1 / 23

BOSTON Police Department Crime Laboratory Amy Reynolds, Erin Corcoran, Emily runt

Instrumental Validation of a Scanning Electron Microscope with Energy Dispersive X-ray Spectroscopy (SEM/EDS). BOSTON Police Department Crime Laboratory Amy Reynolds, Erin Corcoran, Emily runt. SEM/EDS Acquisition. Sept. of 2014 our previous Amray SEM with PGT EDS stopped working

dedwards
Download Presentation

BOSTON Police Department Crime Laboratory Amy Reynolds, Erin Corcoran, Emily runt

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Instrumental Validation of a Scanning Electron Microscope with Energy Dispersive X-ray Spectroscopy (SEM/EDS) BOSTON Police Department Crime Laboratory Amy Reynolds, Erin Corcoran, Emily runt

  2. SEM/EDS Acquisition • Sept. of 2014 our previous Amray SEM with PGT EDS stopped working • After much research and soliciting for funds • JEOL IT300LV SEM with an Oxford X-MaxN 50 EDS • Purchased in 2016 • AZtec software • Trace Evidence analysis • Paint • Tape • Glass • General Unknowns • Gunshot primer residue analysis

  3. SEM/EDS Validation • ANAB 17025 Accreditation • 5.4.5, Supplemental and our own labs validation requirements • Spoke to several Forensic Scientists in the field about their validations • John Druggan, MA State Police Crime Laboratory • Matney Wyatt, US Army Criminal Investigation Laboratory • Sent their validation plan for validating an SEM/EDS for GSR • David Edwards, JEOL • Richard McLaughlin, Oxford Instruments • Started out with validation of the instrument for trace evidence samples other than Gunshot primer residues

  4. Magnification and Measurement Tools • Copper Grid (200 Mesh, Pelco XCS-6 Standard) • 3 locations on grid • One edge of the mesh to the adjacent edge • 2 magnifications (350x and 550x) • JEOL and Oxford image • Average measurements from the 3 locations at each magnification, compared to expected 127µm grid spacing • JEOL- 350x was 126.8µm • JEOL- 550x was 127.6µm • Oxford- 350x was 126.9µm • Oxford- 550x was 126.6µm • Under 1µm difference observed was acceptable

  5. Verify Factory Resolution –127kV for Mn • Manganese (Pelco XCS-6 Standard) • FWHM • Process Time 4 and 6, 20kV • Determine height of K peak, left side of peak at half height subtracted from right side of peak at half height • Number of channels counted at half width, multiplied by the amount of eV’s per channel • Average of 3 spectra for each method determined • Process time 4: approx. 133eV • Process time 6: approx. 127eV • Protocol-Process time 6 is used, any eV under 127eV is acceptable for QC, or +/- 10eV of 127eV Height FWHM

  6. Verify X-ray Range • Full Scale Verification • Molybdenum (Mo) aperture • 3 spectra at 30kV • Mo peaks observed at approx. 2.3keV (2.266keV) & 17.5keV (17.446keV) • Low eV Limit Verification • Boron (Pelco XCS-6 Standard) • 3 spectra at 20kV • B peak observed at 0.179keV (0.179keV) • Beam Energy Range • Manganese (Pelco XCS-6 standard) • 1 spectra each at 20kV & 25kV • Bremsstrahlung Curve observed to drop off at approx. 20kV (20kV run) and 25kV (25kV run)

  7. Verify Consistency of Probe Current • Manufacturer-Probe current over the course of a run <1% difference per hour • Current meter was purchased • rbd 9103 AutorangingPicoammeter • Old aperture placed over the top of a hole in stub holder (Faraday cup) • Focused upon the aperture hole and magnified until the aperture hole completely covered field of view • Filament stabilized for ½ hour • Probe current data collected for 9.69 hours • Difference in probe current data over 9 hours was approx. 2.6% • <1% per hour, 0.28% per hour

