IPRD10 Siena, June 7-10 2010 . Calibration of the CMS Electromagnetic Calorimeter with first LHC data. Maria Margherita Obertino ( Universita ’ del Piemonte Orientale – INFN Torino) On behalf of the CMS Collaboration. OUTLINE. The CMS Electromagnetic Calorimeter (ECAL)
E N D
IPRD10 Siena, June 7-10 2010 Calibration of the CMS Electromagnetic Calorimeter with first LHC data Maria MargheritaObertino (Universita’ del Piemonte Orientale – INFN Torino) On behalf of the CMS Collaboration
OUTLINE • The CMS Electromagnetic Calorimeter (ECAL) • Pre-calibration: crystal intercalibration beforethe LHC startup • In situ calibration: crystal intercalibration with collision data • Results with very first collisions • Calibration data streams • 2010 plans M.M.Obertino – IPRD10
ECAL • Hermetic homogeneous calorimeter made of PbWO4 • Crystals • Barrel (EB) [|h|<1.48 ] • 36 Supermodules (SM) with 1700 crystals each • Photodetectors: APD • Endcaps (EE) [1.48<|h|<3.0 ] • 2 Dees each side, 14648 crystals in total • Photodetectors: VPT • Preshower (ES) [1.65<|h|<2.6 ] • Sampling calorimeter (lead, silicon strips) EE Supermodule ES EE Excellent energy (and position) resolution for photons and electrons crucial for studying interesting physics channel (Hgg, HZZ4e, …) • Target value: 0.5% • Crucial dependence on the channel • to channel inter-calibration precision. M.M.Obertino – IPRD10
PRE-CALIBRATION Extensive10 years long pre-calibration program carried out before ECAL installation in CMS. Test Beam at CERN 90-120 GeV electrons (2004-2007) [Intercalibration of the barrel electromagnetic calorimeter of the CMS experiment at start-up - 2008 JINST 3 P10007 ] Beam • 10 EB SMs • 460/14648 EE crystals Number of crystals Intercalibration precision ~0.3% dominated by statistical uncertainty, determined comparing two independent data sample collected in the same conditions. M.M.Obertino – IPRD10
PRE-CALIBRATION Light yield (LY) measurements with 60Co (1MeV g) during crystal qualification (2000-2006) • EB:LY Measurements • Intercalibration precision ~4.5-6.0% • EE:LY + VPT gain Measurements • Intercalibration precision ~8% M.M.Obertino – IPRD10
PRE-CALIBRATION Light yield (LY) measurements with 60Co (1MeV g) during crystal qualification (2000-2006) • EB:LY Measurements • Intercalibration precision ~4.5-6.0% • EE:LY + VPT gain Measurements • Intercalibration precision ~8% • 36 EB SMsoperated on a cosmic ray stand for a period of ~1 week/SM • Deposited energy per crystal ~ 250 MeV Cosmic muons Channel intercalibration with cosmic muons (2006-2007) Mean intercalibration precision ~1.5% EB SM h=0 h=1.48 M.M.Obertino – IPRD10
PRE-CALIBRATION Beam splash: in September 2008 and November 2009 beam was circulated in LHC, stopped in collimators 150 m away from CMS • All Endcapcrystals Intercalibration precision ~6% BEAM CLOSED COLLIMATOR Combination of LY and beam splash gives ~5% intercalibration precision over the entire EE M.M.Obertino – IPRD10
ACHIEVEMENT WITH PRE-CALIBRATION • EE (from 27% to 5%) • Test beam (460 crytals) • Combination of LY and beam splash • EB(from 11% to 0.3-2.2%) • LY, cosmic, test beam. • Combination strategy: • Select best calibration available • Combine when comparable precision from two sources 27% 11% RMS of the constant distribution gives indication of inhomogeneity at the construction M.M.Obertino – IPRD10
ENERGY SCALE: ADC to GeV • ECAL energy scale fixed by Test Beam data in Barrel and Endcap separately • ~63 MeV/ADC in EE • ~39 MeV/ADC for EB Distribution of 5x5 amplitude sum (S25) with and without intercalibration for EE crystals exposed to 90 GeV electron beam . • Confirmed later in dE/dx • analysis with cosmics(*) and in collisions withp0 and hmass peak. • Waiting for higher mass resonances to check the scale • in-situ at high energy. 90 GeV INTERCALIBRATION (*) presented by F.DeGuio M.M.Obertino – IPRD10
IN-SITU CALIBRATION • Several methods to calibrate in-situ: • f-symmetry method: intercalibration for crystals at the • same pseudorapidity. • Based on invariance around the beam axis of the energy flow in minimum • bias events. • Calibration performed comparing the total energy deposited in each • crystal with the mean of the distribution of the total energies for all crystals • at the same pseudorapidity M.M.Obertino – IPRD10
