Manufacturers and Filters – Is it all still a bit fuzzy?

Manufacturers and Filters – Is it all still a bit fuzzy?. Bill Thomson, James Cullis, Joe O’Brien Physics & Nuclear Medicine City Hospital. Birmingham. Background. MPI Images. Philips Odyssey Butterworth filter: Order = 6.0 Cut-off = 0.24 cycles / pixel.

colum
Download Presentation

Manufacturers and Filters – Is it all still a bit fuzzy?

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Manufacturers and Filters – Is it all still a bit fuzzy? Bill Thomson, James Cullis, Joe O’Brien Physics & Nuclear Medicine City Hospital. Birmingham

  2. Background

  3. MPI Images Philips Odyssey Butterworth filter: Order = 6.0 Cut-off = 0.24 cycles / pixel 6 mm pixel , = 0.4 cycles / cm G* X*l*r*s Manufacturer recommendations Butterworth filter: Power = 5 Critical Freq. = 0.5 cycles/cm G* X*l*r*s Butterworth filter: Power = 6 Critical Freq. = 0.4 cycles/cm

  4. What’s Going on ? Variation in filter definition ? Companies often make subtle changes to things

  5. Butterworth Filter Filter shape set by order and cutoff Fix cut-off = 0.3 pixel-1 Alter the ORDER

  6. Butterworth Filter Fix order = 6 Alter the CUT-OFF

  7. Effect of Varying Cut-off 0.12 pixel-1 0.18 pixel-1 0.24 pixel-1 0.3 pixel-1 0.36 pixel-1

  8. How Does All This Help? • More higher frequency terms, more noise • Need the ‘lower-mid’ frequencies for structure

  9. Butterworth Filter Comparison Philips: Order 6 , Cut-off 0.4 cm-1 GE: Power 6 , Critical Freq. 0.4cm-1

  10. Filter Definitions PhilipsGE 1/2 ( ) Order Number n Power = 2n Cut-off at 0.5 critical freq. = 0.707

  11. Butterworth Filter Comparison Philips: Order 6 , Cut-off 0.4 cm-1 GE: Power 12 , Critical Freq. 0.4cm-1

  12. Butterworth Filter Comparison Philips: Order 6 , Cut-off 0.4 cm-1 GE: Power 12 , Critical Freq. 0.37cm-1

  13. PhilipsGE Short Axis Comparison Order 6.0 Cut-off 0.4 cycles/cm Power 12.0Critical Freq.0.37 cycles/cm

  14. Manufacturers Comparison Cut-off Definition Cut-off units Order number Philips n 0.5 per pixel Philips n 0.5 per 2 pixels GE power = 2n 0.707 per cm-1 Hermes n 0.5 per cm-1 Siemens n 0.5 per Nyquist* * ( Nyquist is 0.5 pixel -1 )

  15. Published recommendations QGS order = 5 , cut-off 0.41 cycles per cm-1 EANM guidelines Says ‘ correct cut-off is at 0.707 ‘ Cut-off = 0.3 – 0.4 Nyquist , pixel size = 6mm Different cut-off values depending on MBq. Definition of Nyquist and pixel size relationship is wrong. BNMS Order 5 or 10 , cut-off 0.5 cycles per cm-1 (? Pixel size ) but cut-off = 0.75 for Hanning ? SNM Useless

  16. Published recommendations (2) SNM Useless G Germano, Cedars-Sinai* Equates cut-off and critical frequency States is at 0.5, but graphs show both 0.5 and 0.707 * “Technical Aspects of Myocardial Imaging”JNM, 42: no 10 1499-1507 , 2001

  17. Summary • Take great care! • Check filter definition and units • Don’t always trust the manufacturer to be right • Convert units and double-check! • Beware! Confusion in other areas eg DaTSCAN