  8. Old vs. New • Do we get the same or better results with the new SEM/EDS vs. the old SEM/EDS? • Used old proficiencies to answer this question • Tape- 2010 Items 1 and 2 (electrical), 2012 Items 1-3 (duct) • Paint- 2012 Items 1-3 (automotive), 2013 Items 1 and 2 (architectural) • Glass- 2011 Items 1 and 2, 2012 Items 1 and 4 • Determine the best parameters • Used different Pa in Low Vacuum (LV) mode (30, 50 and 70) • Observed charging at low Pa • Observed contamination from other paint layers or the environment at high Pa due to beam spread • Used different kV (10, 15, and 20) • Unable to obtain all of the elements found in previous spectra with lower kV • Observed contamination from the other paint layers or the environment with higher kV • Determined that the samples needed to be sputter coated with carbon

  9. Carbon Sputter Coating • Denton Vacuum Desk II Sputter/Etch Unit and Carbon Evaporation accessory • Resurrected and sent off for maintenance • Determine the best height of the carbon chuck for proper carbon coating • Heights- (5.5, 7.5 and 9.5 centimeters above the samples) • Higher the carbon chuck the lighter the coating • Lower heights could burn the sample • 7.5 cm height chosen • Slight charging still observed on the electrical tape and backing of duct tape • Carbon coat twice or use low Pa in LV mode

  10. Proficiency Comparisons • Glass Proficiency • CTS 11-548 and CTS 12-548 old system vs. new system • All the same major peaks observed • Paint Proficiency • CTS 12-546 old system vs. new system • Same major peaks observed plus more peaks with new system • Better Quality Detector • CTS 13-546 old system vs. new system • Same major peaks observed • Tape Proficiency • FTS 10-Tape and FTS 12-Tape old system vs. new system • Same expected peaks observed plus more peaks with the old system • Contamination

  11. Protocol for SEM/EDS • JEOL SEM with Oxford EDS was comparable with the reliability, reproducibility and sensitivity to the old Amray SEM with PGT EDS • Protocol adjustments • Daily QC • Mn is run at 10mm, 20kV, Aperture 3, Process Time 6 • FWHM at 5.9eV should be below 127eV, but at least within +/-10eV from 127eV • Charging samples • May use LV mode below 30Pa, starting at 15kV • Highly recommended to carbon coat the samples • If after carbon coating the sample charging is still observed: • Increase the working distance (possible loss of elemental information) • Decrease the accelerating potential (possible loss of elemental information) • Recoat the sample with another layer of carbon (best option) • Parameters to start: • 10mm, 20kV, Aperture 3, Probe current at a level to observe 40-50% deadtime, Secondary detector for good image quality, Backscattered detector-Compositional for good elemental analysis

  12. Gunshot Primer Residue Validation • AZtec software • Seems like all other labs with Oxford are using INCA software • ENFSI/PLANO standard • Determine the best parameters to be used • 90% of the >1µm particles • 10mm, 20kV, Process time 4, Aperture 2, these parameters stayed the same • Gray level threshold changes first • 8000, 8800, 9000, 10000 at 3µs first pass • First pass µs changed second • 5, 4, 3, 2, 1 µs at 8800 gray level threshold • Magnification changed last • 190x, 250x

  13. Gunshot Primer Residue Validation • Each different parameter run was done in at least triplicate • Data from the runs were exported to Excel • X and Y coordinates were plotted to create an overlay on the PLANO standard plot • Determine the % of the particles >1µm • After countless runs with the different parameters using the PLANO standard • Several discussions and visits with both Dave Edwards (JEOL) and Richard McLaughlin (Oxford Instruments) • We got the best parameters for a Gunshot primer residue run nailed down • 10mm, 20kV, Aperture 2, Process time 4 • Image Scan size: 2048, Dwell time: 2µs, Gray level threshold: 8800 • Second pass imaging: 20µs, Acquisition mode: Live Time of 0.5s, Second Pass EDS filter: Lead containing features with an additional time of 1.5s • 250x magnification

