An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Application of Time-Frequency Analysis :DiagnosticDoppler Ultrasound Date:2013/12/12 Reporter:林宗民

  2. Outline • Basic Concepts • Doppler Signal • Spectral Doppler • Doppler Image(Color Doppler, Power Doppler) • Conclusion

  3. Ultrasound • Longitudinal wave • Frequency > 20 KHz • Diagnosis Ultrasound:1~13 MHz • Velocity : • 343m/s in air • 1540m/s in tissue • 6420m/s in aluminum http://i299.photobucket.com/albums/mm288/wasi142/Biological%20classification/waves.jpg

  4. Doppler Effect • Frequency shift due to relative movement. • Doppler frequency:fd = f – f0

  5. Ultrasound Image – B mode • Brightness-mode • Backscattered image • Tomography http://content.answcdn.com/main/content/img/McGrawHill/Encyclopedia/images/CE757525FG0010.gif

  6. Doppler Signal • Doppler Signal, xd(t) = iFT{ fd = fr – ft} • Typically ~kHz in diagnosis Doppler ultrasound.

  7. Doppler Signal Continuous Wave • Transmitting monochromatic wave. Transmitted signal spectrum

  8. Doppler Signal Continuous Wave • Transmitting monochromatic wave. • Received signals from moving tissue plus static tissue. fstatic fmoving Received signal spectrum

  9. Doppler Signal Continuous Wave • Transmitting monochromatic wave. • Received signals from moving tissue plus static tissue. • Demodulation and filtering. fstatic fmoving Doppler signal spectrum

  10. Doppler Signal Continuous Wave • Transmitting monochromatic wave. • Received signals from moving tissue plus static tissue. • Demodulation and filtering. fmoving Doppler signal spectrum

  11. Doppler Signal Pulse Wave • Transmitting several pulses. • Multi-gate sampling.

  12. Doppler Signal Pulse Wave • Transmitting several pulses. • Multi-gate sampling. • Doppler signal

  13. Spectral Doppler (Sonogram) • Continuous wave or Pulse wave • Combining B-mode ultrasound with pulsed wave Doppler • STFT has been commonly used.Unbalanced Sampling with ↑50% overlap. http://en.wikipedia.org/wiki/File:SpectralDopplerA.jpg

  14. Simulation Analysis • Simulated Doppler signal. • Using STFT, Cohen’s class Distribution. • Choi-Williams Distribution • Bessel Distribution • Cone-Shape Distribution

  15. Simulation Analysis Choi-Williams Distribution STFT Bessel Distribution Cone distribution

  16. Simulation Analysis Choi-Williams Distribution STFT Bessel Distribution Cone distribution

  17. Color Doppler • Combining B-mode and Doppler signal. • Pulse Wave  Multi-gate system. • Velocity and direction of blood flow are the desire information. • 2 spatial dimension +1 temporal dimension amount computation. • Auto-correlation method for calculating average frequency.

  18. Power Doppler • Combining B-mode and Doppler signal. • Pulse Wave  Multi-gate system. • Total power of Doppler signal. • Indicating the flow density information. • Not T-F analysis

  19. Conclusion • Image is 3-D signals. It’s not proper to use T-F analysis. • Spectral Doppler is 1-D signals. T-F analysis has been used to measure the blood flow. • In practice, diagnosis signals are more complex than simulated signals, and more noise. Statistic methods are used.

  20. Reference • “The Difference Between Colour Doppler Velocity Imaging and Power Doppler Imaging,” W. N. McDicken and T. Anderson, Eur J Echocardiography(2002)3,240–244 • “Time-Frequency Analysis for Doppler Ultrasound signals,”陳宗銓, http://djj.ee.ntu.edu.tw/Tutorial_Time-Frequency%20Analysis%20for%20Doppler%20Ultrasound%20Signals.docx. • “Physics and instrumentation of Diagnostic Medical Ultrasound,” PETER FISH, 1990. • “Evaluation of fetal arrhythmias from simultaneous pulsed wave Doppler in pulmonary artery and vein,” Julene S Carvalho, Federico Prefumo, Valentina Ciardelli, ShanthiSairam, AmarnathBhide, Elliot A Shinebourne, 2007. • “The Physical Principles of Doppler and Spectral Analysis,” Peter N. Burns, PhD, J Clin Ultrasound 15567-590. NovemberlDecember1987.