandled. • QC results display as “PASS” or “FAIL” only.

  16. Sample Application

  17. Sample Application • Test strip must be placed in the Inform meter with the yellow test window facing “up” (curved side on the left) before sample is applied. • Apply 4 microliters of sample (patient blood or control solution) to the curved area (yellow test window). • Capillary action draws the sample into the yellow test window. • The yellow test window must be completely filled (no yellow showing). • Additional sample can be applied within 15 seconds to complete filling.

  18. Sample Application • Arterial, venous, and capillary specimens are acceptable. • Note: A comment can be attached to the result identifying the specimen type. • Always wipe away the first drop of blood from a finger stick before touching the test strip to the finger. • Venous blood samples must be collected below an IV. • When using a syringe, always discard a drop before applying the sample to the test strip.

  19. Inappropriate Sample Types • Finger stick or capillary samples are not appropriate from patients exhibiting decreased peripheral blood flow. • In these cases, venous or arterial blood samples should be used. Examples: • Severe dehydration caused by diabetic ketoacidosis • Hyperglycemic hyperosmolar nonketotic state • Hypertension • Shock • Peripheral vascular disease

  20. Inappropriate Sample Types • Certain substances contain Maltose which, when present, cause false elevated glucose results obtained by the Inform. • Patients receiving these products cannot be tested using the Inform. Send all glucose specimens to the laboratory for testing. • Maltose-containing products: • Octagam® 5% • WinRho® SDF liquid • Vaccinia immune globulin • d-xylose • Extraneal® (icodextrin)

  21. Operator and PatientIdentification • Federal, State, and other regulatory entities require positive patient identification and positive test operator identification. • All users must be previously defined using the bar code information located on the reverse of your hospital ID badge. • You can use the Inform bar code scanner to scan the bar code located on your hospital ID badge, OR • Enter your hospital ID using the Inform touch pad. • Non-defined users cannot access any Inform functions.

  22. Operator and PatientIdentification • A patient is identified using their Medical Record Number. • You can use the Inform bar code scanner to scan the patient’s bar coded wristband, OR • Enter the 8 digit MRN using the Inform touch pad. • Patient testing cannot be initiated without entry of an MRN. • Patient test results cannot be posted to the LIS without a valid MRN.

  23. Hospital Policies & Procedures Normal and Critical Ranges • Normal range for a glucose test result obtained by the Inform is: 60 – 100 mg/dL. • Critical glucose test results obtained by the Inform are: • Results <40 mg/dL (adult) • Results <50 mg/dL (neonate) • Results >500 mg/dL (all) • LO test results (<10 mg/dL) and HI test results (>600 mg/dL) are critical results.

  24. Hospital Policies & Procedures Normal and Critical Ranges • The Inform requires entry of a comment to a critical test result. • Multiple pre-defined comments are available. • User can also enter a free text comment. • Up to three comments can be appended to a test result.

  25. Hospital Policies & Procedures Normal and Critical Ranges • Repeat all critical test results obtained by the Inform. • Verify critical test results by sending a specimen to the laboratory. • Glucose test results obtained by the Inform not corresponding to the patient’s medical condition should be confirmed by submitting another blood sample for laboratory testing.

  26. Testing Personnel • Only trained personnel are permitted to perform point of care glucose testing. • Operators will be permitted access to the Inform meter only after successful completion of an approved initial training program. • Each user’s competency using the Inform must be assessed annually.

  27. Accu-Chek®Certification • You must also maintain your Accu-Chek®certification (access) to continue using the Inform. • A new user receives certification upon completion of their initial training. The POC Coordinator is responsible to “certify” the user. The initial certification period is 6 months. • Existing users recertify by testing both Hi and Lo QC on a single Inform within 30 days of their certification expiration. • Recertification allows access for the next 12 months.

  28. Infection Control • The Inform is routinely decontaminated (every 24 hours) using: • 70% isopropyl alcohol pads, OR • pads moistened with 10% bleach solution. • Decontaminate when visibly soiled with blood or after exposure to patients in isolation conditions. • Do not allow excess amounts of cleaning solution to pool in the meter connectors or base unit. • Do not use “Sani-Cloths” to decontaminate the Inform.

  29. Infection Control • Testing personnel must follow hospital infection control guidelines at all times. • Users must wear gloves during sample collection and testing. • Sharps must be discarded in puncture proof containers. • All other biohazardous waste must be discarded in designated biohazard containers.

  30. Support • The POC Coordinator is responsible for maintaining the Inform system. • Please contact the POC Coordinator (215-955-1664) for assistance with device, user access, or other issues.

  31. Next Steps • Upon completion of the course and passing the quiz, please notify your instructor, supervisor, or section head. • The POC Coordinator must be notified so you can be “certified” providing access to the Inform. • You will be instructed to test both Hi and Lo controls on a device before a specific date.

  32. Congratulations • You have completed the course.