  14. Negative Control Study • 5 air blank samples using the optimum parameters • 5 lab spaces, SEM stubs were left open for 3 working days • Serology lab, pre-SEM room, SEM room, Trace lab, Evidence Receiving area • Results • Zero characteristic GSR particles found on any of the stubs • Some consistent GSR particles found • Found on the actual aluminum stub surface not the adhesive tab • Negative control • Air blank from the SEM room or pre-SEM room (sample prep area)

  15. Known GSR stub • Stub collected directly after shooting occurred • Test the optimum parameters • Stopped after 100 characteristic particles detected • Test the reacquisition portion of the software • Change to GSR Reacquisition Program • 15 features ranging from 20µm to 1µm were reacquired-Automatically • Confirmed 3 features >9µm in size • Determined that if feature is <8-9µm in size it will not relocate properly to those features • Point and ID as an end around • Unacceptable • Version 3.4 now has a Manual Reacquisition

  16. Known GSR stub • Version 3.4 was installed in Nov 2017 • Manual selection inthe GSR Reacquisition program • Relocate to a feature • Magnify and focus on this feature • Enter this magnification into the program • Start reacquisition • Need to find and refocus on the particle at the previously determined magnification • Reacquires a better image and spectrum • 10mm WD, 20kV, Aperture 2, Process Time 4 • Image Scan size: 512, Dwell time: 5µs, Gray level threshold 5000 • Acquisition mode: Live Time of 10s • Confirm the reacquired features • Create a report

  17. Weapon and Time of Collection Study • Currently running analysis on a weapon and time of collection study • 5 different weapons, 5 different shooters • Glock 9mm pistol with PMC bronze FMJ 9mm Luger • Smith & Wesson 38 revolver with PMC bronze FMJ 38 special • Hi Point 9mm pistol with Federal American Eagle FMJ 9mm Luger • Walther 380 auto pistol with Winchester FMJ 380 Auto • Ruger 357 Magnum revolver with Blazer Total Metal Jacket 38 Special + P • 4 different times since collection • 0 hrs, 2 hrs, 4 hrs and 6 hrs after shooting

  18. Preliminary Results-Weapon and Time of Collection Study Shooter #1- Glock 9mm pistol

  19. Preliminary Results-Weapon and Time of Collection Study Shooter #2- Smith & Wesson 38 Revolver

  20. Contamination Study • Prep area with the Denton Vacuum was used as our GSR Distance Determination area previously • Adjacent to the SEM room • Since been moved to another location in the lab • Stubs were collected from the lab bench, fume hood and examination table • Before cleaning • After cleaning • Determine if any contamination is observed from these locations for future casework • Clean again? • Retest areas with further stubs?

  21. Training Program • Training Courses • JEOL JSM-IT300LV operator’s training course • Oxford AZtec operator’s training course with AZtec GSR • Gunshot Residue Identification-Wayne Niemeyer, Hooke College • Readings • ASTM Standard Guide for Gunshot Residue Analysis by Scanning Electron Microscopy/Energy Dispersive X-Ray Spectrometry (newest version) • SWGGSR Guide for Primer Gunshot Residue Analysis by Scanning Electron Microscopy/Energy Dispersive X-Ray Spectrometry • Summary of the FBI Laboratory’s Gunshot Residue Symposium, 2005 • Related journal articles • Workshops • Studies using brake pad stubs and fireworks stubs • Run 2-3 of the Time of Collection sets • Old proficiency test(s) • Competency • Practical- Old proficiency or a Mock case created for our analysts • Theoretical- Competency questions and Mock trial

  22. Thank you • MatneyWyatt • US Army Criminal Investigation Laboratory • John Druggan • Massachusetts State Police Crime Laboratory • Rusty White • Texas DPS, Austin Crime Laboratory • Mike Martinez • Bexar County Crime Laboratory • Dave Edwards • JEOL • Richard McLaughlin • Oxford Instruments • Everyone else in the SEM/EDS and GSR community

  23. Questions?

More Related