An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Vibration Analysis of a Cryocooler for CDMS Lauren Coutant SULI, summer 2007

  2. What is CDMS? • Cryogenic Dark Matter Search (CDMS) • It is known that dark matter exists [1] • Best theory: it is Weakly Interacting Massive Particles (WIMPs) [2] • CDMS is a direct detection experiment • Detects the energy transferred when a WIMP has a collision with a nucleus in the detector • Detector is kept at 50mK with a Gifford McMahon cryocooler and dilution refrigerator

  3. Two Types of Cryocoolers • Gifford-McMahon (GM) • Uses a piston to push the working fluid through the system • Piston is located in the cryocooler • Pulse Tube (PT) • Uses orifice and reservoir instead of piston • Has no moving mechanical parts inside the cryocooler [3] [4]

  4. Stirling (GM) versus PT Cycle Stirling Pulse Tube [5]

  5. Potential Problems from Cryocooler • All Cryocoolers have some vibration • Vibration can cause • Heat • Electrical noise, ~T5 • “Vibration free” • Vibration on the order of background seismic motion [6]

  6. What Causes the Vibrations? • GM • Piston causes motion in the system • PT • Fluid traveling through impedances—think of musical instruments • Both • Pressure changes of the working fluid cause pipes to deform—think of blood vessels [7]

  7. Which one is better? [8] p. 314

  8. What type was tested here? • Cryomech Pulse Tube 415 • Two Stage • 1st Stage of the Cryocooler--room temperature to 77K • 2nd Stage--77K to 4K [9]

  9. Vibration Analysis

  10. Accelerometer Adapted from [10] • Piezoelectric material experiences charge separation when a force is applied • Voltage can be measured and related to the magnitude of the force • Force can be used to find acceleration, F=ma

  11. Vibration Testing: Accelerometer Location

  12. Vibration Testing: Block Diagram

  13. Vibration Analysis: Lab 3 Set up

  14. Results As a reference, g=9.8 m/s^2

  15. Comparison • How does this relate to what Tomaru’s group got? • How does this relate to the performance of the current CDMS cryocooler?

  16. Vibration Reduction Methods • Distance • Cold head placed away from cryocooler [3] • Was implemented in Lab 3 • Flexible material to absorb vibrations • Rigid material to resist elastic deformation • Active vibration cancellation

  17. Vibration Reduction Methods: Materials and Counter-weight Adapted from [11] p. 118

  18. Vibration Reduction Methods: Distance and Materials [8] p. 310

  19. Vibration Reduction Methods: Flexible and Rigid Materials [12]

  20. Vibration Reduction Methods: Flexible Connectors Currently in use at Soudan CDMS site.

  21. Vibration Reduction Methods: Flexible Connectors

  22. Vibration Reduction Methods: Active Cancellation [13]

  23. Vibration Reduction Methods: Active Cancellation Single Pipe Four Pipes Six Pipes [7]

  24. Conclusions • Vibration of PT415 should be less than current GM cryocooler at Soudan • More analysis needs to be done • Any of the suggested methods of reduction could be applied. • Adding flexible materials--least work, but also least reduction • Active cancellation--work intensive, but great reduction, may not dampen vibrations at high frequency

  25. Works Cited • Reisetter, Angela. “Results from the Two-Tower Run of the Cryogenic Dark Matter Search.” University of Minnesota. 2005. • “SuperCDMS Development Project.” SuperCDMS Collaboration. Proposal to DOE and NSF. 2004. • Radebaugh, Ray. “Development of the Pulse Tube Refrigerator as an Efficient and Reliable Cryocooler.” Institute of Refrigeration, London. 2000. <http://cryogenics.nist.gov/Papers/Institute_of_Refrig.pdf>. • Courtney, Bill. “Cryocoolers-Refrigerators producing very low temperatures” Accessed on 6-15-2007. <http://www.cheshire-innovation.com/news/cryocooler/htm> Cheshire Innovation. • Yuan, Sidney. “Pulse Tube Coolers (Cycle, Engines, Cryogenics Refrigerators, Cryocoolers).” <http://www.yutopian.com/Yuan/PT.html> 2000. • Uchiyama, T., K. Kuroda, M. Ohashi, et al. “Present status of large-scale cryogenic gravitational wave telescope." Classic Quantum Gravity V 21. 13 Feb. 2004: p. 1161-1172.“Cryomech PT415.” Cryomech, Inc. 2007. <http://www.cryomech.com/DEVELOPMENTS/PT415.htm> • Suzuki, T., T. Tomaru, T. Haruyama, et al. "Ultra-Low Vibration Pulse Tube Cryocooler with a New Vibration Cancellation Method ." Advances in Cryogenic Engineering: Transactions of the Cryogenic Engineering Conference. V. 51. 2006 American Institute of Physics. • Tomaru, Takayuki; Toshikazu Suzuki; Tomiyoshi Haruyama; et al. "Vibration analysis of cryocoolers." Cryogenics. V44, 2004. p. 309-317. • “Cryorefrigerators: Pulse Tube.” Cryomech Inc. 2007. <http://www.cryomech.com/PULSE%20TUBES.htm> • “Function of Piezoelectric Accelerometers.” PCB Piezotronics. PCB Group, Inc. 2007. <http://www.pcb.com/ techsupport/tech_accel.php>. • Veprik, A.M., S.V. Riabzev, G.S. Vilenchik, et al. “UItra-Iow vibration split Stirling linear cryogenic cooler with a dynamically counterbalanced pneumatically driven expander.” Cryogenics. V4, 2005. p 117-122. • Tomaru, Takayuki; Toshikaza Suzuki; Tomiyoshi Haruyama; et al. “Development of a Vibration-Reduction System of Cryocooler for a Cryogenic Interferometric Gravitational Wave Detector.” • Suzuki, T., T. Tomaru, T. Haruyama, et al. "Pulse Tube Cryocooler with Self-Cancellation of Cold Stage Vibration." Cryo Prague 2006. <http://arxiv.org/abs/physics/0611031>.