IN-SITU CALIBRATION • Several methods to calibrate in-situ: • f-symmetry method: intercalibration for crystals at the • same pseudorapidity. • Based on invariance around the beam axis of the energy flow in minimum • bias events. • Calibration performed comparing the total energy deposited in each • crystal with the mean of the distribution of the total energies for all crystals • at the same pseudorapidity • p0 and h method: single crystal intercalibration • Constants derived using unconverted g’s from p0 and h decay reconstructed • in 3x3 matrices • The reconstructed mass is used to determine the photon energy M.M.Obertino – IPRD10
IN-SITU CALIBRATION • Several methods to calibrate in-situ: • f-symmetry method: intercalibration for crystals at the • same pseudorapidity. • Based on invariance around the beam axis of the energy flow in minimum • bias events. • Calibration performed comparing the total energy deposited in each • crystal with the mean of the distribution of the total energies for all crystals • at the same pseudorapidity • p0 and h method: single crystal intercalibration • Constants derived using unconverted g’s from p0 and h decay reconstructed • in 3x3 matrices • The reconstructed mass is used to determine the photon energy • High energy electrons fromWenand Zeedecays (J/yunderstudy) • Higher integrated luminosity required. Helpful at the startup only for energy • scale. M.M.Obertino – IPRD10
SUPERMODULE RELATIVE SCALE First calibration with collisions of the barrel. • First results in 2009 with ~0.01 nb-1 collected at • Scale derived in 2010 with ~0.4 nb-1 collected at • Systematic error: 0.5% (see next slide) • Statistical error negligible w.r.t. systematic one CMS Preliminary CMS Preliminary Tested with beam • f-symmetry data • p0 data M.M.Obertino – IPRD10
SUPERMODULE RELATIVE SCALE In-situ SM relative scale defined as the weighted average of p0 andf-symmetry results. Distribution of the relative scale of 10 SMscalibrated with test beam electrons. Distribution of the relative scale of 26 SMs calibrated only with cosmic rays. CMS Preliminary CMS Preliminary The RMS is 1.2%±0.2% is consistent with the expected precision of cosmic ray scale. The RMS of 0.5%±0.1% estimates the current precision of the combinedin-situSM relativescale calibration. M.M.Obertino – IPRD10
RECONSTRUCTED p0 gg PEAK • L = 0.43 nb-1 • PT (g) > 0.4 GeV, PT(ggpair) > 1.0 GeV(pure ECAL selection) • Good agreement between data and MC. • Number ofp0 from the fit: 1.46 106 MC DATA The fitted mass agrees with the expectation to within 1% (in situ validation of absolute enegry scale) M.M.Obertino – IPRD10
RECONSTRUCTED hgg PEAK • L = 0.43 nb-1 • PT (g) > 0.5 GeV, PT(ggpair) > 2.5 GeV(pure ECAL selection) • Good agreement between data and MC • Number ofh from the fit: 2.55 104 DATA MC M.M.Obertino – IPRD10
CALIBRATION DATA STREAM In CMS dedicated paths of data acquisition (streams) are implemented for calibration. In p0 and f-symmetry streams, only few tens of “useful”crystal hits are stored for each accepted event (output limited to 1kHz per stream). Calibration stream commissioned with collision data. p0 peak from dedicated calibration stream (better S/B because of higher transverse energy cuts in calibration stream) M.M.Obertino – IPRD10
PLANS FOR 2010 In the Barrel • improve crystal calibration down to 1% in the whole EB with few pb-1. • reach the goal of 0.5% precision in EB with ~10pb-1, expected by the end of the year Fluctuation due to material in front of ECAL • Improve Endcapcrystal calibration from 5% to 1-2%. • Set the absolute scale to few permille. M.M.Obertino – IPRD10
SUMMARY • ECAL crystal intercalibration in very good shape. 0.3%-2% in EB 5% in EE • First calibration result with collision data: SM relative scale in EB. • p0and f-symmetry in agreement. • Calibration streams commissioned. • Collected statistics (~17 nb-1 at 26th May) good for regional • intercalibration (5x5 matrix). • 2010 goals: 0.5%intercalibrationprecision inEB • 1-2%intercalibrationprecision in EE M.M.Obertino – IPRD10